Int J Pharm Pharm Sci, Vol 6, Issue 11, 199-202Original Article

HPLC METHOD FOR SIMULTANEOUS DETERMINATION OF KETOPROFEN AND PRESERVATIVES IN GEL FORMULATION

BOYKA TSVETKOVA*, LILY PEIKOVA

Department of Pharmaceutical chemistry, Medical University – Sofia, Faculty of Pharmacy, 2 Dunav st., 1000 Sofia, Bulgaria.
Email: bojka@abv.bg

Received: 08 Sep 2014 Revised and Accepted: 06 Oct 2014


ABSTRACT

Objective: The aim of presented study was to develop a high-performance liquid chromatography (HPLC) method with UV detection for simultaneous determination of ketoprofen, methyl paraben and propyl paraben in a gel formulation.

Methods: The chromatography was carried out on a C18 (250 mm x 4.6 mm, 10 μm) column with methanol, acetonitrile and 1.5 % sodium acetate solution (15:35:50 v/v/v) as mobile phase, at a flow rate of 1.0 ml/min, with detection at 240 nm.

Results: Under these chromatographic conditions, the obtained retention times were approximately 2.41 min for methyl paraben, 3.33 min for propyl paraben and 7.41 min for ketoprofen. Analytical parameters specificity, linearity, accuracy, precision and robustness were determined by validation procedure and found to be satisfactory.

Conclusion: Overall, the proposed method was found to be simple, rapid, precise and accurate for quality control of ketoprofen in the presence of preservatives in gel formulation.

Keywords: Liquid chromatography, Ketoprofen, Methyl paraben, Propyl paraben, Validation, Gel formulation.


INTRODUCTION

Ketoprofen (KET) is a non-steroidal anti-inflammatory drug (NSAID) with well established analgesic and antipyretic properties used for the treatment of rheumatoid arthritis, osteoarthritis, ankylosing spondylitis and gout [1]. Although ketoprofen is poorly water soluble it is rapidly absorbed, metabolized and excreted, it causes some gastrointestinal complaints such as nausea, dyspepsia, diarrhea, constipation and some renal side effects like other NSAIDs. Topically applied NSAIDSs have several advantages in comparison with their systemically administrated counterparts. These drugs are simple for application and deliver high drug concentrations locally into affected tissues with limited side effects at the application site for prolonged periods [2].

Several types of analytical procedures have been proposed for the analysis of ketoprofen in pharmaceutical formulations. The procedures include capillary zone electrophoresis [3], UV-spectrophotometry [4, 5], high-performance liquid chromatography [6,7], flow injection technique with hemiluminiscence [8], flow injection with UV-detection [9], polarography [10], potentiometry [11], and quantitative Fourier transformation infrared spectrophotometry [12]. European Pharmacopoeia recommends acid-base titration for analysis of ketoprofen in substance, UV-spectrophotometry for its determination in capsules as well as liquid chromatography for assay in gel [13].

Methyl 4-hydroxybenzoate (methyl paraben) and propyl 4-hydroxybenzoate (propyl paraben) are well-known preservatives used primarily for their bactericidal and fungicidal properties. Parabens are a group of alkyl esters of p-hydroxybenzoic acid, having a low toxicity profile and a long history of use. It readily absorb from the gastrointestinal tract or through the skin [14]. The use of methyl paraben is toxic at higher concentrations due to estrogenic effect [15]. The estrogenic activity of parabens increases with the length of the alkyl group and it is well known that propyl paraben is estrogenic to a certain degree as well [16].

The determination of these substances in pharmaceuticals and cosmetics is important in quality control, especially considering the numerous reports of allergic reactions caused by preservatives. Some methods of analysis of investigated parabens either alone or in some other formulations are available including HPLC [17-25], solid phase extraction HPLC [26, 27], HPLC-MS [28].

Simultaneous determination of both parabens and ketoprofen is reported by several authors [29-33].

The aim of this paper was to develop a specific, precise and accurate chromatographic method able to be applied in quality control for the determination of ketoprofen in gel formulation in the presence of both preservatives methyl- and propyl parabens.

MATERIALS AND METHODS

Chemicals and Reagents

Ketoprofen, methyl- and propyl parabens were purchased from Sigma-Aldrich (Germany) as standards. A ketoprofen gel containing 2.5 % w/w active compound, 1 % w/w methyl paraben and 0.1 % w/w propyl paraben was obtained commercially. LC-grade methanol and acetonitrile were supplied from Merck (Germany). All other chemical reagents were of analytical grade.

Instrumentation and chromatographic conditions

Chromatographic separation was performed on modular HPLC system LC-10A Shimadzu (Japan) arranged with a LC-10A pump, solvent degasser DGU-3A, Rheodyne injector, column oven CTO-10A, SPD-M10A fixed wavelength detector and communication bus module CBM-10A. A LiChrosorb C18, 250 mm x 4.6 mm, 10 μm column was used as a stationary phase. The components were separated isocratically with a mobile phase consisting of 15 volumes methanol, 35 volumes acetonitrile and 1.5 % sodium acetate solution at a flow rate of 1.0 ml/min. The analysis was carried out at an ambient temperature and injection volume was 20 μl. The UV detector was set at 240 nm.

Preparation of reference solutions

Reference solution (a): The solution was prepared by dissolving 20.0 mg of accurately weighed methyl paraben in methanol in a 50.0 ml volumetric flask (C=400 µg/ml).

Reference solution (b): The solution was prepared by dissolving 20.0 mg of accurately weighed propyl paraben in methanol in a 100.0 ml volumetric flask (C=200 µg/ml).

Working reference solution: The solution was prepared by dissolving of accurately weighed 25.0 mg ketoprofen in the first step and in the second step by diluting 25.0 ml of reference solution (a) and 5.0 ml of reference solution (b) with methanol into a 50.0 ml volumetric flask. The concentrations of the investigated compounds in working reference solution were as follows: ketoprofen – 500 µg/ml, methyl paraben – 200 µg/ml and propyl paraben – 20 µg/ml, respectively.

Sample preparation

2 ml (accurately measured) of ketoprofen gel corresponding to 50 mg ketoprofen, 20 mg methyl paraben and 2 mg propyl paraben were transferred to a 20 ml volumetric flask. Ten ml of methanol were added and the flask was placed in an ultrasonic bath for 20 min. After cooling at the room temperature, the flask was filled with methanol to the volume mark, and the sample was centrifuged at 2000 rpm for 10 min. Five ml of the clear centrifugate was transferred to a 25 ml volumetric flask, and filled with mobile phase to the volume mark. After filtration through a 0.45 mm membrane filter, an aliquot of the sample solution was injected into the HPLC column.

Validation procedure

The analytical method developed was validated with respect to specificity, precision, accuracy, robustness and sensitivity.

Specificity

The ability of an analytical method to unequivocally assess the analyte in the presence of other components can be demonstrated by evaluating specificity. The specificity of the method was evaluated by assessing interference from excipients in the pharmaceutical dosage form prepared as a placebo solution.

Linearity and range

The linearity of the method was determined at six concentration levels ranging from 125 to 1000 µg/ml for ketoprofen, from 50 to 400 µg/ml for methylparaben and from 5 to 40 µg/ml for propyl paraben. The calibration curves were constructed by plotting peak areas versus concentrations of investigated compounds, and the regression equations were calculated. Each response was the average of three determinations.

Precision

Intraday precision (repeatability) was calculated using two concentrations of KET (250, 500 μg/ml), methyl paraben (100, 200 μg/ml) and propyl paraben (10, 20 μg/ml) in triplicate using proposed methods. The inter day precision (reproducibility) was repeated three times on three different days for analysis of two different concentration (250:100:10, 500:200:20 μg/ml) for analyzed drugs.

Accuracy

Accuracy of the method was evaluated by standard addition technique, which was performed by addition of known amounts of pure ketoprofen and both parabens to known concentrations of gel and analysed by proposed methods in triplicate.

Robustness

Thе robustness of the method is a measure of the capacity to remain unaffected by small variations in method parameters and provides indication of its reliability during normal usage. Robustness of the analytical procedure was studied by deliberately varying parameters like analytical wavelength (±2 nm) and flow rate (±0.2 ml/min). Column-to-column reproducibility was also checked by using a C18 column of different make (Nucleosil) with same dimension.

RESULTS AND DISCUSSION

Fig. 1, showed that under the experimental chromatographic conditions ketoprofen, methyl paraben and propyl paraben were completely separated one from each other. In the chromatogram of blank solution there are no interfering peaks at the retention times of the investigated peaks. This indicated that the method is selective and can be used for identification and simultaneous quantification of ketoprofen in the presence of preservatives in topical gel formulation.

Fig. 1: Chromatogram of working standards: methyl paraben, propyl paraben and ketoprofen

Retention times, number of theoretical plates and tailing factors obtained by using of the HPLC method were listed in Table 1.

Table 1: Chromatographic data from HPLC method (system-suitability test)

Parameter methyl paraben propyl paraben ketoprofen
Retention time (min) 2.41 3.33 7.41
Tailing factor 0.91 0.89 0.90
Theoretical plates 2578 1354 1523

Calibration and linearity

We prepared a series of 5 calibration solutions with a concentration range shown in Table 2. It was found that response (peak area) was proportional to concentration over the ranges tested with correlation coefficients greater than 0.9996. Calibration plot data slope (a), intercept (b), and correlation coefficient (r) were listed in Table 2.

Table 2: Validation data for the calibration plots

Compounds methyl paraben propyl paraben ketoprofen
Concentration range (μg/ml) 50-400 5-40 125-1000
Slope 2456.8 34111.7 47854.4
Intercept 678.1 -946.4 7845.1
Correlation coefficient (r) 0.9996 0.9999 0.9998

Precision

The values of % RSD (Table 3) for KET, MP and PP were found to be in the range from 0.32 to 0.87 indicating good repeatability and reproducibility of the analytical procedure.

Robustness

It was found that the elution order and resolution for both components were not significantly affected by small variation of the conditions. Results from study of the robustness of the method were listed in Table 5.

Limit of detection and limit of quantification

The limits of quantification (LOQ) and limit of detection (LOD) were evaluated based on signal-to-noise ratios by serial dilution of working reference solution. The LOQs for ketoprofen, methyl paraben and propyl paraben were found to be 1.0, 0.8 and 0.5 µg/ml respectively; the LODs – 0.2, 0.1 and 0.05 µg/ml, respectively.

Table 3: Precision of the method

Precision

Amount taken (μg/ml)

% Mean*

% RSD

KET

MP

PP

KET

MP

PP

KET

MP

PP

Intra day

250

100

10

99.56

100.5

99.61

0.36

0.32

0.54

Intra day

500

200

20

100.1

100.2

100.9

0.65

0.61

0.61

Inter day

250

50

10

99.48

99.89

99.81

0.84

0.56

0.38

Inter day

500

150

20

99.75

99.52

99.62

0.78

0.87

0.42

*Mean of three determinations

Results presented in Table 4 indicated good accuracy and showed no interference from tablet excipients.

Table 4: Recovery studies of ketoprofen and parabens

Compound Amount taken

(μg/ml)

Amount added

(μg/ml)

Amount recovered ± SD*

(μg/ml)

% RSD
ketoprofen 500.0 125.0 625.7±1.50 0.24
250.0 749.8±1.98 0.26
375.0 873.8±3.41 0.39
methyl paraben 200.0 50.00 248.4±0.95 0.38
100.0 297.1±0.98 0.33
150.0 348.7±1.15 0.33
propyl paraben 20.0 5.00 24.81±0.25 1.01
10.00 29.68±0.31 1.04
15.0 35.87±0.28 0.78

*Average value of three determinations, RSD is relative standard deviation

Table 5: Robustness parameters of LC method

Factor

Level

ketoprofen

methyl paraben

propyl paraben

Analytical wavelength

 

Assay, %

% RSD

Assay, %

% RSD

Assay, %

% RSD

238

-2

99.84

0.54

100.0

0.65

98.8

0.98

240

0

100.5

0.48

99.65

0.93

99.6

1.05

242

+2

99.21

0.97

99.48

0.92

99.1

0.87

Flow rate of mobile phase

 

 

 

 

 

 

 

0.8

-0.2

98.79

1.24

98.70

0.64

98.2

0.56

1.0

0

99.36

0.98

101.0

0.89

99.6

0.85

1.2

+0.2

99.25

0.97

99.82

0.97

99.8

1.21

Stationary phase

 

 

 

 

 

 

 

Nucleosil C 18

-

101.2

1.08

99.81

0.91

99.1

1.04

LiChrosorb C18

-

99.89

0.84

99.55

0.98

99.6

0.95


Conclusion

The validated RP-LC method developed here proved to be simple, specific, accurate, precise, sensitive and robust. It can successfully used for routine analysis of ketoprofen in the presence of described preservatives in gel formulation without any interference from common excipients.

CONFLICT OF INTERESTS

Declared None

REFERENCES

  1. Goodman, Gilman's. The pharmacological basis of therapeutics: Tenth edition, New York 2001;10:2045.
  2. Rajendra R, Chetan C, Rajesh N, Manohar D. Ketoprofen: In-vitro release and percutaneous absorption in rats through polymeric gels. Der Pharm Sin 2013;4:116-20.
  3. Blanko M, Coello J, Iturriaga H, Maspoch S, Perez-Maseda C. Chiral and nonchiral determination of ketoprofen in pharmaceuticals by capillary zone electrophoresis. J Chromatogr A 1998;799:301-7.
  4. Kormosh Z, Hunka I, Basel Y. Spectrophotometric determination of ketoprofen and its application in pharmaceutical analysis. Acta Pol Pharm-Drug Res 2009;66:3-9.
  5. El-Sadec M, El-Adi S, Abou-Kull M. Spectrophotometric determination of ketoprofen in pharmaceutical preparations by means of charge transfer complex formation. Talanta 1993;40:585-8.
  6. Wong C, Yeh M, Wang D. High-performance liquid chromatographic determination of ketoprofen in pharmaceutical dosage forms and plasma. J Liquid Chromatography 1992;15:1215-25.
  7. Bempong D, Bhattacharyya L. Development and validation of a stability-indicating high-performance liquid chromatographic assay for ketoprofen topical penetrating gel. J Chromatogr A 2005;1073:341-6.
  8. Zhuang Y, Cao D, Ge D. Flow injection analysis of ketoprofen based on the order transform second chemiluminiscence reaction. Spectrochim Acta part B 2012;85:139-44.
  9. Aboul-Enein H, Dal A, Tuncel M. A validated method development for ketoprofen by a flow-injection analysis with UV-detection and its application to pharmaceutical formulations. Il Pharm 2003;58:419-22.
  10. Amankwa L, Chatten L. Electrochemical reduction of ketoprofen and its determination in pharmaceutical dosage forms by differential-pulse polarography. Anal 1984;109:57-60.
  11. Kormosh Z, Hunka I, Basel Y, Matviychuk O. Potentiometric determination of ketoprofen and piroxicam at a new PVC electrode based on ion associates of Rhodamine 6G. Materials Sci Eng C 2010;30:997-1002.
  12. Overbeke A, Baeylus W, Van der Bossche W. Quantitative Fourier transform infrared attenuated total reflectance analysis of ketoprofen in some pharmaceutical formulations. Spectrosc 1995;9:121-30.
  13. European Pharmacopoeia, 5th ed. Council of Europe, Strasbourg; 2007.
  14. Soni M, Taylor S, Greenberg N, Burdock G. Evaluation of the health aspects of methyl paraben: a review of the published literature. Food Chem Toxicol 2002;40:1335-73.
  15. Wei G. Toxicity and estrogen effects of methyl paraben on drosophila melanodaster. Food Sci 2009;30:252-4.
  16. Cashman A, Warshaw E. Parabens: a review of epidemiology, structure, allergenicity, and hormonal properties. Dermatitis 2005;16:57-66.
  17. Shabir GA. A new validated HPLC method for the simultaneous determination of 2-phenoxyethanol, methylparaben, ethylparaben and propylparaben in a pharmaceutical gel. Indian J Pharm Sci 2010;72:421-5.
  18. Atemnkeng MA, Marchand E, Plaizier-Vercammen J. Assay of artemether, methylparaben and propylparaben in a formulated paediatric antimalarial dry suspension. J Pharm Biomed Anal 2007;43:727-32.
  19. Satinsky D, Huclova J, Ferreira RL, Montenegro MC, Solich P. Determination of ambroxol hydrochloride, methylparaben and benzoic acid in pharmaceutical preparations based on sequential injection technique coupled with monolithic column. J Pharm Biomed Anal 2006;40:287-93.
  20. Ali MS, Ghori M, Khatri AR. Stability indicating simultaneous determination of domperidone (DP), methylparaben (MP) and propylparaben by high performance liquid chromatography (HPLC). J Pharm Biomed Anal 2006;41:358-65.
  21. Matysova L, Hajkova R, Sicha J, Solich P. Determination of methylparaben, propylparaben, triamcinolone acetonide and its degradation product in a topical cream by RP-HPLC. Anal Biol Chem 2003;376:440-3.
  22. Kreuz DM, Howard AL, Ip D. Determination of indinavir, potassium sorbate, methylparaben, and propylparaben in aqueous pediatric suspensions. J Pharm Biomed Anal 1999;19:725-35.
  23. Hajkova R, Solich P, Dvorak J, Sicha J. Simultaneous determination of methylparaben, propylparaben, hydrocortisone acetate and its degradation products in a topical cream by RP-HPLC. J Pharm Biomed Anal 2003;32:921-7.
  24. Ali MS, Chaudhary RS, Takieddin MA. Simultaneous determination of metronidazole benzoate, methylparaben, and propylparaben by high performance liquid chromatography. Drug Dev Ind Pharm 1999;25:1143-7.
  25. Kokoletsi MX, Kafkala S, Tsiaganis M. A novel gradient HPLC method for simultaneous determination of ranitidine, methylparaben and propylparaben in oral liquid pharmaceutical formulation. J Pharm Biomed Anal 2005;38:763-7.
  26. Pongcharoenkiat N, Wittayanukulluk A, Hem SL. Determination of methylparaben in o/w emulsions by solid-phase extraction and high performance liquid chromatography. J Cosmet Sci 2003;54:47-52.
  27. Rebbeck C, Hammond R, Wong J, Nair L, Raghavan N. Solid phase extraction and HPLC analysis of methylparaben and propylparaben in a concentrated antibiotic suspension. Drug Dev Ind Pharm 2006;32:1095-102.
  28. Ma M, DiLollo A, Mercuri R, Lee T, Bundang M. HPLC and LCMS studies of the transesterification reaction of methylparaben with twelve 3-to 6-carbon sugar alcohols and propylene glycol and the isomerization of the reaction products by acyl migration. J Chromatogr Sci 2002;40:170-7.
  29. Mannucci C, Bertini J, Cocchini A, Salvagninip F, Triolo A. High performance liquid chromatography simultaneous quantitation of ketoprofen and parabens in a commercial gel formulation. J Liquid Chrom 1992;15:327-35.
  30. .Dvořák J, Hájková R, Matysová L, Nováková L, Koupparis MA, Solich P. Simultaneous HPLC determination of ketoprofen and its degradation products in the presence of preservatives in pharmaceuticals. J Pharm Biomed Anal 2004;36:625-9.
  31. Labbozzetta S, Valvo L, Bertocchi P, Alimonti S, Gaudiano M, Manna L. Focused Microwave-Assisted extraction and lc determination of ketoprofen in the presence of preservatives in a pharmaceutical cream formulation. Chromatographia 2009;69:365-8.
  32. Blanco M, Coello J, Iturriaga H, Maspoch S, Alaoui-Ismaili S. UV-Spectrophotometric determination of ketoprofen and paraben in a gel preparation by partial least-squares calibration. Fresenius' J Anal Chem 1997;357:967-72.
  33. Safra J, Pospisilova M. Separation and determination of ketoprofen, methylparaben and propylparaben in pharmaceutical preparation by micellar electrokinetic chromatography. J Pharm Biomed Anal 2008;48:452-5.