Int J App Pharm, Vol 13, Issue 5, 2021, 115-121Original Article

DEVELOPMENT AND VALIDATION OF A STABILITY-INDICATING RELATED SUBSTANCES OF TRANDOLAPRIL BY RP-HPLC AND ITS DEGRADATION

CHALLA SUDHEER REDDY1*, B. TIRUMALESWARA RAO1

1Department of Chemistry, Vikas PG College, Vissannapeta, Andhra Pradesh, India 521215
Email: sudheer192020@gmail.com

Received: 19 May 2021, Revised and Accepted: 15 Jul 2021


ABSTRACT

Objective: A validated stability-indicating RP-HPLC method for Trandolapril was developed by separating its related impurities.

Methods: By using Waters HPLC e-2695 quaternary pump with a PDA detector of 2998 instrument, the chromatographic separation of Trandolapril and its related impurities was achieved on the column of Agilent eclipse C18 (150x4.6 mm, 3.5 µ) using gradient elution with a buffer containing 0.1percent formic acid and acetonitrile as a mobile phase with a flow rate of 1 ml/min at ambient temperature. A detector wavelength of 213 nm utilizing the PDA detector was given in the instrumental settings. The linearity was studied between the concentration range of 4-60 µg/ml of Trandolapril and 0.5-7.5 µg/ml of imp-E, imp-A, imp-B and 0.7-10.5 µg/ml of imp-D were injected with a run time of 17 min. Validation of the proposed method was carried out according to an International Conference on Harmonization (ICH) guidelines.

Results: LOD and LOQ for the Trandolapril and its impurities were established with respect to test concentration. The plotted calibration curves were linear with a regression coefficient of R2>0.999, which indicates that the linearity was within the limit. As a part of method validation, the parameters like specificity, linearity, accuracy, ruggedness, robustness were determined and the results were found to be within the allowable limit.

Conclusion: The method developed was found to be applicable to routine analysis and to be used for the measurement of active pharmaceutical ingredients (i. e, Trandolapril and its related impurities). Since there is no HPLC method reported in the literature for the estimation of Trandolapril and its related impurities; there is a need to develop quantitative methods under different conditions to achieve improvement in specificity, selectivity etc.

Keywords: Trandolapril, Related impurities, HPLC, Development, Validation


INTRODUCTION

Trandolapril is an ACE inhibitor [1, 2] used to treat high blood pressure [3, 4]. It may also be used to treat other conditions. Side effects reported for trandolapril include nausea, vomiting, diarrhea, headache, dry cough, dizziness or lightheadedness when sitting up or standing, hypotension [5], or fatigue [6]. Patients also on diuretics [7] may experience an excessive reduction of blood pressure after initiation of therapy with trandolapril. It can reduce potassium loss caused by thiazide [8] diuretics and increase serum potassium when used alone. Therefore, hyperkalemia [9, 10] is a possible risk. Increased serum lithium levels can occur in patients who are also on lithium. Trandolapril is teratogenic [11] (US: pregnancy category D) and can cause birth defects and even death of the developing fetus. The highest risk to the fetus is during the second and third trimesters. When pregnancy is detected, trandolapril should be discontinued as soon as possible. Trandolapril should not be administered to nursing mothers. Combination therapy with paricalcitol and trandolapril has been found to reduce fibrosis [12] in obstructive uropathy [13, 14]. Trandolapril is a prodrug that is de-esterified to trandolaprilat. It is believed to exert its antihypertensive effect through the renin-angiotensin-aldosterone system [15, 16]. Trandolapril has a half-life of about 6 h, and trandolaprilat has a half-life of about 10 h. Trandolaprilat has about eight times the activity of its parent drug. About one-third of trandolapril and its metabolites [17] are excreted in the urine [18, 19], and about two-thirds of trandolapril and its metabolites are excreted in the feces. Serum protein binding of trandolapril is about 80%. Trandolapril acts by competitive inhibition of  the angiotensin-converting enzyme (ACE), a key enzyme in the renin-angiotensin system, which plays an important role in regulating blood pressure. So, we developed a method for the estimation of Trandolapril by using RP-HPLC.

A B C

D E

Fig. 1: Chemical structures of (A) Trandolapril (B) Impurity-A (C) Impurity-B (D) Impurity-D (E) Impurity-E

MATERIALS AND METHODS

Chemicals

Acetonitrile, HPLC-grade orthophosphoric acid, water were purchased from Merck India Ltd, Mumbai, India. Candila health care ltd, Ahmedabad, India provided the reference criteria for Trandolapril and its related impurities.

The ınstrumentation

Waters alliance liquid chromatography (model e-2695) monitored with empower 2.0 data handling system and a detector of photodiode array (model 2998) [20] was used for this study.

Preparation of mobile phase-A: 1 ml of formic acid is dissolved in 1 lt of HPLC grade water and filter through 0.45 µ filter paper.

Mobile Phase-B: Acetonitrile

Optimization of mobile phase

Different trails have been done, different buffers and different mobile phases were used to develop the method. In all trails, peaks are not separated properly. Finally, for the proposed method, all the peaks are separated and the entire suitability conditions are within the limit.

Table 1: Gradient program

Time (min) Mobile phase-a Mobile phase-b
0.00 80 20
5 50 50
7 20 80
10 20 80
12 80 20
17 80 20

Till today there are no HPLC methods reported in the literature, So, it has more interested to develop a novel and reliable HPLC strategy for the establishment of Trandolapril and its related impurities.

Chromatographic conditions

The HPLC analysis was performed on a reverse-phase HPLC system with isocratic elution mode using a mobile phase of acetonitrile and 0.1% formic acid and Agilent eclipse C18 (150x4.6 mm, 3.5 μ) column with a flow rate of 1 ml/min.

Diluent

Mobile phase was used as a diluent.

Validation procedure

The analytical parameters [21-25] such as system suitability, precision, specificity, accuracy, linearity, robustness, LOD, LOQ, forced degradation and stability were validated according to ICH Q2 (R1) guidelines.

Standard stock solution

Weigh accurately 40 mg of Trandolapril and transferred into a 100 ml volumetric flask; add 70 ml of diluent sonicated for 10 min to dissolve the contents make up to the mark with diluent.

Sample stock solution

Transfer 740 mg (equivalent to 40 mg of Trandolapril and each tablet contains 4 mg of Trandolapril) of sample into a 100 ml volumetric flask diluted to volume with diluent. Filter through 0.45µ nylon syringe filter.

Impurity standard stock solution

Weigh accurately 5 mg each of imp-E, imp-A, imp-B and 7 mg of imp-D into a 100 ml volumetric flask. Add 70 ml of diluent, sonicated to dissolve and make up.

Spiked standard solution

Transfer 5 ml of standard stock into a 50 ml volumetric flask, add 40 ml of diluent and also add 5 ml of impurity standard stock solution and makeup to the mark with diluent. Filter through 0.45µ syringe filter.

Spiked sample solution

Transfer 5 ml of sample stock into a 50 ml volumetric flask, add 40 ml of diluent and also add 5 ml of impurity standard stock solution and makeup to the mark with diluent. Filter through 0.45µ syringe filter.

RESULTS AND DISCUSSION

The main analytical challenge during the development of a new method was to separate active Pharma ingredients. In order to provide good performance, the chromatographic conditions were optimized.

Method validation

The optimized RP-HPLC validated method according to ICH guidelines in terms of system suitability, linearity, accuracy, precision and robustness.

System suitability

Device suitability was performed by injecting a spiked standard solution containing 40 µg/ml of Trandolapril, 5 µg/ml each of imp-E, imp-A, imp-B and 7 µg/ml of imp-D in six replicates. The results show that the machine fitness parameter is within the limit provided by ICH [26]. The results were shown below.

Table 2: Results of system suitability

System suitability parameter Acceptance criteria Drug name trandolapril
USP Plate count NLT 2000 51863
USP Tailing NMT 2.0 1.04
USP Resolution NLT 2.0 22.69
% RSD NMT 2.0 0.82
Retention Time NLT 2.0 9.894

Fig. 2: Chromatogram of standard

Specificity

In this test method, placebo, sample and standard solutions were analyzed individually to examine the interference. The below fig. shows that the active ingredient and its related substances were well separated from blank and their excipients and there was no interference of placebo with the principal peak. Hence the method is specific.

Fig. 3: Chromatogram of blank

Linearity

Linearity was calculated by plotting a calibration curve of the peak area against its respective concentration, linearity was determined. From this calibration curve, it was noticed that the curve was linear between the range of 4-60µg/ml of Trandolapril and 0.5-7.5µg/ml each of imp-E, imp-A, imp-B and 0.7-10.5 µg/ml of imp-D. The regression equations for calibration curve was Y=147215.89x+159148.18 (R2=0.9996) for Trandolapril and Y= 9503.01x+1281.27 (R2=0.9993) for imp-E and Y=7488.75x+397.6 (R2=0.9998) for imp-D and Y=14689.54x+1046.22 (R2=0.9993) for imp-A and Y=14478.22x+389.98 (R2=0.9995) for imp-B respectively.

Table 3: Linearity results of trandolapril and its impurities

A

Linearity Imp-E Imp-D Imp-A
Conc. (µg/ml) Area Conc. (µg/ml) Area Conc. (µg/ml) Area
Linearity-1 0.50 6035 0.70 6484 0.50 9046
Linearity-2 1.25 13458 1.75 13311 1.25 19198
Linearity-3 2.50 25824 3.50 25892 2.50 37293
Linearity-4 5.00 50245 7.00 53572 5.00 77891
Linearity-5 6.25 60822 8.75 65900 6.25 91630
Linearity-6 7.50 71154 10.50 78762 7.50 110125
Slope 9503.01 7488.75 14689.54
Intercept 1281.27 397.6 1046.22
CC 0.9993 0.9998 0.9993

B

Linearity Trandolapril Imp-B
Conc. (µg/ml) Area Conc. (µg/ml) Area
Linearity-1 4.00 781724 0.50 8168
Linearity-2 10.00 1702135 1.25 18324
Linearity-3 20.00 3226387 2.50 36182
Linearity-4 40.00 6049599 5.00 74780
Linearity-5 50.00 7596589 6.25 88636
Linearity-6 60.00 8885327 7.50 109639
Slope 147215.89 14478.22
Intercept 159148.18 389.98
CC 0.9996 0.9995

Accuracy

The accuracy of the system was achieved by measuring the recovery experiments at three stages (50 percent, 100 percent and 150 percent). APIs with concentrations of 20, 40 and 60µg/ml of Trandolapril and 125, 250 and 375µg/ml of Permethrin were prepared. For each spike stage, the test solution was injected three times and the test was performed according to the test process. The recovery results were similar to 100% and also the RSD values were less than±2%. The percentage recovery, mean and relative standard deviations were determined. Recovery values shown within the desired range were correct. The results are summarized below. Accuracy findings have been shown in table 4.

Precision

The precision of the analytical technique is the degree of proximity of the sequence of measurements obtained from multiple homogeneous mixture samplings. The accuracy of the process of the drugs was calculated by injection of six individual determinations of Trandolapril and its related substances. Method precision results were shown in table 4 and sample chromatogram was shown in fig. 5.

Intraday precision

Six replicates of a sample solution containing Trandolapril and is related substances were analysed on the same day. Peak areas were calculated, which were used to calculate mean, SD and %RSD values.

Trandolapril Imp-E

Imp-D Imp-A

Imp-B

Fig. 4: Calibration plots of trandolapril and its related impurities

Table 4: Results of accuracy

S. No. % Level Trandolapril % recovery
1 50 99.82
2 100 100.05
3 150 99.76
Mean 99.88
Std Dev 0.153

mean±SD (n=3)

Table 5: Intraday precision results of allantoin and permethrin

Sample No. % of related substances
Spiked impurities Total impurities % Purity (100-total impurities)
1 1.15 0.58 99.42
2 1.16 0.62 99.38
3 1.12 0.64 99.36
4 1.24 0.69 99.31
5 1.22 0.63 99.37
6 1.28 0.61 99.39
Average 1.20 0.63 99.37
% RSD 5.12 5.82 0.04

mean±SD (n=6)

Fig. 5: Chromatogram of sample

Intermediate precision

Six replicates of the sample solution were analyzed by different researchers and different tools were checked on separate days. The peak regions used to assess the average percent of RSD values have been determined. The findings are shown in the table below.

Interday precision

Six replicates of a sample solution containing Trandolapril and its related substances were analysed on a different day. Peak areas were calculated, which were used to calculate mean, SD and %RSD values. The present method was found to be precise as the RSD values were less than 2% and also, the percentage assay values were close to be 100%. The results are given in table 6 [27].

LOD and LOQ

LOD and LOQ were determined separately using the calibration curve technique. The LOD and LOQ of the compound were measured using the developed RP-HPLC method by injecting lower and lower concentrations of the standard solution. The LOD and LOQ concentrations and their s/n values of Allantoin and Permethrin were represented in the following table. This method is validated as per the ICH guidelines [28, 29].

A

B

Fig. 6: Chromatogram of (A) LOD and (B) LOQ

Robustness

The conditions of the experiment were designed to measure the robustness of the intentionally changed conditions such as flow rate, mobile phase in organic percentage in all these varied conditions. Robustness results for Trandolapril and its impurities were found to be within the limit and results were tabulated in table 8 [30].

Table 6: Inter-day precision results

Sample No. % of related substances
Spiked impurities Total impurities % Purity (100-Total impurities)
1 1.19 0.68 99.32
2 1.25 0.66 99.34
3 1.22 0.59 99.41
4 1.17 0.72 99.28
5 1.13 0.55 99.45
6 1.20 0.62 99.38
Average 1.19 0.64 99.36
% RSD 3.46 9.77 0.06

mean±SD (n=6)

Table 7: Results of LOD and LOQ

Name LOD conc. (µg/ml) S/N LOQ conc. (µg/ml) S/N
Trandolapril 0.05 8 0.165 37
Imp-E 0.0063 4 0.0218 33
Imp-D 0.0088 5 0.029 334
Imp-A 0.0062 4 0.0218 33
Imp-B 0.0061 4 0.0218 33

Table 8: Robustness results of allantoin and permethrin

Parameter name % RSD
Trandolapril
Flow rate (0.8 ml/min) 0.35
Flow rate (1.2 ml/min) 0.98
Org Plus (66:34) (+10%) 1.01
Org Minus (54:46) (-10%) 0.86

Table 9: Stability results of trandolapril

Stability Trandolapril
% Purity % deviation
Initial 99.98 0.01
6 h 99.36 0.24
12 h 98.86 1.14
18 h 98.53 1.47
24 h 98.21 1.79

Stability

Normal solution was kept at room temperature and 2-8 ° C for up to 24 h. These solutions were then pumped into the system and the percent deviation from the initial to 24 h was measured [31]. No major variations were found and verified that the solutions were stable up to 24 h percentage of the assay was not quite 2%. There is no effect in storage conditions for Trandolapril and its related impurities. Stability results were tabulated in table 9.

Degradation studies

Trandolapril and its related substances were subjected to various conditions of forced degradation [32, 33] in order to induce partial degradation of the compound. Forced degradation experiments have been performed to establish that the process is acceptable for degradation materials [34, 35]. In addition, the studies include information on the condition under which the drug is unstable, such that the steps are also taken during formulation to prevent possible instabilities [36].

Acid degradation

Acid degradation was done by using 1N HCl and 15.3% Trandolapril degradation was observed.

Alkali degradation

Alkali degradation was done by using 1N NaOH and 15.1% of Trandolapril degradation was observed.

Peroxide degradation

Peroxide degradation was done by using 30% peroxide and 14.7% of Trandolapril degradation was observed.

Reduction degradation

Reduction degradation was done by using 30% sodium bisulphate solution and 12.4% Trandolapril degradation was observed.

Thermal degradation

Thermal degradation was done at 105 °C and 11.9% of Trandolapril degradation was observed.

Hydrolysis degradation

Hydrolysis degradation was done by using HPLC water and 10.7% Trandolapril degradation was observed.

Table 10: Forced degradation results of allantoin and permethrin

Degradation condition Trandolapril
% purity
Acid deg 15.3
Alkali deg 15.1
Peroxide deg 14.7
Reduction deg 12.4
Thermal deg 11.9
Hydrolysis deg 10.7

CONCLUSION

The developed method gave good results between Torsemide and its four impurities with a run time of 17 min, high efficiency and complies with modified SST specifications of USP. The utilization of Agilent eclipse C18 column within the present work has shown better elution of analytes with good resolution, improved plate count and tailing. Therefore the C18 columns are often wont to achieve high specificity in a shorter time of study of Trandolapril as per ICH Q 3A (R2) guidelines. The proposed method was found to be simple, precise, accurate, linear, robust and rapid for simultaneous determination and quantification of Trandolapril and its impurities. The sample recovery was in good agreement with their respective label claims suggested non-interference within the estimation. Hence, the technique is often easily and conveniently adopted for routine analysis of Trandolapril in the combined dosage form.

ACKNOWLEDGEMENT

I thankful to Shree Icon Pharmaceutical Laboratories for providing laboratory facilities to finish this research work.

FUNDING

Nil

AUTHORS CONTRIBUTIONS

All authors have contributed equally.

CONFLICTS OF INTERESTS

Author declares that there have been no conflicts of interest.

REFERENCES

  1. Van Vark LC, Bertrand M, Akkerhuis KM, Brugts JJ, Fox K. Angiotensin-converting enzyme inhibitors reduce mortality in hypertension: a meta-analysis of randomized clinical trials of renin-angiotensin-aldosterone system inhibitors involving 158,998 patients. Eur Heart J 2012;33:2088–97.

  2. Augenstein, Vedra A, Heniford B Todd, Sing, Ronald F. ıntestinal angioedema ınduced by angiotensin-converting enzyme ınhibitors: an underrecognized cause of abdominal pain. J Am Osteopath Assoc 2013;113:221–3.

  3. Musini VM, Gueyffier F, Puil L, Salzwedel DM, Wright JM. Pharmacotherapy for hypertension in adults aged 18 to 59 y. Cochrane Database Syst Rev 2017;8:CD008276.

  4. Arguedas, Jose Agustin, Leiva, Viriam, Wright, James M. Blood pressure targets in adults with hypertension. Cochrane Database Syst Rev 2020;12:CD004349.

  5. Freeman Roy, Abuzinadah, Ahmad R, Gibbons, Christopher, Jones, et al. Orthostatic hypotension: JACC state-of-the-art review. J Am Coll Cardiol 2018;72:1294–309.

  6. Shen J, Barbera J, Shapiro CM. Distinguishing sleepiness and fatigue: focus on definition and measurement. Sleep Med Rev 2006;10:63–76.

  7. Rejnmark L, Vestergaard P, Pedersen AR, Heickendorff L, Andreasen F, Mosekilde L. Dose-effect relations of loop-and thiazide-diuretics on calcium homeostasis: a randomized, double-blinded latin-square multiple cross-over study in postmenopausal osteopenic women. Eur J Clin Invest 2003;33:41–50.

  8. Magaldi, Antonio J. New insights into the paradoxical effect of thiazides in diabetes insipidus therapy. Nephrol Dial Transplant 2000;15:1903–5.

  9. Smellie, W Stuart A. Spurious hyperkalemia Br Med J 2007;334:693–5.

  10. Kovesdy CP. Updates in hyperkalemia: outcomes and therapeutic strategies. Rev Endocr Metab Disord 2016;18:41–7.

  11. Bellinger DC. Teratogen update: lead and pregnancy. Birth Defects Res (Part A) Clin Mol Teratol 2005;73:409–20.

  12. Chute, Michael, Aujla, Preetinder, Jana. The non-fibrillar side of fibrosis: contribution of the basement membrane, proteoglycans, and glycoproteins to myocardial fibrosis. Cardiovasc Dev Dis 2019;6:35.

  13. Woodford HJ, George J. Diagnosis and management of urinary infections in older people. Clin Med 2011;11:80–3.

  14. Simerville, Jeff A. Urinalysis: a comprehensive review. Am Fam Physician 2005;71:1153–62. 

  15. Kumar R, Singh VP, Baker KM. The intracellular renin-angiotensin system in the heart. Curr Hypertens Rep 2009;11:104–10.

  16. Brown MJ. Success and failure of vaccines against renin-angiotensin system components. Nat Rev Cardiol 2009;6:639–47.

  17. Obach R Scott, Esbenshade, Timothy A. Pharmacologically active drug metabolites: ımpact on drug discovery and pharmacotherapy. Pharmacol Rev 2013;65:578–640.

  18. Yang PJ, Pham J, Choo J, Hu DL. Duration of urination does not change with body size. Proc Natl Acad Sci USA 2014;111:11932–7. 

  19. Rose C, Parker A, Jefferson B, Cartmell E. the characterization of feces and urine: a review of the literature to ınform advanced treatment technology. Crit Rev Environ Sci Technol 2015;45:1827–79. 

  20. Cijo M Xavier, Kanakapura Basavaiah. RP-UPLC development and validation of metformin hydrochloride in pure drug and pharmaceutical formulations. World J Pharm Pharm Sci 2015;4:1649-68. 

  21. Shalini K, Ilango K. Development, evaluation and RP-HPLC method for simultaneous estimation of quercetin, ellagic acid and kaempferol in a polyherbal formulation. Int J Appl Pharm 2021;13:183-92.

  22. Sri Girija K, Bikshal Babu K, Venkateswara Rao A. A new high-performance liquid chromatography method for the separation and simultaneous quantification of eptifibatide and its impurities in pharmaceutical injection formulation. Int J Appl Pharm 2021;13:165-72.

  23. VLN Balaji Gupta T, Venkateswara Rao B, Kishore Babu B. RP-HPLC (stability-indicating) based assay method for the simultaneous estimation of doravirine, tenofovir disoproxil fumarate and lamivudine. Int J Appl Pharm 2021;13:153-9.

  24. Murali Krishnam Raju P, Venkata Narayana B, Shyamala P, Srinivasu K, HSN Raju D. A validated RP-HPLC method for impurity profiling of Sodium nitroprusside in an injection dosage form. Int J Appl Pharm 2021;13:160-9.

  25. Sanathoiba Singha L, Srinivasa Rao T. Development and validation of an RP-HPLC method for the determination of ulipristal acetate in the pharmaceutical dosage form. Asian J Pharm Clin Res 2021;14:83-9.

  26. Asha Eluru, Surendra Babu K. A study of method development, validation and forced degradation for simultaneous quantification of povidone ıodine and ornidazole in bulk and pharmaceutical dosage form by using RP-HPLC. IJPSR 2021;12:1217-22.

  27. Malathi S, Devakumar D. Development and validation of RP-HPLC method for the estimation of escitalopram oxalate and flupentixol dihydrochloride in combined dosage form and plasma. Int J Pharm Pharm Sci 2021;13:61-6.

  28. International Conference on Harmonization (ICH). Harmonized tripartite guideline validation of analytical procedures: Text and methodology Q2 (R1). Geneva: IFPMA, Switzerland; 2005.

  29. Ravichandran V, Shalini S, Sundaram KM, Rajak H. Validation of analytical methods-strategies and importance. Int J Pharm Pharm Sci 2010;2:18-22.

  30. Raviteja G, Rambabu K. A study of development and validation of a method for simultaneous estimation of cidofovir and famciclovir using RP-HPLC. Int J Res Pharm Sci 2020;11:7878-84.

  31. Vijayakumari M, Balasekhar Reddy Ch. Stability indicating validated HPLC method for the determination of zanubrutinib in bulk and pharmaceutical dosage form. Asian J Pharm Clin Res 2020;13:159-62.

  32. Mohinish Sahai, N Devanna. Validated stability-indicating HPLC approach for quantifying tricholine citrate and cyproheptadine simultaneously in syrup forms. Int J Appl Pharm 2021;13:207-13.

  33. Abdul Raziq, Syed Umer Jan. Relative comparison of stability and degradation of methylcobalamin tablets of different brands at different storage settings. Int J Appl Pharm 2021;13:171-5.

  34. Rajakumari R, Sreenivasa Rao S. Stress degradation studies and development of a validated RP-HPLC method for determination of tiagabine in the presence of its degradation products. Int J Pharm Pharm Sci 2016;8:230-6.

  35. Charu Pandya P, Sadhana Rajput J. Development and validation of stability-indicating method RP-HPLC method of acotiamide. Int J Pharm Pharm Sci 2018;10:1-8.

  36. Birva Athavia A, Zarna Dedania R. Stability indicating HPLC method for determination of vilazodone hydrochloride. Int J Curr Pharm Res 2017;9:123-9.