Department of Chemistry, Acharya Nagarjuna University, Guntur-522510, Andhra Pradesh, India
*Corresponding author: Rudraraju Ramesh Raju; *Email: rrraju1@gmail.com
Received: 14 Mar 2023, Revised and Accepted: 27 May 2023
ABSTRACT
Objective: This investigation entitles the development and authentication of a rapid, selective and explicit RP-HPLC technique to assay tegafur (TGR), gimeracil (GMR), and oteracil (OTR) simultaneously in bulk and formulations of capsule type.
Methods: The separation, detection and assessment of TGR, GMR and OTR were achieved using a C18 Agilent Zorbax (25 cm; 4.6 mm; 5 µm particle dimension) reverse phase column. The acetonitrile (40% by volume) and 0.1% triethylamine in distilled water (pH 2.5, 60% by volume) was utilized as mobile phase. The validation of the method and degradation study was performed as per the strategy given by ICH.
Results: The retention periods in Agilent Zorbax column for OTR, TGR, and GMR were 2.458 min, 7.236 min and 8.629 min, respectively. Linearity was seen in the concentration series of 5.0-30.0 µg/ml (TGR), 1.45-8.70 µg/ml (GMR), and 3.95-23.70 µg/ml (OTR). The regression coefficient was greater than 0.999. The LOQ values were 0.606 µg/ml (TGR), 0.175 µg/ml (GMR), and 0.478 µg/ml (OTR). The percent comparative standard deviation (exactness) values were bestowed to be 0.243%-0.676%, 0.293%-1.894% and 0.269%-0.615% for TGR, GMR and OTR, respectively. The percent recoveries (accuracy) were in the range of 100.044%-100.493 for TGR, 99.730%-100.335% for GMR and 100.064%-100.543% for OTR.
Conclusion: The research results of the degradation investigation proved the technique's specificity as well as stability indicating feature. The process could be used for routine evaluation of OTR, TGR, and GMR in formulations of capsule type.
Keywords: Tegafur, Tegonat capsule, Gimeracil, HPLC, Oteracil, Stability
© 2023 The Authors. Published by Innovare Academic Sciences Pvt Ltd. This is an open access article under the CC BY license (https://creativecommons.org/licenses/by/4.0/)
DOI: https://dx.doi.org/10.22159/ijap.2023v15i4.47831. Journal homepage: https://innovareacademics.in/journals/index.php/ijap
Gastric cancer is the fourth largely prevalent cancer in the overall world and is the second leading cause of cancer-related mortality globally [1, 2]. Once the gastric cancer condition has advanced, symptoms start to appear. The incidence probability of gastric cancer varies greatly by region. More than half of all new cases are figured-out in underdeveloped nations. The high-jeopardy regions included East Asia, Central America, Eastern Europe, and South America. The low-jeopardy regions included North America, North Africa, New Zealand Southern Asia, East Africa, and Australia [3].
The clinical therapy of metastatic gastric cancer needed the use of innovative therapeutic medications with greater effectiveness. To manage severe gastric cancer, a set of three pharmaceutical substances-tegafur (TGR), gimeracil (GMR), and oteracil (OTR) was authorized [4, 5]. After being taken orally, TGR progressively transforms into the DNA synthesis-inhibiting compound 5-fluorouracil. Additionally, RNA function is disturbed by the amalgamation of 5-fluorouracil into RNA [6]. GMR, a dihydropyrimidine dehydrogenase blocker, stops 5-fluorouracil against being degraded. In order to lessen toxicity to the healthy gastrointestinal mucosa, OTR, an orotate phosphoribosyltransferase blocker, is designed to limit the action of 5-fluorouracil in the stomach [7, 8]. The chemical architectures of TGR, GMR, and OTR appear in fig. 1.
Fig. 1: Chemical architectures of TGR, GMR, and OTR
Pharmaceutical assessment in drug development largely focuses on ways to discover as well as quantify prospective novel drug candidates, assess purity, recognize bye-products and degradation components in compatibility, including stability studies, and to establish the destiny of the drug ingredient in humans [9, 10]. These kinds of difficult activities need advanced approaches, specialized tools, and procedures. Liquid chromatography, specifically high-performance liquid chromatography, has recently emerged as the paramount analytical method employed in both normal quality control agencies and drug development [11, 12]. The fundamental benefit of liquid chromatography is its extensive use in academics, education, and routine development over decades, leading to generally recognized and extensively implemented ways of method innovation, method improvement, and problem-solving. Given that liquid chromatography methodologies are the preferred analytical strategy, many researchers have worked to develop a variety of liquid chromatography techniques for the simultaneous estimate of various active elements in multi-constituent medications [13-21].
The quantification of TGR in combination with uracil and 5-fluorouracil by UPLC-MS/MS [22]; TGR in combination with uracil by LC-MS/MS andLC-UV [23]; TGR alone by competitive ELISA [24]; and TGR in combination with 5-fluorouracil, GMR and oxonic acid by LC–MS/MS [25] were reported. The aforementioned procedures were used only on human plasma [22-25]. The simultaneous estimate of TGR, GMR, and OTR in the pharmaceutical capsule preparation hasn't yet been recorded using any approach.
This study intends to set up an efficient, simple, accurate, and reproducible RP-HPLC approach for the quantitative investigation of TGR, GMR, and OTR. Another objective of the study is to evaluate the validity of the devised RP-HPLC technique and by what means this designed new RP-HPLC approach can identify the stabilities of these three molecules (TGR, GMR, and OTR) under different ICH-mentioned stress situations.
Drug and capsule formulations
The active pharmaceutical ingredients TGR, GMR and OTR were supplied by “Natco Pharma Limited”, India. Tegonat capsules from Natco Pharma Limited, India, were bought at a local market and were branded as having amounts of TGR, GMR, and OTR of 20 mg, 5.8 mg, and 15.8 mg, respectively.
Chemicals
Acetonitrile, triethyl amine, NaOH, HPLC grade water, orthophosphoric acid, HCl, and peroxide were supplied by “Merck Life Science Private Limited” India.
Instruments
The analytes (TGR, GMR and OTR) were investigated employing HPLC system (“Water-2695/Alliance”) couple with PDA detector (“Water-2998/Alliance”). The photostability compartment (“Newtronic NLPS4SI”) was employed to investigate TGR, GMR and OTR photodegradation. Hot air oven (“Sun life”) was employed to investigate TGR, GMR and OTR thermal degradation. The pH of the analytes solutions and the mobile phase was checked with pH meter (“Eutech Instruments-ECPH70042GS”). “Unichrome sonicator” was used to sonicate TGR, GMR and OTR samples.
TGR, GMR and OTR solutions
The active pharmaceutical ingredients TGR (200 mg), GMR (58 mg) and OTR (158 mg) were dissolved in ultrapure acetonitrile to create stock solutions of TGR (2000 µg/ml), GMR (580 µg/ml), and OTR (1580 µg/ml), which were then completed in 100 ml calibrated measurement flasks. Acetonitrile was used to further dilute aliquots of the produced stock TGR, GMR and OTR solutions, resulting in a final volume of 100 ml. These diluted TGR, GMR and OTR solutions were used as working samples having 20 µg/ml, 5.8 µg/ml and 15.8 µg/ml quantities of TGR, GMR and OTR, respectively.
The analytes (TGR, GMR and OTR) were separated employing C18 Agilent Zorbax type column (dimensions: 250 mm × 4.6 mm, particle dimension: 5 µm) arranged with ambient temperature using isocratic kind elution, with the mobile phase being a mixture of acetonitrile (40% by volume) and 0.1% triethylamine in distilled water (pH 2.5, 60% by volume) with 1.0 ml per min flow scale. For the quantitative study, the areas of TGR, GMR, and OTR obtained at 282 nm utilising PDA sensor module were employed. For analysis, 10 µl of TGR, GMR, and OTR samples were infused.
Calibration curves of TGR, GMR and OTR
Standard solutions containing 5.0–30.0 µg/ml TGR, 1.45–8.70 µg/ml GMR and 3.95–23.70 µg/ml OTR, were made ready separately in ultra-pure acetonitrile. Measured the peak areas of 5.0–30.0 µg/ml TGR, 1.45–8.70 µg/ml GMR and 3.95–23.70 µg/ml OTR solutions at 282 nm employing the suggested HPLC approach. The peak areas of TGR, GMR, and OTR are exactly proportionate to the corresponding concentrations of each; therefore, calibration graphs for TGR, GMR, and OTR were built and regression equations for TGR, GMR, and OTR were calculated.
Analysis of TGR, GMR and OTR in capsule formulation
One Tegonat capsule's worth of material (20 mg TGR, 5.8 mg GMR, and 15.8 mg OTR) was precisely placed into a calibrated flask of volume 100 ml, and 25.0 ml of acetonitrile was added while the mixture was continuously stirred in a sonicator for almost 30 min. The flask is allowed to cool before being filled with acetonitrile to finish the volume and filtrated with 0.45 µ syringe nylon filter. Acetonitrile was used to further dilute an aliquot (1 ml) of the produced stock Tegonat capsule solutions, resulting in a final volume of 10 ml. These diluted Tegonat capsule solutions were used as working Tegonat capsule samples having 20 µg/ml, 5.8 µg/ml and 15.8 µg/ml quantities of TGR, GMR and OTR, respectively. The suggested approach was then applied to analyse a pharmaceutical Tegonat preparation, and the concentrations of TGR, GMR, and OTR in Tegonat capsule were determined using the associated regression equations.
Stress investigation TGR, GMR and OTR
The tegonat capsule material underwent stress examinations in conformity with ICH norms [26].
Acid degradation
One Tegonat capsule's worth of material (20 mg TGR, 5.8 mg GMR, and 15.8 mg OTR) was appended with 5 ml of acid (1N strength HCl) in a volumetric flask (100 ml capacity) and boiled in a water bath thermostated at 60 °C for 30.0 min. Then, once it had reached room temperature, 5 ml of alkali (1N strength NaOH) was added. Finally, 70 ml of acetonitrile was appended, and the mixture was ultrasonically processed for 20 min before being made up to the required level (100 ml) with acetonitrile.
Alkali degradation
One Tegonat capsule's worth of material (20 mg TGR, 5.8 mg GMR, and 15.8 mg OTR) was subjected to alkali (1 N strength NaOH, 5 ml) and a temperature of 60 °C in a thermostated water bath for 30.0 min. Once the contents had reached room temperature, 5 ml of acid (1N strength HCl) was added. Finally, 70 ml of acetonitrile was appended, and the mixture was ultrasonically processed for 20 min before being made up to the required level (100 ml) with acetonitrile.
Oxidative degradation
This was carried out on one Tegonat capsule's worth of material (20 mg TGR, 5.8 mg GMR, and 15.8 mg OTR) under an oxidant (30% peroxide, 5 ml) at 60 °C in a thermostated water bath for 30.0 min. After bringing the contents to ambient temperature, 70 ml of acetonitrile was appended, and the mixture was ultrasonically processed for 20 min before being made up to the required level (100 ml) with acetonitrile.
Hydrolytic degradation
This was made on one Tegonat capsule's worth of material (20 mg TGR, 5.8 mg GMR, and 15.8 mg OTR) with water (Milli Q, 5 ml) at 60 °C in a thermostated water bath for 30.0 min. After adding 70 ml of acetonitrile, the mixture was ultrasonically treated for 20 min before being topped off with acetonitrile to the necessary level (100 ml). The prepared sample was then filtered using a 0.45 syringe nylon filter.
Photodegradation
A Tegonat capsule comprising 20 mg TGR, 5.8 mg GMR, and 15.8 mg OTR was used during these experiments. The substance (Tegonat capsule powder) was treated in a photostability chamber over 72 lux h. The stressed Tegonat capsule sample was brought to ambient thermal conditions and precisely placed into a calibrated flask of volume 100 ml, and twenty-five ml of acetonitrile was added while the mixture was continuously stirred in a sonicator for almost 20 min. The flask is allowed to cool before being filled with acetonitrile to finish the volume and filtrated with 0.45 µ syringe nylon filter.
Thermal degradation
A Tegonat capsule comprising 20 mg TGR, 5.8 mg GMR, and 15.8 mg OTR was used during these experiments. The substance (Tegonat capsule powder) was treated at 105 °C in the thermostated oven for three hr. The sample of the stressed Tegonat capsule was cooled to ambient temperature and carefully placed in a calibrated flask with a capacity of 100 ml. An 25.0 ml of acetonitrile was further added, and the mixture was constantly agitated in ultrasonication for almost 20 min. After allowing the flask to cool, the remaining volume is replenished with acetonitrile. The prepared sample was then filtered using a 0.45 syringe nylon filter.
In each instance, an appropriate amount of the stressed Tegonat capsule (1 ml) was diluted to 10 ml with acetonitrile before being analysed using the indicated HPLC methodology.
RESULTS
To separate TGR, GMR and OTR well within a manageable run time, the chromatographic conditions first needed to be optimised. The isocratic mode was adopted for the HPLC assessments of TGR, GMR, and OTR.
Inertsil ODS (dimensions: 250 mm × 4.6 mm, particle dimension: 5 µm), Waters symmetry C18 (dimensions: 250 mm × 4.6 mm, particle dimension: 5 µm), Aligent C18 zorbax (dimensions: 250 mm × 4.6 mm, particle dimension: 5 µm) were put on trails as stationary phase. With Aligent C18 zorbax (dimensions: 250 mm × 4.6 mm, particle dimension: 5 µm), smoother symmetrical peaks, excellent system efficiency are attained. Improved separation selectivity for TGR, GMR and OTR are also attained.
In the proposed investigation, the mobile phase was made up of an eluent that included varied amounts of 0.1% phosphoric acid buffer (pH 2.5) and 0.1% triethyl amine buffer (pH 2.5) with acetonitrile. Adequate retention, better system efficacy, symmetrical peak shapes, and selectivity in a 12 min separation time for TGR, GMR and OTR were obtained with a mixture of acetonitrile (40% by volume) and 0.1% triethylamine in distilled water (pH 2.5, 60% by volume). The same mixture was maintained with 1.0 ml per min flow scale. The elution times were 2.456 min for OTR, 7.236 min for TGR and 8.629 min for GMR (fig. 2).
Fig. 2: TGR, GMR and OTR chromatogram
As depicted in fig. 3, the absorbance spectrum of the TGR, GMR, and OTR got scanned across the 200-400 nm range. The data information obtained was saved in the computer. The 282 nm was shown to represent the ideal wavelength for determining TGR, GMR, and OTR.
Fig. 3: Absorbance spectrum of the TGR, GMR, and OTR
Fig. 4a: Mobile phase blank (without TGR, GMR, and OTR) chromatogram
Validation
When validating the established HPLC: TGR, GMR, and OTR analysis technique, the ICH recommendations were considered taking into account [27, 28].
Selectivity
The selectivity was evaluated in order to identify compounds that might interfere with TGR, GMR, and OTR elution in the chromatogram. Mobile phase blank, working TGR, GMR, and OTR solution (20 µg/ml TGR, 5.8 µg/ml GMR and 15.8 µg/ml OTR) and Tegonat capsule solution (20 µg/ml TGR, 5.8 µg/ml GMR and 15.8 µg/ml OTR) were evaluated to look after selectivity. Fig. 4a to 4c show the comparable chromatograms.
System suitability
Five injections of the working TGR (20 µg/ml), GMR (5.8 µg/ml), and OTR (15.8 µg/ml) sample solution were made in a volume of 10 µl. Peak areas, theoretical plate numbers, elution durations, tail factors for TGR, GMR and OTR were calculated. The standard deviation (SD) and relative standard deviation (RSD) of these variables for TGR, GMR and OTR were calculated were also calculated (table 1).
Fig. 4b: Working TGR, GMR, and OTR solution chromatogram
Fig. 4c: Tegonat capsule solution chromatogram
Table 1: System suitability of HPLC: TGR, GMR, and OTR analysis technique
Statistics ↓ | Drug area | Drug tailing | Column plate number | Resolution among drugs | Drug’s elution time |
OTR (15.8 µg/ml) | |||||
Mean† | 2269534 | 1.126 | 5083.6 | - | 2.482 |
SD‡ | 15440.18 | 0.016733 | 5.176872 | - | 0.002236 |
%RSD | 0.6803 | 1.4861 | 0.1018 | - | 0.0901 |
TGR (20.0 µg/ml) | |||||
Mean† | 2652826 | 0.974 | 12832.8 | 21.784 | 7.2244 |
SD‡ | 17495.66 | 0.011402 | 177.7321 | 0.126016 | 0.007987 |
%RSD | 0.6595 | 1.1706 | 1.3850 | 0.5785 | 0.1106 |
GMR (5.8 µg/ml) | |||||
Mean† | 1263457 | 0.976 | 14619.6 | 5.14 | 8.596 |
SD‡ | 22650.61 | 0.013416 | 222.8493 | 0.057446 | 0.040305 |
%RSD | 1.7927 | 1.3746 | 1.5243 | 1.1176 | 0.4689 |
SD–standard deviation; † three number of experiment average; ‡ Deviation for three number of experiment; RSD–relative standard deviation
Linearity
While the linearity test, six calibration solution standards from the ranges of 5.0-30.0 µg/ml TGR, 1.45-8.70 µg/ml GMR, and 3.95-23.70 µg/ml OTR were analysed for the calibration curves of TGR, GMR, and OTR. The peak area and drug quantity were plotted in order to create the drug calibration curve. The intercept, R2 and slope for TGR, GMR, and OTR were obtained by the application of linear regression. Fig. 5, illustrates the data's linear fit. Linear equation obtained were: y = 14080x+12823 for OTR; y = 13236x+8507.6 for TGR; and y = 21422x+12540 for GMR. The R2 for OTR was 0.9994; for TGR was 0.9996; and for GMR was 0.9995.
Fig. 5: Data's linear fit for TGR, GMR, and OTR
Sensitivity
In order to fig. out the LOD and LOQ values for TGR, GMR, and OTR, the “signal-to-noise” (STR) ratio was exercised. The LOD of TGR, GMR, and OTR were 0.2 µg/ml, 0.058 µg/ml, and 0.158 µg/ml, respectively. The LOQ of TGR, GMR, and OTR were 0.606 µg/ml, 0.175 µg/ml, and 0.478 µg/ml, respectively. Fig. 6 illustrates the chromatograms of TGR, GMR, and OTR at corresponding LOD and LOQ quantities.
Fig. 6: TGR, GMR, and OTR chromatograms at corresponding LOD and LOQ concentrations
Precision
Mixed standard TGR, GMR, and OTR solutions (TGR-20 g/ml; GMR–5.8 g/ml; and OTR–15.8 g/ml) were evaluated six times within a single day for evaluating system precision and method precision. The peak area and assay of TGR, GMR, and OTR on a single day were used to compute the RSD percentile of TGR, GMR, and OTR. Regarding system precision, the RSD percentile of peak areas of TGR, GMR, and OTR were employed. But for method precision, the RSD percentile of TGR, GMR, and OTR assay was used (table 2).
Table 2: Precision of HPLC: TGR, GMR, and OTR analysis technique
Statistics ↓ |
System precision | ||
TGR | GMR | OTR | |
Mean† | 2649280 | 1268729 | 2270375 |
SD‡ | 17898.44 | 24024.98 | 13962.59 |
%RSD | 0.676 | 1.894 | 0.615 |
Statistics ↓ |
Method precision | ||
TGR | GMR | OTR | |
Mean†† | 99.227 | 100.860 | 99.871 |
SD‡ | 0.2416 | 0.2952 | 0.2684 |
%RSD | 0.243 | 0.293 | 0.269 |
SD–standard deviation; † six number of experiment average for peak area; †† six number of experiment average for percentile assay; ‡ Deviation for six number of experiment; RSD–relative standard deviation
Ruggedness
The investigation of ruggedness used conscious and significant observable changes, such as analyst-analyst and day-day, while keeping the other experimental HPLC: TGR, GMR, and OTR analysis technique circumstances and parameters constant. Mixed standard TGR, GMR, and OTR solutions (TGR-20 g/ml; GMR–5.8 g/ml; and OTR–15.8 g/ml) were evaluated six times for evaluating ruggedness. In table 3, data for ruggedness were displayed.
Table 3: Ruggedness of HPLC: TGR, GMR, and OTR analysis technique
Experiment | GMR assay (%) | OTR assay (%) | TGR assay (%) | |||
Analyst 1 Lab 1 |
Analyst 2 Lab 2 |
Analyst 1 Lab 1 |
Analyst 2 Lab 2 |
Analyst 1 Lab 1 |
Analyst 2 Lab 2 |
|
1 | 101.098 | 100.086 | 100.139 | 101.487 | 99.294 | 98.848 |
2 | 101.066 | 99.813 | 100.102 | 101.172 | 99.44 | 99.031 |
3 | 101.098 | 100.154 | 100.091 | 101 | 99.531 | 99.191 |
4 | 100.979 | 99.962 | 99.915 | 100.821 | 98.988 | 98.896 |
5 | 100.775 | 99.086 | 99.798 | 100.74 | 99.103 | 98.406 |
6 | 100.382 | 99.008 | 99.448 | 100.293 | 99.074 | 98.027 |
Mean† | 100.292 | 100.417 | 98.986 | |||
SD‡ | 0.7456 | 0.6175 | 0.4207 | |||
%RSD | 0.743 | 0.615 | 0.425 |
SD–standard deviation; † twelve number of experiment average for percentile assay; ‡ Deviation for twelve number of experiment; RSD–relative standard deviation
Table 4: Recovery of TGR with HPLC: TGR, GMR, and OTR analysis technique
µg/ml TGR added | µg/ml TGR found | % TGR recovered | Recovered (%) mean† | SD‡ | RSD% |
50% study level | |||||
10.19 | 10.250 | 100.589 | 100.493 | 0.1703 | 0.169 |
10.15 | 10.180 | 100.296 | |||
10.11 | 10.170 | 100.593 | |||
100% study level | |||||
19.89 | 19.890 | 100.000 | 100.168 | 0.2102 | 0.210 |
19.85 | 19.870 | 100.101 | |||
19.78 | 19.860 | 100.404 | |||
150% study level | |||||
30.05 | 30.160 | 100.366 | 100.044 | 0.2829 | 0.283 |
30.21 | 30.160 | 99.834 | |||
30.02 | 30.000 | 99.933 |
SD–standard deviation; † three number of experiment average for percentile assay; ‡ Deviation for three number of experiments; RSD–relative standard deviation
Table 5: Recovery of GMR with HPLC: TGR, GMR, and OTR analysis technique
µg/ml GMR added | µg/ml GMR found | % GMR recovered | Recovered (%) mean† | SD‡ | RSD% |
50% study level | |||||
2.99 | 2.990 | 100.000 | 100.335 | 0.3355 | 0.334 |
2.98 | 3.000 | 100.671 | |||
2.99 | 3.000 | 100.334 | |||
100% study level | |||||
5.89 | 5.920 | 100.509 | 100.226 | 0.3520 | 0.351 |
5.91 | 5.930 | 100.338 | |||
5.94 | 5.930 | 99.832 | |||
150% study level | |||||
8.65 | 8.610 | 99.538 | 99.730 | 0.1760 | 0.176 |
8.62 | 8.610 | 99.884 | |||
8.58 | 8.560 | 99.767 |
SD–standard deviation; † three number of experiment average for percentile assay; ‡ Deviation for three number of experiments; RSD–relative standard deviation
Recovery study/accuracy
The recovery study of TGR, GMR, and OTR was determined by incorporating additional quantities of TGR, GMR, and OTR into the Tegonat capsule solution, which contains 20 µg/ml TGR, 5.8 µg/ml GMR, and 15.8 µg/ml OTR. TGR recovery (table 4) was seen to range from 100.044% to 100.493, with a minimal RSD (0.169%-0.283%). The recoveries of GMR (table 5) ranged from 99.730% to 100.335%, with an RSD percentile of 0.176% to 0.351%. OTR recovery (table 6) was seen to range from 100.064% to 100.543%, with a percentile minimal RSD of 0.131% to 0.260%.
Robustness
The flow rate as well as the acetonitrile percentage in the mobile phase, were tweaked to assess robustness. Mixed standard TGR, GMR, and OTR solutions (TGR-20 µg/ml; GMR–5.8 µg/ml; and OTR–15.8 µg/ml) were evaluated in each case three times for evaluating robustness. In each case, the assay percentiles of TGR, GMR, and OTR were determined (table 7). The percentage recoveries were 98.987% to 100.467% for GMR, 99.233% to 100.267% for OTR and 98.500% to 100.733% for TFR. The RSD percentiles were found as 0.058% to 0.304% for GMR, 0.058% to 1.202% for OTR and 0.115% to 0.703% for TFR.
Table 6: Recovery of OTR with HPLC: TGR, GMR, and OTR analysis technique
µg/ml OTR added | µg/ml OTR found | % OTR recovered | Recovered (%) mean† | SD‡ | RSD% |
50% study level | |||||
7.99 | 8.040 | 100.626 | 100.543 | 0.2616 | 0.260 |
8.01 | 8.030 | 100.25 | |||
7.97 | 8.030 | 100.753 | |||
100% study level | |||||
15.80 | 15.790 | 99.937 | 100.064 | 0.1684 | 0.168 |
15.76 | 15.760 | 100.000 | |||
15.68 | 15.720 | 100.255 | |||
150% study level | |||||
23.34 | 23.400 | 100.257 | 100.286 | 0.1314 | 0.131 |
23.31 | 23.410 | 100.429 | |||
23.34 | 23.380 | 100.171 |
SD–standard deviation; †three number of experiment average for percentile assay; ‡Deviation for three number of experiments; RSD–relative standard deviation
Table 7: Robustness of HPLC: TGR, GMR, and OTR analysis technique
Robustness parameters → | Flow (+): 1.1 ml a min |
Flow (-): 0.9 ml a min |
Acetonitrile (+): 45% by volume |
Acetonitrile (-): 35% by volume |
GMR (5.8 µg/ml) | ||||
Recovered (%) Mean† | 98.967 | 100.467 | 100.033 | 100.367 |
SD‡ | 0.0577 | 0.3055 | 0.1155 | 0.2082 |
%RSD | 0.058 | 0.304 | 0.115 | 0.207 |
OTR (15.8 µg/ml) | ||||
Recovered (%) Mean† | 99.933 | 99.767 | 100.267 | 99.233 |
SD‡ | 0.0577 | 0.0577 | 0.2082 | 1.1930 |
%RSD | 0.058 | 0.058 | 0.208 | 1.202 |
TGR (20.0 µg/ml) | ||||
Recovered (%) Mean† | 100.033 | 100.733 | 100.367 | 98.500 |
SD‡ | 0.3215 | 0.5686 | 0.1155 | 0.6928 |
%RSD | 0.321 | 0.564 | 0.115 | 0.703 |
SD–standard deviation; † three number of experiment average for percentile assay; ‡ Deviation for three number of experiments; RSD–relative standard deviation
Degradation studies
On treatment with 5 ml of acid (1N strength HCl) in a water bath thermostated at 60 °C for 30.0 min, 13.913% of GMR, 18.672% of OTR and 15.159% of TFR were found degraded. The acid-degraded Tegonat capsule chromatogram showed two additional peaks with elution times 3.874 min and 6.167 min (fig. 7). At a temperature of 60 °C in a thermostated water bath for 30.0 min, TGR, GMR, and OTR peaks showed nearly12.832%, 15.681% and 17.836% degradation underneath the alkaline (1 N strength NaOH, 5 ml) condition, respectively. Fig. 7 displays the elution patterns of the degradation products TGR, GMR, and OTR. Two additional peaks with elution times 3.774 min and 6.057 min were seen. Degradations of TGR (15.159%), GMR (18.548%), and OTR (19.600%) were also detected over oxidative condition (30% peroxide, 5 ml) at 60 °C in a thermostated water bath for thirty min. The oxidized Tegonat capsule chromatogram disclosed one additional peaks with an elution time of 4.941 min (fig. 7). An 13.814% degradation of GMR, 10.004% degradation of OTR and 9.651% degradation of TFR were found with exposure of TGR, GMR, and OTR to photostability chamber over 72 lux hrs. After degradation of TGR, GMR, and OTR in the photostability chamber over 72 lux hrs, two degradant peaks with 19.328 min retention time and 20.751 min retention time (fig. 7). GMR degradation was 2.097%, OTR degradation was 12.549% and TFR degradation was 13.049% at 105 °C in a thermostated oven for three hr. One degradant peak having 14.724 elution time (fig. 7) was found with TGR, GMR, and OTR degradation at 105 °C in a thermostated oven for three hrs. Hydrolytic degradation of TGR (0.070%), GMR (3.804%), and OTR (0.040%) at 60 °C in a thermostated water bath for 30.0 min did not yield significant degradation. The chromatogram of hydrolytic degradation of TGR, GMR, and OTRat 60 °C in thermostated water bath for 30.0 min did not display any additional peaks (fig. 7).
Selectivity
By assessing the peak purities of TGR, GMR, and OTR using LabSolution software, the method's selectivity was explored. The peak purities of TGR, GMR, and OTR were evaluated to make absolutely sure that neither comigration components affected the responsiveness of the TGR, GMR, and OTR peaks. The analytes (TGR, GMR, and OTR) were clearly segregated across all degradation compounds (fig. 7), and the peak purities, as well as peaks of TGR (fig. 8), GMR (fig. 9), and OTR (fig. 10) obtained across all stress conditions throughout forced degradation experiments, were pure and homogenous.
Fig. 7: TGR, GMR, and OTR chromatograms after accelerated degradation tests
Fig. 8: Peak purity plots of TGR after accelerated degradation tests
Fig. 9: Peak purity plots of GMR after accelerated degradation tests
Fig. 10: Peak purity plots of OTR after accelerated degradation tests
Application
Commercially accessible capsule products containing TGR, GMR, and OTR were analysed in order to assess the usability of this newly devised HPLC approach. The TGR, GMR, and OTR recovery values were ascertained in an attempt to appraise the method's applicability. The contents of two determinations, expressed as percentages, were 100.185% and 100.147%, with an RSD percentile of 0.026 and mean content being 100.166% for OTR. The percentages were 99.491% and 99.637%, with RSD percentile of 0.104 and mean content being 99.564% for TGR. The percentages were 100.510% and 100.478%, with RSD percentile of 0.022 and mean content being 100.494% for GMR.
DISCUSSION
Shiraiwa et al. reported UPLC-MS/MS [22], Peer et al. and Ki et al. reported LC-MS/MS [23, 25], Marta et al. reported immunoassay [24] methodologies to quantity TGR in uracil/5-fluorouracil/GMR/oxonic acid on human plasma. The simultaneous measurement of TGR, GMR, and OTR was not addressed by any of the described approaches [22–25]. The HPLC method developed in our work is the first to concurrently measure TGR, GMR, and OTR in bulk and formulations of the capsule type.
The ICH recommendations were considered taking into account while validating the established HPLC: TGR, GMR, and OTR analysis technique [27]. Excipients in capsule formulation and mobile phase ingredients were not observed to interfere with TGR, GMR, and OTR elution. The selectivity was supported by the chromatograms (fig. 4a to 4c) of selectivity [29].
After analysing the results of five analytical replications, it was concluded that there were no discernible variations in the responses. The high degree of instrument precision was reflected by the relative standard deviation (RSD) during system appropriateness testing of the approach, which was ascertained to be<2 (0.0901% to 1.7927%) [30]. The data disclosed an acceptable fit to the regression line with R2 values of 0.9994 for OTR, 0.9996 for TGR, and 0.9995 for GMR [31]. The LOD is the quantity where the signal intensity of the TGR/GMR/OTR is at minimum three times that of the baseline signal noise [32]. The LOQ is the quantity where the signal intensity of the TGR/GMR/OTR is at minimum ten times that of the baseline signal noise [32]. The low fig. of LOD and LOQ for TGR, GMR, and OTR disclosed ample sensitivity of HPLC: TGR, GMR, and OTR analysis technique
For TGR, GMR, and OTR, the RSD percentile of peak area variability varied from 0.615% to 1.894%, demonstrating admissible system precision [33, 34]. While the RSD percentile of assay variability for TGR, GMR, and OTR varied from 0.243% to 0.293%, exhibiting excellent technique precision. The RSD percentile of assay variability for TGR, GMR, and OTR was consistently determined to be under 2%, exhibiting excellent technique ruggedness [33, 34]. The TGR, GMR, and OTR mean recoveries, which are close to 100%, and the lower RSD (lower than 1%), show that the HPLC: TGR, GMR, and OTR analysis technique was accurate [33, 34]. The RSD (0.058%-1.202%) and percentile assays (98.987%-100.733%) for TGR, GMR, and OTR obtained under all robustness experimental circumstances fell within allowable ranges for all modifications to the analytic conditions. The procedure is hence robust [33, 34].
The accelerated degradation tests demonstrate the TGR, GMR, and OTR's vulnerability to deterioration in basic, heat, UV, acidic, as well as oxidative circumstances [35, 36]. Chromatographic separation of opted analytes (TGR, GMR, and OTR) from distinct degradation products was achieved. This shows that the new approach has stability indicating efficiency and specificity for determining TGR, GMR, and OTR in pharmaceutical capsule formulations and bulk raw samples.
The established HPLC: TGR, GMR, and OTR analysis methodology is quick and simple, and it has been successfully used to assure the TGR, GMR, and OTR in their formulation capsule form meet quality standards. The time-and cost-efficient, well-established HPLC: TGR, GMR, and OTR analytical approach stand out as a potent method for achieving effective and reliable separation of TGR, GMR, and OTR. These benefits, which have been scientifically verified, support the quality control laboratories' attempts to use this HPLC: TGR, GMR, and OTR analytical approach. Another benefit of this approach is that it accomplished excellent separation of all analytes (TGR, GMR, and OTR) and degradation products, as seen by the chromatograms of TGR, GMR, and OTR following degradation tests.
Nil
Declared none
Sung H, Ferlay J, Siegel RL, Laversanne M, Soerjomataram I, Jemal A. Global cancer statistics 2020: GLOBOCAN estimates of incidence and mortality worldwide for 36 cancers in 185 countries. CA Cancer J Clin. 2021;71(3):209-49. doi: 10.3322/caac.21660, PMID 33538338.
Jemal A, Bray F, Center MM, Ferlay J, Ward E, Forman D. Global cancer statistics. CA Cancer J Clin. 2011;61(2):69-90. doi: 10.3322/caac.20107, PMID 21296855.
Sitarz R, Skierucha M, Mielko J, Offerhaus GJA, Maciejewski R, Polkowski WP. Gastric cancer: epidemiology, prevention, classification, and treatment. Cancer Manag Res. 2018;10:239-48. doi: 10.2147/CMAR.S149619. PMID 29445300.
Yen HH, Chen CN, Yeh CC, Lai IR. Adjuvant tegafur-uracil (UFT) or S-1 monotherapy for advanced gastric cancer: a single center experience. World J Surg Oncol. 2021;19(1):124. doi: 10.1186/s12957-021-02233-2, PMID 33865416.
Mahlberg R, Lorenzen S, Thuss Patience P, Heinemann V, Pfeiffer P, Mohler M. New perspectives in the treatment of advanced gastric cancer: S-1 as a novel oral 5-FU therapy in combination with cisplatin. Chemotherapy. 2017;62(1):62-70. doi: 10.1159/000443984, PMID 27643822.
Matt P, Van Zwieten Boot B, Calvo Rojas G, Ter Hofstede H, Garcia Carbonero R, Camarero J. The European medicines agency review of tegafur/gimeracil/oteracil (Teysuno™) for the treatment of advanced gastric cancer when given in combination with cisplatin: summary of the scientific assessment of the committee for medicinal products for human use (CHMP). Oncologist. 2011;16(10):1451-7. doi: 10.1634/theoncologist.2011-0224, PMID 21963999.
Liu H, Wang Y, Li G, Song W, Wang R. Clinical study of tegafur-gimeracil-oteracil potassium capsule (s-1) and oxaliplatin combination chemotherapy in advanced colorectal cancer. J Cancer Res Ther. 2015;11(2):331-5. doi: 10.4103/0973-1482.157339, PMID 26148595.
Chen J, Wang J, Xu T. Comparison of efficacy and safety of S-1 and capecitabine in patients with metastatic colorectal carcinoma: a systematic review and meta-analysis. Medicine. 2019;98(30):e16667. doi: 10.1097/MD.0000000000016667, PMID 31348323.
Ahmed S, Islam MS, Ullah B, Kanti Biswas S, Samad Azad MA, Hossain MS. A review article on pharmaceutical analysis of pharmaceutical industry according to pharmacopoeias. Orient J Chem. 2020;36(1):1-10. doi: 10.13005/ojc/360101.
Zacharis CK, Markopoulou CK. Recent trends in pharmaceutical analytical chemistry. Molecules. 2020;25(16):3560. doi: 10.3390/molecules25163560, PMID 32764423.
Nikolin B, Imamovic B, Medanhodzic Vuk S, Sober M. High performance liquid chromatography in pharmaceutical analyses. Bosn J Basic Med Sci. 2004;4(2):5-9. doi: 10.17305/bjbms.2004.3405, PMID 15629016.
Martins YA, Gonçalves TM, Lopez RFV. HPLC methods for choloroquine determination in biological samples and pharmaceutical products. Daru. 2021;29(1):223-39. doi: 10.1007/s40199-021-00391-y, PMID 33738722.
Fares MY, Hegazy MA, El-Sayed GM, Abdelrahman MM, Abdelwahab NS. Quality by design approach for green HPLC method development for simultaneous analysis of two thalassemia drugs in biological fluid with pharmacokinetic study. RSC Adv. 2022;12(22):13896-916. doi: 10.1039/d2ra00966h, PMID 35548387.
Kowalska M, Wozniak M, Kijek M, Mitrosz P, Szakiel J, Turek P. Management of validation of HPLC method for determination of acetylsalicylic acid impurities in a new pharmaceutical product. Sci Rep. 2022;12(1):1. doi: 10.1038/s41598-021-99269-x, PMID 34992227.
Yadav PK, Tiwari AK, Saklani R, Chauhan D, Rana R, Yadav P. HPLC method for simultaneous estimation of paclitaxel and baicalein: pharmaceutical and pharmacokinetic applications. Bioanalysis. 2022;14(14):1005-20. doi: 10.4155/bio-2022-0100, PMID 36066029.
Mallavarapu R, Katari NK, Dongala T, Rekulapally VK, Marisetti VM, Vyas G. A validated stability-indicating reversed-phase-HPLC method for dipyridamole in the presence of degradation products and its process-related impurities in pharmaceutical dosage forms. Biomed Chromatogr. 2022;36(1):e5247. doi: 10.1002/bmc.5247, PMID 34541698.
Naresh P, Kumar KV. Leflunomide tablet formulation: development and validation of an RP-HPLC technique. Int J Pharm Pharm Sci. 2023;15(1):16-21. doi: 10.22159/ijpps.2023v15i1.46505.
Attia KA, Madkour AW, Almrasy AA, Abdelfatah AM. Determination of toltrazuril in the presence of complete alkaline degradation product by RP-HPLC and TLC-densitometric methods. Int J Pharm Pharm Sci. 2022;14(5):49-56. doi: 10.22159/ijpps.2022v14i5.44179.
Chandar PB, Madhuri MS, Hyma KN, Anusha S, Rasool SN. ARP-HPLC stability indicating method development and validation for the simultaneous determination of netupitant and palonosetron in pharmaceutical dosage form. Int J Curr Pharm Sci. 2022;14(4):60-7. doi: 10.22159/ijcpr.2022v14i4.1985.
Katolkar P, Jaiswal S. Analytical method development and validation for the estimation of cyamemazine tartrate in formulation by RP-HPLC with stability indicating. Asian J Pharm Clin Res. 2022;15(9):28-32. doi: 10.22159/ajpcr.2022.v15i9.45154.
Ramoller IK, Abbate MTA, Vora LK, Hutton ARJ, Peng K, Volpe Zanutto F. HPLC-MS method for simultaneous quantification of the antiretroviral agents rilpivirine and cabotegravir in rat plasma and tissues. J Pharm Biomed Anal. 2022;213:114698. doi: 10.1016/j.jpba.2022.114698. PMID 35259714.
Shiraiwa K, Suzuki Y, Uchida H, Iwashita Y, Tanaka R, Iwao M. Simultaneous quantification method for 5-FU, uracil, and tegafur using UPLC-MS/MS and clinical application in monitoring UFT/lV combination therapy after hepatectomy. Sci Rep. 2021;11(1):3132. doi: 10.1038/s41598-021-82908-8, PMID 33542452.
Peer CJ, McManus TJ, Hurwitz HI, Petros WP. Development and utilization of a combined LC-UV and LC-MS/MS method for the simultaneous analysis of tegafur and 5-fluorouracil in human plasma to support a phase I clinical study of oral UFT®/leucovorin. J Chromatogr B Anal Technol Biomed Life Sci. 2012;898:32-7. doi: 10.1016/j.jchromb.2012.04.010, PMID 22565063.
Broto M, McCabe R, Galve R, Marco MP. A high throughput immunoassay for the therapeutic drug monitoring of tegafur. Analyst. 2017;142(13):2404-10. doi: 10.1039/c7an00418d. PMID 28555688.
Liu K, Zhong D, Zou H, Chen X. Determination of tegafur, 5-fluorouracil, gimeracil and oxonic acid in human plasma using liquid chromatography-tandem mass spectrometry. J Pharm Biomed Anal. 2010;52(4):550-6. doi: 10.1016/j.jpba.2010.01.026. PMID 20138454.
International conference on harmonization. ICH stability testing of new drug substances and products. Vol. Q1A(R2); 2003.
International conference on the harmonization. ICH harmonized tripartite guideline. Validation of analytical procedures: Text and methodology Q2 (R1); 2005.
Sharma S, Goyal S, Chauhan K. A review on analytical method development and validation. Int J App Pharm. 2018;10(6):8-15. doi: 10.22159/ijap.2018v10i6.28279.
Snyder LR, Dolan JW. Optimizing selectivity during reversed-phase high-performance liquid chromatography method development: prioritizing experimental conditions. J Chromatogr A. 2013;1302:45-54. doi: 10.1016/j.chroma.2013.05.082, PMID 23830242.
Sabir AM, Moloy M, Bhasin PS. HPLC method development and validation: a review. Int Res J Pharm. 2015;4(4):39-46. doi: 10.7897/2230-8407.04407.
Ismail R, Lee HY, Mahyudin NA, Abu Bakar F. Linearity study on detection and quantification limits for the determination of avermectins using linear regression. J Food Drug Anal. 2014;22(4):407-12. doi: 10.1016/j.jfda.2014.01.026. PMID 28911453.
Armbruster DA, Pry T. Limit of blank, limit of detection and limit of quantitation. Clin Biochem Rev. 2008;29Suppl 1:S49-52. PMID 18852857.
Ira SK, Michael ES. Analytical method validation: back to basics, Part II. LCGC North Am. 2011;29(1):44-52.
Suresh R, Anarthanan S, Manavalan R, Valliappan K. Aspects of validation in HPLC method development for pharmaceutical analysis-comparison of validation requirements by FDA, USA and ICH. Int J Pharm Sci Res. 2010;12:123-33.
Rochani AK, Wheatley M, Oeffinger BE, Eisenbrey JR, Kaushal G. LC-MS based stability-indicating method for studying the degradation of lonidamine under physical and chemical stress conditions. Res Pharm Sci. 2020;15(4):312-22. doi: 10.4103/1735-5362.293509, PMID 33312209.
Henry TR, Penn LD, Conerty JR, Wright FE, Gorman G, Pack BW. Best practices in stability indicating method development and validation for non-clinical dose formulations. AAPS J. 2016;18(6):1418-23. doi: 10.1208/s12248-016-9976-0, PMID 27600135.