1Department of Pharmaceutical Analysis, Krishna Teja Pharmacy College, Tirupathi, Andhra Pradesh, India, 2Department of Quality Assurance, Krishna Teja Pharmacy College, Tirupathi, Andhra Pradesh, India
Email: prashanthi.chengalva87@gmail.com
Received: 11 Jan 2016 Revised and Accepted: 27 Feb 2016
ABSTRACT
Objective: To develop an accurate, precise and linear RP-HPLC method for simultaneous quantitative estimation of Metformin hydrochloride and Nateglinide in tablets and validate as per ICH guidelines.
Methods: The method used a reverse phase column, Inertsil C18-ODS 3V (2504.6 mm, 5 m), a mobile phase comprising of phosphate buffer (pH 4.0): Acetonitrile: methanol (30:60:10) flow rate of 1.0 ml/min and a detection wavelength of 221 nm using a UV detector.
Results: The developed method resulted in elution of Metformin hydrochloride at 2.45 min and Nateglinide at 4.21 min. The calibration curves were linear (r=0.999) in the concentration range of 60-140 g/ml and 14.4-33.2 g/ml for Metformin hydrochloride and Nateglinide respectively. The percentage recoveries were found to be 99.59-101.36 for Metformin hydrochloride and 98.43-101.38 for Nateglinide. The LOD was found to be 2.18 g/ml and 1.55 g/ml for Metformin hydrochloride and Nateglinide respectively. LOQ was found to be 8.52g/ml and 4.69g/ml for Metformin hydrochloride and Nateglinide respectively.
Conclusion: A simple, accurate, precise, linear and rapid RP-HPLC method was developed for simultaneous quantitative estimation of Metformin hydrochloride and Nateglinide in bulk and pharmaceutical formulation and validated as per ICH guidelines. Hence, the method holds good for the routine analysis of Metformin hydrochloride and Nateglinide in various pharmaceutical industries as well as in academics.
Keywords: Metformin hydrochloride, Nateglinide, RP-HPLC, Method development, Validation
2016 The Authors. Published by Innovare Academic Sciences Pvt Ltd. This is an open access article under the CC BY license (http://creativecommons.org/licenses/by/4.0/)
INTRODUCTION
Metformin hydrochloride (fig. 1) is 1, 1-Dimethyl biguanide monohydrochloride. It is the first-line medication for the treatment of type 2 diabetes. It is a biguanide antihyperglycemic agent used for treating noninsulin dependent diabetes mellitus (NIDDM). It improves glycemic control by decreasing hepatic glucose production, decreasing glucose absorption and increasing insulin-mediated glucose uptake. Metformin is the only oral anti-hyperglycemic agent that is not associated with weight gain.
Fig. 1: Structure of metformin hydrochloride
Nateglinide (fig. 2) is 2-(4-isoopropylcyclohexanecarboxamido)-3-phenylpropanoic acid. It is an oral antihyperglycemic agent used for the treatment of NIDDM. It belongs to the Meglitinide class of short-acting insulin secretagogues. It lowers blood glucose by stimulating the release of insulin from the pancreas, by closing ATP-dependent potassium channels in the membrane of the cells. This depolarizes the cells and causes voltage-gated calcium channels to open. The resulting calcium influx induces fusion of insulin-containing vesicles with the cell membrane and insulin secretion occurs. Nateglinide is extensively metabolized in the liver and excreted in urine (83%) and feces (10%).
A detailed literature survey revealed that there was a spectrophotometric method for simultaneous estimation of Metformin with other combination [1] and Nateglinide in tablet dosage form [2], and HPTLC method [3] have been established for simultaneous analysis of Metformin hydrochloride and Repaglinide. There were various RP-HPLC methods have been developed for the determination Metformin hydrochloride in combination with other drugs [4-11]. However, there was no RP-HPLC Method has been reported for simultaneous estimation of Metformin hydrochloride and Nateglinide, we presented easy, rapid, accurate and specific HPLC method for simultaneous estimation of RP-HPLC assay procedure for the analysis of Metformin hydrochloride and Nateglinide in bulk and tablet dosage form. The developed method was validated as per ICH guidelines [12].
Fig. 2: Structure of nateglinide
MATERIALS AND METHODS
Pharmaceutical grade Metformin hydrochloride and Nateglinide were supplied as a gift sample by Chandra labs, Hyderabad and marketed formulation (Netop plus, Metformin-500 mg, and Nateglinide-120 mg) was purchased from the local market. Methanol, Orthophosphoric acid (OPA), Acetonitrile and HPLC grade water were obtained from Merck. All solvents used in this work are HPLC grade. RP-HPLC Shimadzu (LC 20ATVP) model with Spin chrome (LC SOLUTIONS) software was employed in this method. Analytical column used for the separation of analytes is Inertsil ODS C18 (250x4.6 mm, 5).
Methods
Selection of wavelength
Standard solutions of Metformin hydrochloride and Nateglinide were prepared at the concentration of 10 g/ml scanned by UV/Vis spectrophotometer at the range of 200-400 nm. UV spectrums of Metformin hydrochloride (fig. 3) and Nateglinide (fig. 4) were shown below. The isosbestic point selected for simultaneous estimation was 221 nm (fig. 5).
Fig. 3: UV spectrum of Nateglinide
Fig. 4: UV spectrum of metformin hydrochloride
Fig. 5: UV overlaps spectrum of Nateglinide and Metformin hydrochloride
Chromatographic conditions
The developed method used a reverse phase C18 column, Waters Inertsil ODS 3V (250x4.6 mm, 5), a mobile phase of phosphate buffer (pH 4.0): Acetonitrile: methanol (30:60:10), flow rate of 1.0 ml/min and a detection wavelength of 221 nm using a UV detector.
Preparation of phosphate buffer
Accurately 2.7 g of potassium dihydrogen phosphate (KH2PO4) was weighed and dissolved in water and volume was made up to 1000 ml with water. The PH was adjusted to 4.0 by using Orthophosphoric acid. The buffer was filtered to remove all fine undissolved particles.
Preparation of mobile phase
A mixture of 30 volumes of phosphate Buffer, 60 volumes of Acetonitrile and 10 volumes of methanol was prepared. The mobile phase was sonicated for 10 min to remove gasses.
Diluent
The mobile phase was used as diluent.
Preparation of standard solutions
Standard stock solutions of Metformin hydrochloride and Nateglinide were prepared by dissolving 100 mg of Metformin and 24 mg of Nateglinide insufficient mobile phase. After that, the solution was filtered and sonicated for 5 min and diluted to 100 ml with mobile phase. Further dilutions were prepared in 5 replicates of 100 g/ml of Metformin and 24 g/ml of Nateglinide, which was made by adding 1 ml of stock solution to 10 ml of mobile phase. This has been treated as 100 % target concentration.
Preparation of sample solution
20 tablets each containing 500 mg of Metformin hydrochloride and 120 mg of Nateglinide were weighed and taken into a mortar and crushed to fine powder and uniformly mixed. Primary stock solutions of Metformin hydrochloride and Nateglinide (g/ml) were prepared by dissolving a weight equivalent to 500 mg of Metformin hydrochloride and 120 mg of Nateglinide and dissolved in sufficient mobile phase. After that, the solution was filtered by syringe filter and sonicated for 5 min and dilute to 100 ml with mobile phase. Further dilutions were prepared in 5 replicates of 100 g/ml of Metformin hydrochloride, and 24 g/ml of Nateglinide was made by adding 1 ml of stock solution to 10 ml of mobile phase. That has been treated as 100 % target concentration.
RESULTS AND DISCUSSION
Method development
Different chromatographic conditions were tried for better separation and resolution. Inertsil C18-ODS 3V (2504.6 mm, 5 m) column was found satisfactory. Peak purity of Metformin hydrochloride and Nateglinide was checked using UV detector and 221 nm was considered satisfactory for detecting both the drugs with adequate sensitivity. A number of solvents in the different ratio over a wide range of pH were tried, but either peak shape was broad or resolution was not good. Repeated trials to obtain good, sharp peak with an efficient resolution between two peaks of Metformin hydrochloride and Nateglinide done on a C18 column in isocratic HPLC gave satisfactory results. The run time was good in isocratic trial with mobile phase consisting of phosphate buffer (pH 4.0): Acetonitrile: methanol (30:60:10) and C18-ODS 3V (2504.6 mm, 5 m) column, flow rate 1.0 ml/min and detection wavelength 221 nm gave the satisfactory results in terms of retention time, resolution, symmetry and sensitivity. A typical RP-HPLC chromatogram for simultaneous determination of Metformin hydrochloride and Nateglinide from standard preparation and from pharmaceutical formulation was shown in (fig. 6 and 7).
Fig. 6: Typical chromatogram of standard solution
Fig. 7: Typical chromatogram of sample solution
Method validation
The developed RP-HPLC method was validated for parameters like system suitability, linearity, accuracy, precision, limit of detection (LOD), limit of quantitation (LOQ) and robustness according to ICH guidelines.
System suitability
Standard solutions were prepared as per the test method and injected into the chromatographic system. The system suitability parameters like theoretical plates, resolution and asymmetric factor were evaluated. The system suitability parameters were tabulated in table 1. All the parameters were found to be within the limits.
Precision
Method precision
The precision of the method was verified by precision method studies. The sample solution was prepared at working concentration and analysis was carried out at replicating. The sample solutions of Metformin hydrochloride and Nateglinide were prepared as per the test method and injected 6 times into the column. The results of precision were tabulated in table 2. The average was taken, and % RSD was calculated and reported. % RSD values were within the limits, and the method was found to be precise.
Linearity
The linearity of the test solutions for the assay method was prepared from Metformin hydrochloride and Nateglinide standard stock solution at five concentration levels from 50% to 150% of assay concentration. The peak area versus concentration data was treated by least-squares linear regression analysis (fig. 8 and 9). The results have shown an excellent correlation between peak areas and concentration within the concentration range of 60140 g/ml for Metformin hydrochloride, 14.433.2 g/ml for Nateglinide (tables 3-4). The correlation coefficients were found to be 0.999 for both the drugs, which meet the method validation acceptance criteria and hence the method was said to be linear for both the drugs.
Table 1: Results of system suitability studies
Parameters |
Acceptance limits |
Metformin hydrochloride |
Nateglinide |
Retention time (min) |
- |
2.453 |
4.217 |
Resolution |
NLT 2 |
- |
14.232 |
Theoretical plates |
NLT 2000 |
8489 |
7583 |
Tailing factor |
NMT 2 |
1.706 |
1.556 |
# min: minutes
Table 2: Method precision data
n |
Metformin hydrochloride |
Nateglinide |
||
Rt |
Peak area |
Rt |
Peak area |
|
Injection 1 |
2.459 |
3984.852 |
4.253 |
399.316 |
Injection 2 |
2.456 |
3983.028 |
4.243 |
396.285 |
Injection 3 |
2.453 |
3986.756 |
4.227 |
380.420 |
Injection 4 |
2.457 |
3983.864 |
4.210 |
390.391 |
Injection 5 |
2.459 |
3955.954 |
4.300 |
385.999 |
Injection 6 |
2.460 |
3915.924 |
4.367 |
385.427 |
Average |
2.456 |
3968.396 |
4.267 |
389.640 |
SD |
0.004 |
28.176 |
0.058 |
7.132 |
% RSD |
0.164 |
0.710 |
1.35 |
1.83 |
# Rt: Retention times, # n: number of injections=6
Fig. 8: Linearity chart for Metformin hydrochloride
Accuracy
The accuracy of the method was determined by recovery studies by the determination of % mean recovery of both the drugs at three different levels (80 %, 100 % and 120%). At each level, three determinations were performed. The percentage recovery and mean percentage recovery were calculated for the drug was shown in table 4. The observed data were within the required range, which indicates good recovery values and hence the accuracy of the method developed.
Fig. 9: Linearity chart for Nateglinide
Table 3: Linearity data for Metformin hydrochloride and Nateglinide
% Level |
Metformin hydrochloride concentration (g/ml.) |
Metformin peaks hydrochloride area |
Nateglinide concentration (g/ml) |
Nateglinide peak area |
50 |
60 |
2088.713 |
14.4 |
201.895 |
75 |
80 |
3002.121 |
19.2 |
292.713 |
100 |
100 |
3941.042 |
24 |
385.139 |
125 |
120 |
4890.891 |
28.8 |
476.353 |
150 |
140 |
5854.513 |
33.2 |
560.788 |
Correlation Coefficient |
0.9998 |
0.9996 |
||
Slope |
37.68 |
3.60 |
Table 4: Results of accuracy
Level (%) |
Metformin hydrochloride |
Nateglinide |
|||
% Recovery |
% Mean |
% Recovery |
% Mean |
||
80 |
98.59 |
99.59 |
100.74 |
100.42 |
|
80 |
99.63 |
99.90 |
|||
80 |
100.57 |
100.63 |
|||
100 |
99.48 |
100.46 |
99.53 |
99.30 |
|
100 |
100.54 |
98.43 |
|||
100 |
101.36 |
99.95 |
|||
150 |
101.20 |
100.09 |
101.38 |
101.14 |
|
150 |
99.08 |
100.78 |
|||
150 |
100.01 |
101.27 |
Table 5: Results of robustness
Parameter |
Metformin hydrochloride |
Nateglinide |
||
Plate count |
Tailing |
Plate count |
Tailing |
|
Less flow rate (0.8 ml/min) |
8828 |
1.765 |
7336 |
1.500 |
More flow rate (1.2 ml/min) |
8789 |
1.706 |
7114 |
1.556 |
Less wavelength 216 nm |
8828 |
1.765 |
7180 |
1.519 |
More wavelength 226 nm |
8451 |
1.667 |
7123 |
1.538 |
Table 6: LOD and LOQ
Sample name |
LOD(g/ml) |
LOQ(g/ml) |
Metformin hydrochloride |
2.18 |
8.52 |
Nateglinide |
1.55 |
4.69 |
Robustness
To determine the robustness of the developed method, experimental conditions were deliberately altered, and the system suitability parameters were evaluated. The solutions prepared as per the test method and injected at different variable conditions like flow rate (0.8, 1.2 ml/min.) and wavelength (219 nm, 223 nm), system suitability parameters were compared with that of method precision. The results were tabulated in table 5. At the flow rate of 1.0 ml/min shows, a sharp peak with good resolution and rest of the flow rates were found to be not satisfactory. The method passed all system suitability parameters indicating that the method was robust.
Detection limit and quantification limit
Limit of detection (LOD) which represents the concentration of the analyte at S/N ratio of 3 and limit of Quantification (LOQ) at which S/N was 10 were determined experimentally for the proposed methods and results were given in table 6.Hence, the detection limits and quantitation limits of the drugs were given S/N ratios of 3 and 10 respectively.
CONCLUSION
The proposed RP-HPLC method was found to be simple, specific, accurate, precise, robust, rapid and economical. This method gives good resolution between all the two compounds with a short analysis time. The proposed RP-HPLC method can be useful for routine analysis of Metformin hydrochloride and Nateglinide in the tablet dosage form.
ACKNOWLEDGEMENT
The author expresses sincere thanks to the Principal and Head of Pharmaceutical Analysis department, Krishna Teja Pharmacy College for excellent guidance and encouragement.
CONFLICT OF INTERESTS
Declare none
REFERENCES