Int J App Pharm, Vol 9, Issue 6, 2017, 34-38Original Article

SIMPLE ANALYTICAL METHOD FOR THE SIMULTANEOUS ESTIMATION OF HYDROCHLOROTHIAZIDE AND CANDESARTAN BY RP-HPLC

MADHAVI K.1*, NAVAMANI M.1, PRASANTHI C.2

1Institute of Pharmaceutical Technology, Sri Padmavati Mahila Visvavidyalayam (Women’s University), Tirupati, A. P., India, 2Krishna Teja Pharmacy College, Tirupati, A. P. India
Email: kuchanamadhavi@yahoo.co.in

Received: 15 Jun 2017, Revised and Accepted: 10 Oct 2017


ABSTRACT

Objective: To develop a simple, rapid, economic, accurate and precise reverse phase-high performance liquid chromatographic (RP-HPLC)method for the determination of hydrochlorothiazideand candesartan in thepharmaceutical dosage form and to validate as per international conference on harmonization(ICH) guidelines.

Methods: The chromatographic separation was performed on Silanol BDS C18 column (250 x 4.6 mm, 5 μm), a mobile phase consisting of water (pH adjusted to 2.8 with orthophosphoricacid):acetonitrile (30:70 %v/v), with a flow rate 1 ml/min and the detection wavelength of 210 nm using photodiode array (PDA)detector.

Results: The developed method resulted in elution ofhydrochlorothiazide at2.28 min and candesartanat 4.28 min. The calibration curves were linear (r2=0.999) in the concentration range of 6.25-18.75 μg/ml and 8-24μg/ml for hydrochlorothiazide andcandesartanrespectively.The percentage recoveries were found to be 99.78-100.39 for hydrochlorothiazide and 99.87-100.64 for candesartan. The limit of detection (LOD) was found to be 0.410 μg/ml and 0.699 μg/ml for hydrochlorothiazide and candesartanrespectively. The limit of quantitation (LOQ) was found to be 1.367 μg/ml and 2.330 μg/ml for hydrochlorothiazide and candesartanrespectively.

Conclusion: A simple, economic, accurate, precise, linear and rapid RP-HPLC method was developed for simultaneous quantitative estimation of hydrochlorothiazide and candesartan in bulk and pharmaceutical formulation and the method was validated as per ICH guidelines. Hence, the method holds good for the routine analysis ofhydrochlorothiazide and candesartanin various pharmaceutical industries as well as in academics.

Keywords: Hydrochlorothiazide, Candesartan,RP-HPLC, Method development, Validation


INTRODUCTION

Hydrochlorothiazide [fig. 1] is chemically 6-chloro-1,1-dioxo-3,4-dihydro-2H-1,2,4-benzothiadiazine-7-sulfonamide, belongs to the thiazide class of diuretics.The drug reduces extra fluid in thebody caused by conditions such as heart failure and kidney disease.

Fig.1: Chemical structure of hydrochlorothiazide

Candesartan [fig. 2] is chemically 1-((2'-(1H-tetrazol-5-yl)-[1,1'-biphenyl]-4-yl)methyl)-2-ethoxy-1H-benzo[d]imidazole-7-carboxylic acid, belongs to the class of angiotensin IIreceptor blockers. It is used for treating high blood pressure (hypertension).

Fig.2: Chemical structure of candesartan

A detailed literature survey revealed that there were analytical methods for the estimation of specified drugs with other combinations by ultraviolet spectroscopy (UV) [1-7], high-performance thin layer chromatography (HPTLC) [8], liquid chromatography/mass spectrometry/mass spectrometry (LCMS/MS) [9] and HPLC [10-12].There were few RP-HPLC methods for the estimation ofhydrochlorothiazide and candesartan[13-17] but the developed methods involveuse of buffers as mobile phases which decrease the life of the column and has longer retention times, leading to more analysis time. Hence, an RP-HPLC method has been developed using simple mobile phase with less runtime.

In the present work an easy, rapid, accurate and specific method was developed by using low-cost solvent water and acetonitrile for the simultaneous estimation of hydrochlorothiazide and candesartan in bulk and pharmaceutical dosage form. The developed method was validated as per validation of analytical procedures i.e.ICH guidelines: Q2[R1] Validation of Analytical Procedures: Text and Methodology.

MATERIALS AND METHODS

Chemicals and reagents

Pharmaceutical grade hydrochlorothiazide and candesartan were provided as a gift sample by Spectrum Pharma Labs, Hyderabad and the marketed formulation (Candesar-H, candesartan-16 mg, hydrochlorothiazide-12.5 mg) was purchased from local market,acetonitrile, orthophosphoric acid and HPLC grade water purchased from Merck.

Instrument

Chromatography was performed on Waters HPLC 2695 equipped with quaternary pumps with PDA detector. The chromatographic separation was performed using Silanol C18 column (4.5 x 250 mmx5 μm). The data acquisition and integration were performed using Empower 2 software.

Chromatographic conditions

The developed method used a reverse phase C18 column, Silanol C18 column (4.5 x 250 mmx5 μm), a mobile phase consisting of water (pH adjusted to 2.8 with orthophosphoricacid): acetonitrile (30:70 %v/v), flow rate of 1.0 ml/min and a detection wavelength of 210 nm using PDA detector.

Preparation of standard solutions

Standard stock solutions of hydrochlorothiazide and candesartanwere prepared by dissolving 12.5 mg of hydrochlorothiazide and 16 mg of candesartaninsufficientamount of mobile phaseand sonicated for 5 min. After that, the solution was filtered and diluted to 10 ml with mobile phase. Further dilutions in five replicates were made to get the finalconcentration of 12.5 μg/ml and 16 μg/ml forhydrochlorothiazide and candesartanrespectively. This has been treated as 100 % target concentration[18].

Preparation of sample solutions

20 tablets were weighed, crushed and quantity of powder equivalent to 12.5 mg of hydrochlorothiazide and 16 mg of candesartanwere weighed and transferred into a 10 ml volumetric flask, sufficient amount of mobile phase was added and sonicated for 5min.After that, the solution was filtered and the volume was made up to 10 ml with mobile phase. Further dilutions in five replicates were made to get the finalconcentration of 12.5 μg/ml and 16 μg/ml forhydrochlorothiazide and candesartanrespectively.

Validation of the developed method

The proposed analytical method was validated for systemsuitability, linearity and range, precision, LOD, LOQ and accuracy in accordance with ICHguidelines for analytical procedures Q2[R1] [19].

System suitability

System suitability parameters were studied to verify the systemperformance. Six replicate samples containing hydrochlorothiazide (12.5 μg/ml) and candesartan (16 μg/ml) were analyzed using thedeveloped method. Factors such as theoretical plate count, tailingfactor,percent relative standard deviation of peak area andretention time were taken into consideration for testing systemsuitability.

Linearity and range

The linearity was evaluated at six concentration levels in the rangebetween 6.25-18.75 μg/ml and 8-24μg/ml for hydrochlorothiazide and candesartanrespectively. A calibration curve was obtained by plotting concentration against corresponding peak area and linearity wasdetermined using least square regression analysis. The analyticalrange was established by the highest and lowest concentrations ofanalyteand the acceptable linearity was obtained with the specified analytical range.

Precision

The precision of the developed analytical method was carried out forsame concentration level. Six determinations were performed and were expressed in terms of percentrelative standard deviation [% RSD].

LOD and LOQ

Limit of detection and quantitation of the developed method werecalculated from the standard deviation of the y-intercept and slopeof the calibration curve of hydrochlorothiazide and candesartanusing thefollowing formula:

Limit of detection = 3.3 α/s

Limit of quantitation = 10 α/s

Where α is the standard deviation of the y-interceptand‘s’ is theslope of the calibration curve.

Accuracy

Accuracy of the method has been studied by recovery experiment by applying the standard addition method. A known quantity of drug substance corresponding to 50%, 100%, and 150% of the label claim of drug were added, to determine if there are positive or negative interferences from excipients present in the formulation. Each set of addition were repeated three times. The accuracy was expressed as the percentage of analytes recovered by the assay.

Application of validated method for assay of hydrochlorothiazide and candesartan in pharmaceutical dosage form

Tablet powder equivalent to 12.5 mg of hydrochlorothiazide and 16 mg ofcandesartan were weighed and transferred into a 10 ml volumetricflask, 7 ml of diluent was added and sonicated for 5 min. After that, the solution was filtered and thevolume was made up to the mark with diluent. From this solution, further dilution was made to get the final concentration of 12.5μg/ml of hydrochlorothiazide and 16 μg/ml of candesartan.The prepared solutions were injected into the HPLC system, chromatogramswere recorded and from the peak areas of hydrochlorothiazide and candesartan, the amount of drug present in thesample was computed.

RESULTS AND DISCUSSION

Method development

Different chromatographic conditions were tried for better separation and resolution.Silanol BDS C18 column (250×4.6 mm, 5 μm) was found satisfactory. Detection of analytes was carried out using PDA detector and 210 nm was considered satisfactory for detecting both the drugs with adequate sensitivity. A number of trails were performed with different solvents in the different ratios over a wide range of pH, with different flow rates and column temperatures.

But, either peak shape was broad or resolution was not good. Repeated trials to obtain good, sharp peaks with better retention times and efficient resolution between hydrochlorothiazide and candesartanwere performed on Silanol BDS C18 column.The runtime was good in isocratic trial with a flow rate of 1 ml/min using mobile phase containingwater (pH adjusted to 2.8 with orthophosphoric acid):acetonitrile (30:70 %v/v)and the detection wavelength of 210 nm using PDA detector gavesatisfactory results in terms of retention time, resolution, symmetry and sensitivity.A typical RP-HPLC chromatogram for simultaneous determination ofhydrochlorothiazide and candesartan from standard preparation and sample preparation was shown in fig. 3 and 4.

Fig.3: Standard chromatogram of hydrochlorothiazide and candesartan

Fig.4: Sample chromatogram of hydrochlorothiazide and candesartan

On comparison with literature, it is found that the mobile phase used by Rahul et al. [13] was phosphate buffer and acetonitrile. Quatabet al.[14] and Anand Rao et al.[15] used phosphate buffer and methanol. Whereas Mathrusri et al. [16] and Balamuralikrishna et al. [17] used 0.01 M tetrabutylammonium hydrogen sulphate: methanol and 0.02% triethylamine: acetonitrile respectively. All the methods used buffers in their mobile phases takes more preparation time and also theusage of buffers reduces the life of column. The retention time of drugs in the method developed by Rahul et al. [13] was 4.3 min and 16.1 min, Anand Rao et al.[15] was 2.1 min and 7.2 min and Mathrusri et al. [16] was 2.7 min 8.1 min thereby increasing the analysis time. In the present study, a simple mobile phase consisting of water (pH adjusted with OPA) and acetonitrile was used which elute the hydrochlorothiazide and candesartan with lower retention time.

Method validation

System suitability

Standard solutions were prepared as per the test method and injected into the chromatographic system. The system suitability parameters like theoretical plates, resolution andtailing factorwere evaluated. The system suitability parameters were summarized in table 1. All the parameters were found to be within the limits.

Table 1: Results of system suitability study

Parameters Acceptance limits Hydrochlorothiazide Candesartan
Retention time* (min) - 2.30±0.16 4.40±0.18
Resolution* NLT 2 - 14.2±0.03
Theoretical plates* NLT 2000 6493±2 10170±4
Tailing factor * NMT 2 1.23±0.02 1.05±0.03

*=results of six determinations

Linearity and range

The linearity of the test solutions for the assay method was prepared fromhydrochlorothiazide and candesartan standard stock solution at five concentration levels from 50% to 150% of assay concentration. The peak area versus concentration data was treated by least-square linear regression analysis. The results showed an excellent correlation between peak areas and concentration within the concentration range of 6.25-18.75 μg/ml and 8-24μg/ml for hydrochlorothiazide and candesartan respectively. The results of linearity of hydrochlorothiazide and candesartan weresummarized in table 2.The correlation coefficients were found to be 0.999 for both hydrochlorothiazide and candesartan, which meet the method validation acceptance criteria and hence the method was said to be linear for both the drugs. The results showed that a linear relationship between peak area and concentration of the drug in the calibration curve.

Table 2: Linearity data for hydrochlorothiazide and candesartan

% level Hydrochlorothiazide concentration (μg/ml) Hydrochlorothiazide peak area Candesartan concentration (μg/ml) Candesartan peak area
50 6.25 292718 8 626266
75 9.37 408707 12 909947
100 12.50 544937 16 1213266
125 15.62 681175 20 1516578
150 18.75 812002 24 1881404
Correlation coefficient 0.999 0.999
Slope 43031 77306

Accuracy

The accuracy of the method was determined by recovery studies by the determination of % mean recovery of both the drugs at three different levels (50 %, 100 % and 150%). At each level, threedeterminations were performed. The percentage recovery and mean percentage recovery were calculated for the drugs and presented in table 3. The observed data were within the specified range. Hence good recovery values indicate that the developed method is accurate for the determination of specified drugs.

Precision

Method precision

The precision of the method was verified by precision method studies. The sample solution was prepared at working concentration and analysis was carried out at replicates. The sample solutions of hydrochlorothiazide and candesartanwere prepared as per the test method and injected six times into the column. The results of precision were tabulated in table 4. The mean of peak area was calculated, % RSD was calculated and reported. Method precision % RSD values lower than 2% clearly assured that the developed method was found to be fairly precise and reproducible.

Table 3: Results of accuracy

Level (%) Hydrochlorothiazide Candesartan
% recovery % mean recovery % recovery % mean recovery*
50 99.59 99.78 99.89 99.87
50 99.75 99.78
50 100.01 99.96
100 99.50 99.57 100.87 101.08
100 99.99 101.63
100 99.23 100.76
150 100.10 100.39 101.45 100.64
150 101.00 100.66
150 100.08 99.83

*Average of triplicate determinations; Acceptence criteria: %recovery must be 98%-102%

Table 4: Method precision data for hydrochlorothiazide and candesartan

No. ofinjections Hydrochlorothiazide Candesartan
Rt peak area Rt peak area
Injection 1 2.286 544936 4.462 1213265
Injection 2 2.287 544933 4.485 1213266
Injection 3 2.291 544940 4.491 1213267
Injection 4 2.301 544936 4.488 1213259
Injection 5 2.302 544939 4.487 1213258
Injection 6 2.299 544938 4.480 1213262
Mean*±SD 544937±2.52 0.4 1213263±3.76 0.3
% RSD#

*The value is represented as a mean±SD of 6 observations (n=6), SD: Standard Deviation, #RSD: Relative Standard Deviation, Acceptance criteria:<2

Limit of detection and limit of quantitation

LOD and LOQ were estimated from the standard deviation of the y-intercept and slopeof the calibration curve of hydrochlorothiazide and candesartan.

The LOD andLOQ were found to be 0.410μg/mland 1.367 μg/ml for hydrochlorothiazide and0.699 μg/mland 2.330μg/ml for candesartan. The results of LOD and LOQ specified the sensitivity of the developed method.

Robustness

Robustness of the developed method was determined bydeliberately altering the experimental conditions and the system suitability parameters were evaluated. The solutions prepared as per the test method and injected at different variable conditions like flow rate (0.8, 1.2 ml/min) and wavelength (208 nm, 212 nm), system suitability parameters were compared with that of method precision.

The results were tabulated in table 5. Even though, small changes were made in the conditions there was no significant effect on tailing, plate count and retention time of specified drugs. Hence, the developed method was found to be robust.

Application of validated method for assay of hydrochlorothiazide and candesartan in pharmaceutical dosage form

The developed method was successfully implemented in the assay of hydrochlorothiazide and candesartan in pharmaceutical dosage form. Assay of hydrochlorothiazide and candesartan was found to be 99.16% and 99.60% respectively. The results of the assay were summarized in table 6.

Table 5: Results of robustness

Parameter Hydrochlorothiazide Candesartan
Plate count# Tailing* Plate count# Tailing*
Less flow rate (0.8 ml/min) 5416 1.21 9606 1.11
More flow rate (1.2 ml/min) 6692 1.19 8807 1.04
Less wavelength (208 nm) 3284 1.09 9875 1.12
More wavelength (212 nm) 6217 1.17 9294 1.09

Acceptance criteria (Limits): #Peak Asymmetry: >2000, *Tailing: <2

Table 6: Assay of hydrochlorothiazide and candesartan in pharmaceutical dosage form

Name Standard average area* Sample average area* % Assay
Hydrochlorothiazide 544837 534811 99.16
Candesartan 1213262 1208408 99.60

*Mean of six determinations

CONCLUSION

In the present study, a simple and efficient RP-HPLC method was developed for the simultaneous analysis of hydrochlorothiazide and candesartan in bulk and in pharmaceutical dosage form.The method was validated as per ICH guidelines and found to be applicable for routine quality control analysis. The results of analysis of pharmaceutical formulation by the proposed method were highly reproducible and reliable. Hence, it can be concluded that the proposed RP-HPLC method was simple, cost-effective, specific, accurate, precise, robust and rapid.

ACKNOWLEDGMENT

The authors wish to express their gratitude tothe Institute of Pharmaceutical Technology, Sri PadmavatiMahilaVisvavidyalayam (Women’s University), Tirupati and the Spectrum Pharma Lab, Hyderabad for providing necessary facilities to carry out the research work.

CONFLICT OF INTERESTS

Declared none

REFERENCES

  1. Mathrusri A, Mukthinuthalapati, Sai Phani Kumar J. Simultaneous derivative spectrophotometric determination of candesartan cilexetil and hydrochlorothiazide. Pharm Methods 2015;6:148-51.

  2. Tamboli A, Jamadar M, Khan N, Manure J, Bathe R. UV-spectrophotometric determination of telmisartan and hydrochlorothiazide in combined tablet dosage form using simultaneous equation method.Int J Adv Pharm Anal 2014;4:18-22.

  3. Chaitali T, Jyoti D, Pawar P. Simultaneous estimation and validation of losartan potassium and hydrochlorothiazide in bulk and tablet dosage form by using thedifferent spectrophotometric method. Der Pharm Chem 2014;6:24-30.

  4. Varsha R, Baheti G, Indraksha S, Dehghan M. Estimation of amlodipine besylate, valsartan and hydrochlorothiazide in bulk mixture and tablet by UV spectrophotometry. Indian J Pharm Sci 2012;74:18-23.

  5. BhadkeTejaswini K, MohiteShrinivas K, MagdumChandrakant S. Simultaneous estimation of candesartan cilexetiland hydrochlorothiazide in tablet dosage form by UV spectrophotometric method. Int J PharmTech Res 2012;4:786-90.

  6. Gopala Swamy T, Nagaraju K, Lakshmana Rao A. Simultaneous estimation of telmisartan and hydrochlorothiazide in pharmaceutical dosage form. Int J Drug Dev Res 2011;3:362-8.

  7. Patel J, Dave JB, Patel CN, Patel D. Q-analysis spectrophotometric methods for estimation of candesartan cilexetil and hydrochlorothiazide in tablet dosage form. J Chem Pharm Res 2010;2:10-4.

  8. Bipin H, Sachin B. HPTLC-densitometric analysis of candesartan cilexetil and hydrochlorothiazide in tablets. J Planar Chromatogr Mod TLC2008;21:173-6.

  9. VijayaBharathi D, Kishore H, Pankaj K, Satyanarayana V, Venkateswarlu V. LC–MS/MS method for simultaneous estimation of candesartan and hydrochlorothiazide in human plasma and its use in clinical pharmacokinetics. Bioanalysis2012;4:1195-204.

  10. Revathi R, Saravanan V, Ethiraj T, Jhansi Lakshmi M. RP-HPLC analysis for quantitation of candesartan cilexetil in solid dosage forms. Asian J Pharm Anal 2013;3:115-8.

  11. Subbarao DV, Radha Krishnan P, Suryanarayana MV, Himabindu V. A stability-indicating LC method for candesartan cilexetil. Chromatographia2007;66:499-507.

  12. Ayyakannu A, Gangadhara A, Raj A. HPLC method development and validation for simultaneous estimation of olmesartan medoxomil, hydrochlorothiazide and amlodipine besylate tablets. Pharm Lett2015;7:182-96.

  13. Rahul R, Narayanaswamy VB. Development ofa RP-HPLC method for the simultaneous estimation of candesartan cilexetil and hydrochlorothiazide in pharmaceutical dosage forms. Int JRes Pharm Chem 2015;5:452-69.

  14. Quatab S, Razzaq S, Ashfaq M, Shuja Z, Khan I. Simple and sensitive LC-UV method for simultaneous analysis of hydrochlorothiazide and candesartan cilexetil in pharmaceutical formulations. Acta Chromatogr 2007;19:119-29.

  15. Ananda Rao B, Surya Kumar T. Development and validation of an RP-HPLC method for estimation of hydrochlorothiazide and candesartan cilexetil in pharmaceutical dosage form. Int J Pharm 2013;3:166-9.

  16. Mathrusri Annapurna M, Narendra A, Ravi kumar K. Liquid chromatographic method for the simultaneous quantitative determination of candesartan cilexetil and hydrochlorothiazide in pharmaceutical dosage forms. JDrug Delivery Ther 2012;2:48-54.

  17. Balamuralikrishna K, Syamasundar B. Development and validation of ahigh performance liquid chromatographic method for the simultaneous estimation of candesartan cilexetil and hydrochlorothiazide in thecombined tablet dosage form. Pharm Chem 2010;2:231-7.

  18. Prasanthi C, AngalaParameswari S, Aruna G.Development and validation of aRP-HPLC method for metformin hydrochloride and nateglinide in bulk and combined dosage form.Int J Pharm PharmSci 2016;8:267-71.

  19. International conference on harmonization tripartite guidelineon validation of analytical procedures text and methodology: Q2 (R1); 2005.