Department of Pharmaceutical Chemistry, Bharati Vidyapeeth Deemed University, Poona College of Pharmacy Erandwane, Pune, India.
Email: raojanhavi@rediffmail.com
Received: 11 Jun 2014 Revised and Accepted: 09 Sep 2014
ABSTRACT
Objective: The present study was designed with an objective of a simple, fast, precise, selective and accurate RP-HPLC method was developed and validated for the simultaneous determination of Aspirin, Ramipril, Hydrochlorothiazide, Simvastatin and Atenolol from bulk drug and formulation.
Methods: The separation was achieved on a Hypersil Gold column (250 mm X 4.6 mm, 5 µ) as stationary phase with a mobile phase consisting of methanol: water in the ratio of 95:5% v/v at a flow rate of 1 mL/min and UV detection at 230 nm.
Results: The retention times were observed to be 1.983, 2.525, 3.108, 3.867 and 7.833 minutes for Aspirin, Ramipril, Hydrochlorothiazide, Simvastatin and Atenolol, respectively. The method was statistically validated for linearity, recovery, the limit of detection, limit of quantification, accuracy and precision.
Conclusion: Thus, proposed method was found sensitive, precise, accurate and specific and be used for quantitative estimation of Aspirin, Ramipril, hydrochlorothiazide, Simvastatin and Atenolol in commercial pharmaceutical dosage form.
Keywords: Aspirin, Atenolol, Ramipril, Hydrochlorothiazide, Simvastatin, RP-HPLC.
INTRODUCTION
Cardiovascular disease (also called heart disease) is a class of diseases that involve the heart, the blood vessels (arteries, capillaries, and veins) or both [1]. There are many risk factors associated with heart diseases. Hypertension is the single biggest risk factor nowadays. Hence, antihypertensive therapy considerably reduces the risk of developing cardiovascular complications that cause a high mortality rate; use of evidence-based multidrug regimens for patients at high risk for cardiovascular disease would be cost-effective in low-income and middle-income countries. For prevention of high risk for cardiovascular disease various combinations of antihypertensive drugs along with aspirin, a calcium-channel, an angiotensin-converting-enzyme inhibitor and a statin are available. Combinations of drugs are used in the treatment of hypertension to improve the tolerability and decrease toxicity profile of the therapy [2-3].
Aspirin is chemically, 2-(acetyloxy) benzoic acid and exerts its anti-inflammatory, analgesic and antipyretic actions. Aspirin and other non-steroid anti-inflammatory drugs (NSAIDs) inhibit the activity of the enzyme called cyclooxygenase (COX) which leads to the formation of prostaglandins (PGs) that cause inflammation, swelling, pain and fever [4].
Ramipril is (2S, 3aS, 6aS)-1[(S)-N-[(S)-1-carboxy-3-phenylpropyl] alanyl] octa hydro cyclohepta [b] pyrrole-2-carboxylic acid, 1-ethyl ester. Ramipril is a prodrug belonging to the angiotensin-converting enzyme (ACE) inhibitor class of medications. It is metabolized to ramiprilat in the liver and, to a lesser extent, kidneys. Ramiprilat is a potent, competitive inhibitor of ACE, the enzyme responsible for the conversion of angiotensin I (ATI) to angiotensin II (ATII). ATII regulates blood pressure and is a key component of the renin-angiotensin-aldosterone system (RAAS). Ramipril may be used in the treatment of hypertension, congestive heart failure, nephropathy, and to reduce the rate of death, myocardial infarction and stroke in individuals at high risk of cardiovascular events [5].
Hydrochlorothiazide is 6-Chloro-3, 4-dihydro-2H-l, 2, 4-benzothiadiazine-7-sulfonamide 1, 1-dioxide, a thiazide diuretic, by blocking the sodium-chloride symporter and effectively reduces the osmotic gradient and water reabsorption throughout the nephron [6]. Simvastatin is a prodrug and is hydrolyzed to its active β-hydroxyacid form, simvastatin acid, after administration. Simvastatin is a specific inhibitor of 3-hydroxy-3-methylglutaryl-coenzyme A (HMG-CoA) reductase, the enzyme that catalyzes the conversion of HMG-CoA to mevalonate, an early and rate limiting step in the biosynthetic pathway for cholesterol [7].
Atenolol is chemically 4-(2-hydroxy-3-[(1-methylethyl) amino] propoxy] benzeneacetamide and belongs to a class of medications called beta-blockers. Atenolol is a beta-adrenergic (beta-1 selective) agent that blocks beta receptors on the heart [8].
A literature survey revealed that spectrophotometric, chromatographic methods have been reported for determination of ASP [9-14], RAM [15-19], HCTZ [20-27], SIM [28-30] and ATN [31-36] in single and multicomponent pharmaceutical formulations or from biological fluids. However, there were few HPLC methods for simultaneous estimation of ASP, RAM, HCTZ, SIM and ATN reported.
Analysis of ASP, RAM, HCTZ, SIM and ATN has been carried out by gradient HPLC with the flow rate of 1.5 mL/min [37] while the proposed method employs isocratic elution and hence is simpler. Another method reported for analysis of ASP, RAM, HCTZ, SIM and ATN required 13 min of HPLC runtime in comparison to the proposed method [38].
The proposed method is relatively better or comparable in terms of sensitivity, accuracy, and precision to the methods reported for analysis of ASP, RAM, HCTZ, SIM and ATN. Rapid simultaneous estimation of ASP, RAM, HCTZ, SIM and ATN using a simple isocratic HPLC system with high sensitivity of estimation indicates easy application for analysis. The structures of the drugs are shown in Fig.1.
MATERIALS AND METHODS
Chemicals and reagents
Aspirin (ASP), Hydrochlorothiazide (HCTZ), Ramipril (RAM), Simvastatin (SIM) and Atenolol (ATN), were kindly supplied by Wockhardt Pharmaceuticals Ltd., Aurangabad, Emcure Pharmaceuticals Ltd, Pune and Smilax Laboratories Limited, Hyderabad, India. Marketed sample of ASP, RAM, HCTZ, SIM and ATN (Polycap) in their combined capsule dosage form of Cadila Pharmaceuticals Ltd. India, was used for analysis. Each capsule contained 50mg of ATN, 100mg of ASP, 12.5mg of HCTZ, 5mg RAM and 20mg of SIM. For HPLC work, double distilled water was prepared in the laboratory. Methanol used was of HPLC grade and were purchased from Merck Chemicals, Mumbai, India.
Instrumentation
The separation of five drugs was carried out on an isocratic JASCO RP-HPLC system using Hypersil ODS C18 column (250× 4.6 mm i. d., particle size 5 mm) as a stationary phase and methanol: water (95:5% v/v) as mobile phase. The HPLC system consisted of intelligent HPLC pump model (Jasco PU 1580). The solutions were injected into the chromatograph through a Rheodyne valve, with a 20 μL loop with auto sampler (AS 1555).
The detector consisted of a UV/ VIS (Jasco UV 1575). Data were integrated using Jasco Borwin version 1.5, LC-Net II/ADC system. The mobile phase flow rate was 1 mL/min and a detection wavelength of 230 nm was selected for analysis (Fig.2). An ultrasonic bath was used to remove the air from the mobile phases, operating at ambient temperature.
ASP RAM
HCTZ SIM
ATN
Fig. 1: Chemical Structures of ASP, RAM, HCTZ, SIM, and ATN.
Fig. 2: Overlay Spectrum for ASP, RAM, HCTZ, SIM and ATN (λmax = 230 nm)
Preparation of Standard and Sample solutions
Accurately weighed 50mg of ATN, 100mg of ASP, 12.5mg of HCTZ, 5mg RAM and 20mg of SIM were dissolved in and diluted with methanol up to 50 ml to obtain a standard stock solution of ATN (1000 μg/ml), ASP (2000 μg/ml), HCTZ (250 μg/ml), RAM (100 μg/ml) and SIM (400 μg/ml).
To determine the content of ASP, RAM, HCTZ, SIM and ATN simultaneously in capsule (label claim: 50mg of ATN, 100mg of ASP, 12.5mg of HCTZ, 5mg RAM and 20mg of SIM per tablet), twenty capsule were weighed, their mean weight determined and they were finely powdered and powder equivalent to 50mg of ATN, 100mg of ASP, 12.5mg of HCTZ, 5mg RAM and 20mg of SIM was weighed. Then equivalent weight of the drug was transferred into a 50 ml volumetric flask containing 20 ml methanol, sonicated for 10 min and diluted to 50 ml with methanol to obtain solution of ATN (1000 μg/ml), ASP (2000 μg/ml), HCTZ (250 μg/ml), RAM (100 μg/ml) and SIM (400 μg/ml). The mixture was filtered using whatmann filter.
Method validation
The method was validated in accordance with ICH guidelines [39].
System suitability
System suitability of the HPLC method was determined by making six replicate injections from freshly prepared standard solutions and analyzing each solute for their resolution (Rs), retention time, theoretical plates number (N) and tailing factors (T).
Specificity
The specificity of the method was ascertained by analysis of drug standards and samples. The mobile phase resolved all the drugs very efficiently, as showed in Fig. 3. The identities of the peak for ASP, RAM, HCTZ, SIM and ATN were confirmed by comparing the Rt with those of standards.
Fig.3: Chromatogram of ASP, RAM, HCTZ, SIM and ATN
Linearity
Linearity is generally evaluated by visual inspection of a plot of signals as a function of analyte concentration or content. For determining linearity, calibration curves were plotted over a concentration range of 20-120 μg/mL for ASP, 1-6 μg/mL for RAM, 2.5-15 μg/mL for HCTZ, 4-24 μg/mL for SIM and 10–60 μg/mL for ATN, respectively. A 20 μL of sample solution was injected into the chromatographic system using fixed volume loop injector. Chromatograms were recorded. Calibration plots were constructed by plotting peak area against the corresponding amount of each drug.
Limit of detection and limit of quantitation
The LOD and LOQ were calculated according to the 3.3 σ/s and 10 σ/s criteria, respectively; where σ is the standard deviation of the peak area and s is the slope of the corresponding calibration curve.
Precision
The precision of the proposed method was assessed as repeatability and intermediate precision by preparing three different sample solutions at low, medium and high concentrations, which were freshly prepared and analyzed daily. These experiments were repeated 3 different days over a period of a week to evaluate day-to-day variability (intermediate precision).
Accuracy
To check the accuracy of the developed method and to study the interference of formulation additives, analytical recovery experiments were carried out by standard addition method, at 80, 100 and 120% level. The experiment was conducted in triplicate. Percentage recovery and relative standard deviation were calculated.
Robustness
The robustness of an analytical procedure is a measure of its capacity to remain unaffected by small, but deliberate variations in method parameters and provide an indication of its reliability during normal usage.
RESULTS AND DISCUSSION
Method development
The HPLC procedure was optimized for simultaneous determination of LOS, HCTZ and AMLO. Good resolution of both the components was obtained with methanol: water at ratio 95: 5 v/v. The flow rate of 1 mL/min was optimum. UV detection was made at 230 nm. At this wavelength ASP, RAM, HCTZ, SIM and ATN can be quantified. Hence, 230 nm determined empirically has been found to be optimum. The average retention times for ASP, RAM, HCTZ, SIM and ATN was found to be 1.983, 2.525, 3.108, 3.867 and 7.833 min, respectively.
System suitability
To ascertain its effectiveness, system suitability tests were carried out on freshly prepared stock solutions. The parameters obtained are shown in Table 1.
Table 1: System Suitability Parameters of RP-HPLC for capsule Analysis
Parameter | ASP | RAM | HCTZ | SIM | ATN |
Resolution (Rs) | 0.00 | 2.446 | 3.146 | 4.460 | 8.235 |
Retention Time in min | 1.983 | 2.525 | 3.108 | 3.867 | 7.833 |
Theoretical plates number (N) | 3423 | 5528 | 6782 | 4152 | 4567 |
Tailing Factor | 1.02 | 1.20 | 1.16 | 1.12 | 1.32 |
Table 2: Precision studies of proposed HPLC method
Conc. (μg/mL) |
Intra-day precision | Inter-day precision | ||||
Measured Conc. ± SD | (%) RSD | Recoverya(%) | Measured Conc. ± SD | (%) RSD | Recoverya(%) | |
Aspirin | ||||||
40 80 120 |
39.90 ± 0.18 79.92 ± 0.31 119.96 ± 0.84 |
0.45 0.39 0.70 |
99.75 99.90 99.97 |
39.87 ± 0.35 79.80 ± 0.64 119.82 ± 0.95 |
0.88 0.80 0.79 |
99.67 99.75 99.85 |
Ramipril | ||||||
2 4 6 |
1.98 ± 0.012 3.98 ± 0.051 5.98 ± 0.034 |
0.60 1.28 0.56 |
99.00 99.50 99.67 |
1.97 ± 0.020 3.95 ± 0.039 5.95± 0.061 |
1.01 0.98 1.02 |
98.50 98.75 99.17 |
Hydrochlorothiazide | ||||||
5 10 15 |
4.96 ± 0.058 9.96 ± 0.091 14.89 ± 0.068 |
1.17 0.91 0.46 |
99.20 99.60 99.27 |
4.94 ± 0.064 9.94 ± 0.025 14.86 ± 0.017 |
1.29 0.25 0.11 |
98.80 99.40 99.07 |
Simvastatin | ||||||
8 16 24 |
7.98 ± 0.090 15.97 ± 0.062 23.95 ± 0.078 |
1.13 0.39 0.33 |
99.75 99.81 99.79 |
7.95 ± 0.080 15.93 ± 0.050 23.93 ± 0.024 |
1.00 0.31 0.10 |
99.38 99.56 99.71 |
Atenolol | ||||||
20 40 60 |
19.82 ± 0.092 39.77 ± 0.082 59.85 ± 0.192 |
0.46 0.20 0.32 |
99.10 99.43 99.75 |
19.87 ± 0.080 39.86 ± 0.35 59.55 ± 0.21 |
0.40 0.88 0.35 |
99.35 99.65 99.25 |
aMean from three analysis
Fig. 4: Linear regression and residual plot for ASP
Fig. 5: Linear regression and residual plot for RAM
Fig. 6: Linear regression and residual plot for HCTZ
Fig. 7: Linear regression and residual plot for SIM
Fig. 8: Linear regression and residual plot for ATN
Linearity
Linearity was accessed by visualizing the calibration graph and plot of the residuals. The points distributed equally above and below the trend line showed linearity.
The linear regression equations were Y=5254X + 105362 (r2= 0.9996) for ASP, Y= 42392X + 6327 (r2= 0.9990) for RAM, Y= 21553X + 56927 (r2= 0.9995) for HCTZ, Y= 14133X + 47241 (r2= 0.9996) for SIM and Y= 9089X + 62236 (r2= 0.9996) for ATN.
The plots obtained from linear regression and residuals analysis are given in Figures 4, 5, 6, 7 and 8 for ASP, RAM, HCTZ, SIM and ATN, respectively.
Limit of Detection (LOD) and Limit of Quantitation (LOQ)
The LOD for ASP, RAM, HCTZ, SIM and ATN were found to be 6.60, 0.2, 0.23, 1.32 and 1.98 µg/mL, respectively.
The LOQ for ASP, RAM, HCTZ, SIM and ATN were found to be 14, 0.5, 0.6, 3.97 and 5 µg/mL, respectively.
Precision
Precision was determined by analysis of standard solutions containing concentrations of ASP, RAM, HCTZ, SIM and ATN covering the entire calibration range. The precision of the method, as intra-day variation (RSD %) was determined by analysis of these solutions three times on the same day. Inter-day precision (RSD %) was assessed by analysis of these solutions on three different days over a period of one week. The results of the precision studies are as shown in Table 2.
Accuracy
The difference between theoretical added amount and practically achieved amount is called accuracy of an analytical method. Accuracy was determined at three levels 80%, 100% and 120% of the target concentration in triplicate. The results are presented in Table 3.
Robustness
The results of robustness in the present method showed no significant changes occurring over changes which are summarized in Table 4. As the changes are not significant, we can say that the method is robust.
Table 3: Standard edition techniques for determination of ASP, RAM, HCTZ, SIM and ATN (n= 3)
Drug | Label claim
(mg/capsule) |
Amount
Added (%) |
Total amount
(mg) |
Amount
recovered (mg) |
Recovery
(%) |
ASP | 100 | 80
100 120 |
180
200 220 |
178.5
199.8 219.4 |
99.17
99.90 99.72 |
RAM | 5 | 80
100 120 |
9
10 11 |
8.95
9.95 10.92 |
99.44
99.50 99.27 |
HCTZ | 12.5 | 80
100 120 |
22.5
25 27.5 |
22.30
24.80 27.25 |
99.11
99.20 99.09 |
SIM | 20 | 80
100 120 |
36
40 44 |
35.85
39.75 43.75 |
99.58
99.37 99.43 |
ATN | 50 | 80
100 120 |
90
100 110 |
89.52
99.82 109.27 |
99.47
99.82 99.34 |
Table 4: Robustness evaluation of ASP, RAM, HCTZ, SIM and ATN
Chromatographic factors | Level | Chromatographic changes in tRa | ||||
ASP | RAM | HCTZ | SIM | ATN | ||
A: Flow rate mL/min. | ||||||
0.9 | -0.1 | 2.021 | 2.560 | 3.152 | 3.890 | 7.875 |
1 | 0.0 | 1.983 | 2.525 | 3.108 | 3.867 | 7.833 |
1.1 | +0.1 | 1.981 | 2.500 | 3.008 | 3.821 | 7.801 |
Mean ± SD | 1.995 ± 0.023 | 2.528 ± 0.030 | 3.089 ± 0.074 | 3.859 ± 0.035 | 7.836 ± 0.037 | |
B: % of methanol in the mobile phase (± 5%) | ||||||
90 | -5.0 | 2.024 | 2.562 | 3.154 | 3.891 | 7.876 |
95 | 0.0 | 1.983 | 2.525 | 3.108 | 3.867 | 7.833 |
100 | +5.0 | 1.982 | 2.501 | 3.007 | 3.822 | 7.803 |
Mean ± SD | 1.996 ± 0.024 | 2.529 ± 0.031 | 3.089 ± 0.075 | 3.860 ± 0.035 | 7.837 ± 0.037 |
aMean from three estimates
Analysis of marketed formulation
The proposed method was used to determine the potency of commercially available capsule (Polycap) containing 50mg of ATN, 100mg of ASP, 12.5mg of HCTZ, 5mg RAM and 20mg of SIM. Three replicate determinations (n=3) were carried out and Polycap Capsule were analysed, sharp and well defined peaks for ASP, RAM, HCTZ, SIM and ATN were obtained at Rt 1.983, 2.525, 3.108, 3.867 and 7.833 min, respectively, when scanned at 230 nm.
The amount of the label claim measured were 99.92 ± 1.03 % for ASP, 99.60 ± 1.14 % for RAM, 99.23 ±1.22% HCTZ, 99.65 ±1.40% and 99.82 ±1.37% for ATN.
CONCLUSION
A simple isocratic RP-HPLC method was developed and validated for the simultaneous quantitative assay of ASP, RAM, HCTZ, SIM and ATN in capsule dosage form. The validation results reveal that, method is precise, linear, robust and accurate, which proves the reliability of the proposed method. The short runtime and low solvent consumption are helpful for applying routine quality control analysis.