Int J Pharm Pharm Sci, Vol 8, Issue 1, 255-261Original Article


APPLICATION OFF-LINE SPE-HPLC/UV METHODS IN ANALYSIS OF OFLOXACIN IN HUMAN URINE (IN VITRO)

MUCHTARIDI MUCHTARIDI*, ESTER YULIANI, IYAN SOPYAN

1Department of Pharmaceutical Analysis and Medicinal Chemistry, 2Department of Pharmaceutical and Formulation Technology, Faculty of Pharmacy, Universitas Padjadjaran, JlKM 21.5 Bandung-Sumedang, Jatinangor
Email: muchtaridi@unpad.ac.id
 

 Received: 15 Jul 2015 Revised and Accepted: 25 Nov 2015


ABSTRACT

Objective: The objective of this study is to determine the validity of analytical methods in OFX antibiotic study in human urine (in vitro) using an SPE-HPLC/UV. In this study, SPE was applied in preparing the analysis of ofloxacin using HPLC embedded UV detector.

Methods: C-18 (octadecylsilane) cartridge (100 mg, particle size 10 µm) of SPE was employed in preparing a sample to determine further of analytes using HPLC with phosphate buffer 0.025 M (pH 2.5) and acetonitrile (85.5:14.5) as mobile phase and a flow rate of 1.2 m l/min. UV detector was adjusted at 295 nm with the internal standard ciprofloxacin.

Results: The calibration curves for the ofloxacin were linear over concentrations ranging from 1.15 to 36.0 µg/ml with a correlation coefficient (r) from 0.9998 to 0.9999. The coefficients of variation obtained from ofloxacin were less than 10 %. Ofloxacin on the area ratio of peak height and a segment of the chromatogram, LOD and LOQ of ofloxacin were 0.12 and 0.4 μg/ml, respectively. The recovery of ofloxacin from spiked human urine was 96.0 %.

Conclusion: The validation methods that including parameters: selectivity, repeatability, linearity, detection limit, quantification limit, precision, accuracy, and suitability of the system. The methods used have validity according to the requirements that might be used to analyze ofloxacin in human urine.

Keywords: Ofloxacin, Solid Phase Extraction, HPLC.


INTRODUCTION

Sample preparation is an important part of the analysis of drug in a material biologic sample which able to determine the efficacy of analysis because it can establish reproducibility and recovery of the matrix interference [1-4]. The parent metabolite compound (such as protein, carbohydrate, and lipid) should be reduced because the existence of a drug may give misinterpreting results in analysis methods [5]. Therefore, it is needed the method that could accurately identify both the parent drug and metabolites.

Application of SPE (Solid Phase Extraction) in sample preparation can reduce solvent volume and time [2, 6, 7]. In the previous research, application of SPE in the determination of lead compounds aromatherapy in blood plasma of mice after essential inhalation oil obtain good reliability (recovery 90 %) and reproducibility (variation coefficient less than 15 %).

Oflaxocin (OFX) (9-fluoro-2,3-dihydro-3-methyl-10-(-methyl-1-piperzinyl)-7-oxo-7H-pyrido-[1,2,3-de]1,4-bonzoxazine-6-carboxylic acid)as shown fig. 1 is which second generation quinolones are broad spectrum [8-10]. It has activity against both gram-positive and gram-negative bacteria [11]. The OFX inhibits DNA gyrase activity [12].

Fig.1:Structure of ofloxacin

The methods of determination of OFX in human plasma were published by many researchers [13-17] for bioavailability and bioequivalence study. Application of Solid Phase in preparation sample of OFX analysis in human urine has been reported [18-20].

Here, the preliminary study was carried out prior to the clinical study, using human urine. The simple and validated method for measuring the concentration of the OFX in spiked human urine was required in a clinical test.

The aim of this study is to develop a simple and reliable HPLC method for measurement of ofloxacin concentrations in spiked human urine with application SPE for preparation sample. The further recommendation will help in clinical study and routine analysis.

MATERIALS AND METHODS

Chemicals

OFX and ciprofloxacin were purchased from Zhejiang Jinxin, China. All chemicals were used as received without further purification and all solvents were of reagent grade: sodium dihydrogen phosphate monohydrate (Merck), acetonitrile, and phosphoric acid (Merck), methanol pa (Merck), aqua bidest (IPHA), human urine (volunteer, man 20-24 age).

Tools

HPLC (Shimadzu LC-10 ATVP) embedded with UV-VIS detector SPD, auto-injector Shimadzu system controller SCL-A, the HPLC column (Phenomenex); length of 250 mm, 4.6 mm internal diameter, particle size 10 µm, UV-Vis spectrophotometer (Analytical Jena, specord 200), pH meter (Oh meter), ultrasonic bath (Ney 1510), HLB 30 mg SPE cartridge 1 cc (Oasis), an analytical balance (Sartorius) sensitivity of 0.1 mg, filters vacuum with 0.4 to 0.45 µm pore filter, and an unusual glassware.

Methods

The mobile phase was a mixture of 0.025 M phosphate buffer pH 2.2 and acetonitrile (85:15). The mixture was filtered using 0.45 p. m millipore with vacuum assistance and ultrasonic bath for 15-20 min.

Standard solution preparation

OFX 100 mg dissolved in 200 ml measuring flask with mobile phase to achieve the final concentration of 0.5 mg/ml, diluted with mobile phase to obtain concentrations of 5 mg/ml. The in-scanning solution with a UV-spectrophotometer at a wavelength of 200-320 nm, so the obtained spectrum maximum wavelength (λ max) of absorption and OFX. The same procedure conducted on ciprofloxacin.

Determination of molar extinction

OFX standard solution with a concentration of 6.9, 13.5, and 18.0 pM measured at a wavelength of maximum absorbance OFX, and the calculated values molar extinction.

Optimization of HPLC conditions

OFX standard solution 0.1 mg/ml containing the internal standard ciprofloxacin 0.1 mg/ml was injected with 10 p. l (auto-injector) into the HPLC mobile phase composition of 85:15, 85,5:14,5, and 86: 14 v/v and flow rate was 1.2 and 1.3 ml/min. Viewed retention time and separation of the two peaks (OFX and ciprofloxacin) were produced.

Extraction by solid phase extraction (SPE)

The cartridge of SPE was conditioned by 1 ml of methanol and 1 ml aqua bidest with vacuum assistance. OFX was spiked into plasma with various concentration (0.10, 0.25, 1.00, 2.00, 3.00, 4.00, 5.00, and 6.00 µg/ml). Subsequently, 1 ml of plasma was put into a cartridge of SPE eluted by 1 ml acetonitrile 20% (in phosphate buffer) further injected into the HPLC. The efficiency of SPE extraction was calculated.

Method validation analysis

Selectivity was measured by looking at the chromatogram OFX and ciprofloxacin that obtained from HPLC separation, further calculated the value of the resolution. Repeatability was evaluated by generating a solution of OFX 0.25µg/ml in blood plasma and further extracted by using SPE. 10 µl of analyte was injected into the HPLC equipment in optimum condition; the experiment was repeated six times and then calculated the coefficient of variation. The linearity was determined by making the standard curve of five serial concentrations of OFX (0.10, 0.25, 2.00, 4.00, and 6.00 µg/ml) and the internal standard ciprofloxacin 3 µg/ml in human urine, further extracted using SPE.

HPLC system embedded with UV (294 nm) to be used as followed column C-18 (octadecyl silane), length of 250 mm, diameter in 4.6 mm, and the particle size of 10 im, mobile phase 0.025 M phosphate buffer pH 2.5 and acetonitrile with ratio of 85.5: 14.5 v/v, and flow rate 1.2 ml/min.

Precision, accuracy, and recovery

Calibration curve equation with the best correlation coefficient was used to specify the sample. LOD and LOQ were determined statistically from the calibration curve equation using linear regression. Accuracy and precision were obtained by making the sample solution OFX 1, 3, and 5 lag/ml and the internal standard ciprofloxacin 3 lag/ml in blood plasma was extracted using SPE. 10 µl of analyte injected into the HPLC equipment in optimum condition; the experiment was repeated three times and then calculated percent accuracy (recovery) and precision (coefficient of variation). System suitability test conducted on samples OFX 0.25 lag/ml and the internal standard ciprofloxacin 3µg/ml in blood plasma, and then extracted using SPE. Ten microliters of analyte were injected into the HPLC equipment in optimum condition, done six times a repetition then calculated the coefficient of variation of retention time, area ratio, and peak height ratio chromatogram.

RESULTS AND DISCUSSION

The determination of maximum wavelength (λ max)

The result of scanning using UV at the wavelength of 200-380 nm of OFX solutions in the mobile phase (phosphate buffer pH 2.5 and acetonitrile with a ratio of 85.5: 14.5) showed maximum absorption of OFX at λ max of 294 nm. This result was in line with a previous study [21]. Subsequently, the spectrum absorption of ciprofloxacin as a standard was observed. It showed a quite similar λ max to OFX (277 nm). However, the previous study obtained that the λ max of OFX is 275 nm [22] and 278 nm [23].

On the other hand, the result of the combined spectra of both compounds showed the point of intersection at a wavelength of 286.2 nm, as shown in fig. 1.

4

Fig.1:Spectrum of ciprofloxacin (blue line) was 277 nm dan OFX (green line) was 294 nm

Ciprofloxacin was used as an internal standard due to it has identical chemical structure and properties OFX thus it could be eluted as OFX. The purpose of using internal standard is to reduce errors during the analysis process, particularly, for samples undergoing pre- treatment's, such as extraction, and filtration [24]. y applying the internal standard technique, it was expected to produce sensitive, relatively fast, and accurate method for the analysis of OFX in biological  the internal standard technique.

The λ max of OFX was set as the wavelength used in the detection of the analysis result by HPLC, as OFX was the compound of target analysis, ciprofloxacin then provided large absorption at the λ max of OFX.

The determination result of molar extinction value of OFX

The determination of molar extinction value has been conducted to obtain the sensitivity value of OFX. It could be calculated by comparing the absorptivity value or of molar OFX absorptivity towards the thickness of cuvette (usually 1 cm), with the OFX concentration measured[25]. The extinction value of molar OFX from three varied concentrations was 6.9; 13.5; and 18 μM in 0.1 M HCl in a row. It showed that the molar extinction value ​​had an average value of 3.2 104M-1 cm-1. This molar extinction value was ​​greater than 10.000 M-1 cm-1, indicating that OFX was possible to detect the ultraviolet detector on the HPLC system. The extinction value of OFX could be seen in table 1. The other study measured the extinction value of OFX in 0.1 M HCl at 293 nm and produced 3.5. 104 M-1 cm-1 [26]. However, this result was determined with different diluted concentration.

Table 1: The extinction molar (e) value of OFX

Extinction molar data OFX in mobile phase at 295 nm

No.

Molarity (M)

Absorbance

Extinction molar e (M-1 cm-1)

1

2

3

0.0000069

0.0000135

0.0000180

0.2231

0.4734

0.5817

3.2.104

3.5.104

3.2. 104

Jumlah

   

9.9.104

45

   

3,3.104

* Mobile phase = Buffer phosphat: acetonitril (85: 15)


Optimization result of HPLC condition

Optimization result of HPLC condition has been employed chromatography parameters, including the retention time, resolution or separation (Rs), efficiency (N) and column efficiency (HETP) from various compositions. The flow rate of the mobile phase was presented in table 2. The main priority in selecting the method was the resolution value result ≥ 1.5[27]. Based on the Rs value ≥ 1.5, it showed that the two peaks were completely separated. The second priority was the retention time, meaning that the faster retention time was better because the analysis time required will be faster. Efficiency (N) ≥ 2500 showed that the peaks produced were sharp. The theoretical chip value would increase by lowering the flow rate of the mobile phase or by increasing the length of the column, but the analysis time would remain longer.

The mobile phase composed 0.025 M phosphate buffer pH 2.5 and acetonitrile (85: 15) with a flow rate of 1.2 ml/min, it obtained the retention time of OFX in 11.767 min. This condition was in a good state, but there was a disruption of urine that used to be separated in the early minutes of separation. Therefore, it was necessary to examine another mobile phase condition to slow the retention time of OFX.

Furthermore, the observation continued to try a mobile phase composed of 0.025 M phosphate buffer pH 2.5 and acetonitrile (85.5: 14.5) with the flow rate of 1.2 ml/min. The retention time of OFX obtained was 12.533 min while; at the flow rate of 1.3 ml/min, the retention time of OFX obtained was 11.375 min. The retention time of OFX obtained with the mobile phase composed of 0.025 M phosphate buffer pH 2.5 and acetonitrile (86: 14) with a flow rate of 1.2 ml/min is 14,250 min, while at the flow rate of 1.3 ml/min, the retention time of OFX obtained is 12.750 min. The mobile phase composed of 0.025 M phosphate buffer pH 2.5 and acetonitrile (85.5: 14.5) was chosen, since it produces a good resolution, 1,77 (≥ 1.5). The flow rate of 1.2 ml/min was chosen to optimize the analysis condition because the other slower flow rate would take longer time. The previous study, the mobile phase used in OFX analysis was sodium lauryl sulfate (0.024% aqueous solution)-acetonitrile-glacial acetic acid (500:480:20) [28] or water-acetonitrile-triethylamine (83:14:0.45, v/v, pH 2.30) [29].

The number of theoretical chips (N) in each condition was ≥ 2,500. It indicated that the peak produced was quite sharp [30]. In addition, the number of theoretical chips could be used to determine the quality and performance of the column. The theoretical chip value would rose by lowering the flow rate of the mobile phase or by increasing the length of the column, thereby taking the analysis time much longer.

The result of condition optimization using solid phase extraction (SPE)

The first step of extraction using SPE was conditioned with adding 1 ml of methanol and 1 ml aqua bidest to clean impurities (exposure) in SPE cartridge during storage and also to wet the SPE cartridge [31]. Urine sample spiked by OFX was put in SPE cartridge as much as 1 ml (Sample Loading). The addition of sample has been carried out by drop wise while being energized by negative air pressure using a vacuum to speed up the extraction process and prevent clogged SPE cartridge.

The washing process in SPE phase has been done by adding 1 ml of methanol 3% (in aqua bidest) to clean impurities (endogenous substances) in the urine and conserve the peaks of OFX and ciprofloxacin when analyzed by HPLC. Elution process has been conducted by adding 1 ml of acetonitrile 20% (in 0.025 M phosphate buffer pH 2.5). It was expected that the OFX and ciprofloxacin remained in the SPE cartridge could be eluted or completely pushed out. The output analyte was accommodated in a container, to be analyzed by HPLC.

Elution process was utilizing acetonitrile 20% in 0.025 M phosphate buffer pH 2.5 was sufficient to elute OFX and ciprofloxacin from SPE cartridges. It could be seen from the generated value of the extraction efficiency, which is ≥ 85% for two concentrations, respectively 1.8 and 36 μg/ml. Organic solvents with stronger elution could be used to obtain a better elution result, otherwise using the composition of acetonitrile that is greater than 20%. Due to the composition of the elution containing phosphate buffer pH 2.5 and acetonitrile is in accordance with the composition of the mobile phase of the HPLC system used, the elution result could be directly injected into the HPLC system.

Table 2: Recovery of extraction of OFX 1.8 and 36 mg/ml with internal standard of ciprofloxacin 10 μg/ml based on to area under curve of chromatogram (n = 3)

Recovery of OFX (%)

Recovery of ciprofloxacin (%)

Replication

OFX (μg/ml)

Replication

ciprofloxacin (μg/ml)

1.8

36

1.8

36

1

2

3

96.70

101.41

101.19

99.09

99.49

99.21

1

2

3

92.94

96.26

84.06

99.50

91.02

96.75

5

99.77

99.27

6

91.09

95.76

CV

0.0217

0.0017

CV

0.0565

0.0369

To determine the recovery efficiency of the extraction result, the evaluation of extraction efficiency was conducted towards two samples with the concentrations of 1.8 and 36 mg/ml. After that, the injection result in the form of chromatogram area and peak chromatogram height of each sample in the mobile phase without the SPE between the SPE phases in the urine was compared. In table 2, it could be seen that SPE result of the low concentration is 1.8 mg/ml based on the chromatogram area provides 99.77% of the extraction efficiency of OFX and 91.09% of ciprofloxacin while the high concentration of 36 μg/ml is 99.27% ​​and 95.76% respectively.

Extraction result using the SPE as a whole for μg/ml was very satisfying, because of the extraction efficiency is ≥ 85% and the value has met the extraction efficiency requirement>80%[32]. Chromatogram extraction recovery could be seen in fig. 2.


77

Fig.2: Chromatogram of recoveryextraction OFX 36 µg/ml, (purple) in moble phase and (blue) in urine

Selectivity

Selectivity test has been done by observing the resolution indicated at the peak of Rt = 9.867 min for OFX that was perfect apart from the top of ciprofloxacin peak Rt = 11.100 min, with a resolution value of 1.7 exceeding the recommended requirement for a good resolution value, which is ≥ 1.5. According to the resolution value, the HPLC method used had a good selectivity, and could be used to determine the OFX levels by using internal ciprofloxacin standard.

Repeatability tests result

Repeatability test was applied to the OFX sample with a concentration of 1.8 μg/ml, using ciprofloxacin of 10 μg/ml as an internal standard. The replication was performed 6 times, while the retention time, chromatogram area, and high peak chromatogram are observed afterward. The average retention time for OFX was 10.07 min and CV value being<0.67. It indicated a good repeatability value, whereas the required value is<20% for biological fluid sample analysis [27]. The results were presented in table 3.

Table 3: Repeatability analysis results of OFX 1,8 μg/ml with Internal Standard of Ciprofloxacin 10 μg/ml in Urine (n = 6) based on retention time

OFX

(μg/ml)

Retention time (min)

OFX

Ciprofloxacin

Ratio

1.8

1.8

1.8

1.8

1.8

1.8

10.042

10.025

10.200

10.025

10.092

10.058

11.300

11.292

11.483

11.267

11.375

11.325

0.8886726

0.8877967

0.8882696

0.8897666

0.8872088

0.8881236

1

10.07

11.34

0.8883063

CV

0.66267

0.69581

0.09769


Table 4: Repeatability analysis results of OFX 1,8 μg/ml with internal standard of ciprofloxacin 10 μg/ml based on area under curve

OFX

(μg/ml)

Area under curve

OFX

Ciprofloxacin

Ratio

1.8

1.8

1.8

1.8

1.8

1.8

97501

70535

77786

80065

80121

79496

149397

145583

149236

152391

146883

147747

0.652630

0.484500

0.521228

0.525392

0.545475

0.538055

2

80917.33

148539.5

0.54454673

CV

0.10998

0.01598

0.10466

The repeatability results were good for the CV was<0.11 (table 4), while the required value is<0.2 for the analysis of biological fluid sample[33]. Likewise, the result of repeatability based on chromatogram peak height ratio indicated a good value with a coefficient of variation (CV)<0.05 (table 4).

In particular literature, strict criteria were obtained if the method provides relative standard deviation, or 2% or less of CV. However, this value was very flexible depending on the concentration of the analyte analyzed, the number of samples, and laboratory condition. The variation coefficient would increase as the analyte level analyzed decrease [27, 33, 34]. The data proved that the method used has a good repeatability, with CV value ​​<20% for the analysis of biological fluid sample[27, 34].

Linearity result

Linearity test was conducted to observe the capability of analytical method in giving a good response to various analyte concentrations on a calibration curve in order to produce a straight line. The parameter concerning linear relationship was expressed by the correlation coefficient and a valid analytical method which has a correlation coefficient more than 0.998[33].

3

Fig. 3: Calibration curve of OFX with ranged concentration (1,8-36 μg/ml) that obtained internal standard of ciprofloxacin 10 μg/ml based on ratio of Area Under Curveof chromatogram

OFX extraction result of ciprofloxacin in the urine was ranged from 1.8-36 mg/ml obtaining a linear calibration curve with the line equation y = 0,2172x+0.2656 and the correlation coefficient (R) = 0.9982, containing ciprofloxacin internal standard 10 μg/ml based on the ratio of chromatogram area (fig. 3).

Limit of detection (LOD) and limit of quantitation (LOQ)

The absolute limit of detection (LOD) was determined when the concentration of the analyte analyzed was relatively small as in biological matrix [30, 35]. The result of the limit of detection (LOD) test was calculated based on the calibration curve from an equation with the best correlation coefficient (r). The LOD value was depending on the calibration curve of OFX towards the ratio of the chromatogram area[29]. The LOD value of the area ratio was 0.12 mg/ml.

The absolute limit of quantitation (LOQ) is determined when the concentration of the analyte analyzed relatively small as in biological matrix [30, 35]. The result of the limit of quantitation (LOQ) test was calculated based on the calibration curve of OFX in accordance with an equation that had the best correlation coefficient (r). LOQ value was determined from the calibration curve of OFX towards the ratio of chromatogram area. LOQ value of the area ratio was 0.4 mg/ml.

Precision and accuracy

According to the calculation of the sample levels based on chromatogram area, it provided precision values ​​expressed as CV from concentrations of 1.8 CV; 9 and 36 mg/ml, 0.16; 0,025; and 0.038, respectively. The value was fair as required (<20%) for the analysis of biological fluid sample [30]. While the accuracy values in % sample recovery with concentrations of 1.8; 9 and 36 mg/ml were 80.09%; 109.89% and 90.01%, respectively. The values ​​obtained were fairly good as required as 80-110% [33].

4

Fig.4:Chromatogram in system suitability of OFX 1.8 μg/ml with internal standard ciprofloxacin 10 μg/ml in human urine


Table 5:System suitability analysis of OFX in concentration 1.8 μg/ml with internal standard of ciprofloxacin 10 μg/ml in human urine (n = 6)

Parameters

OFXIn 1.8 μg/ml

Variation coefficient (CV)

Retention time

OFX

Ciprofloxacin

0,66267

0,69581

Ratio

0,09769

Area under curve

OFX

Ciprofloxacin

0,10998

0,01598

Ratio

0,10466

Peak height

OFX

Ciprofloxacin

0,04592

0,00088

Ratio

0,04492

Asymmetry

OFX

Ciprofloxacin

1,06-1,37

1,02-1,11

Tailing factor

OFX

Ciprofloxacin

1,02-1,16

1,02-1,05


System suitability

Table 5 showed that the Consistent Variation (CV) from the retention time, the ratio of ​​the chromatogram area, and the chromatogram peak height ratio ≤ 10% for the analysis of the biological fluid sample. It indicated that the method used had good system suitability.

The suitability of the system was also used to determine the asymmetry factor and tailing factor of the peak. The asymmetry and tailing factors were used to determine the column condition and the experimental condition. The OFX peak of asymmetry with a concentration of 1.8 μg/ml was equal to 0.91-1.04 while the ciprofloxacin was equal to 1.02-1.11. The asymmetry value has met the criteria of the<2[27]. The tailing factor value for the peak of OFX with a concentration of 1.8 μg/ml was equal to 1.02 to 1.16 and for ciprofloxacin was equal to 1.02 to 1.05. The tailing factors obtained have met the requirements value, which was < 2 (the chromatogram could be seen in fig. 4).

CONCLUSION

OFX in human urine could be extracted by using SPE Oasis HLB 1 thus further analysis by HPLC embedded UV detector with recovery more than 96 %. The validation methods that including parameters: selectivity, repeatability, linearity, detection limit, quantification limit, precision, accuracy, and suitability of the system, the methods used were valid according to the requirements that might be used to analyze OFX in human urine.

ABBREVIATION

HPLC/UV: High-Performance Liquid Chromatography-Ultra Violet, SPE: Solid Phase Extraction, LOD: Limit of Detection, LOQ: Limit of Quantification, OFX: ofloxacin.

CONFLICT OF INTERESTS

Declared none

REFERENCES

  1. Deng C, Liu N, Gao M, Zhang X. Recent developments in sample preparation techniques for chromatography analysis of traditional Chinese medicines. J Chromatogr A 2007;1153:90-6.
  2. Saari-Nordhaus R; Nair LM, Anderson Jr, JM. Elimination of matrix interferences in ion chromatography by the use of solid-phase extraction disks. J Chromatogr A 1994;671:159-63.
  3. Montgomery RM, Saari-Nordhaus R, Nair LM, Anderson JJM. On-line sample preparation techniques for ion chromatography. J Chromatogr A 1998;804:55-62.
  4. Muchtaridi M, Musfiroh I, Rambia I, Suganda H, Nasrudin E. Chemical composition and locomotors activity of essential oils from the rhizome, Stem, and leaf of Alpinia malaccencis (Burm F.) of Indonesian Spices. J Appl Pharm Sci 2014;04:52-6.
  5. McDowall RD, Doyle E, Murkitt GS, Picot VS. Sample preparation for the HPLC analysis of drugs in biological fluids. J Pharm Biomed Anal 1989;7:1087-96.
  6. Henderson IK, Saari-Nordhaus R, Anderson Jr JM. Sample preparation for ion chromatography by solid-phase extraction. J Chromatogr 1991;546:61-71.
  7. Muchtaridi M, Musfiroh I. Off-Line SPE-GC/MS analysis of lead compounds aromatherapy in blood plasma of mice of essential oils materials from Indonesian aromatic plants. Asian J Chem 2012;24:5124-8.
  8. Sultana N, Arayne MS, Yasmeen N. In vitro availability of ofloxacin in the presence of metals essential to human body. Pak J Pharm Sci 2007;20:42-7.
  9. Reddy BV, Kumar V, Chandra SR, Chandra AS, Babu GD, Prakash C. Preparation and in-vitro evaluation of ofloxacin mucoadhesive microspheres. Int J Pharm Pharm Sci 2012;4:93-6.
  10. Zacharia S, Peter A, Mathew J. Ntibiotic resistance profile of bacterial pathogens in the gut of P. Americana. Asian J Pharm Clin Res 2013;6:42-6.
  11. Drew RH, Gallis HA. Ofloxacin: its pharmacology, pharmacokinetics, and the potential for clinical application. Pharmacother 1988;8:35-46.
  12. Imamura M, Shibamura S, Hayakawa I, Osada Y. Inhibition of DNA gyrase by optically active ofloxacin. Antimicrob Agents Chemother 1987;31:325-7.
  13. Gao XX, Yao GC, Guo N, An F, Guo XJ. A simple and rapid high-performance liquid chromatography method to determine levofloxacin in human plasma and its use in a bioequivalence study. Drug Discoveries Ther 2007;1:136-40.
  14. Zhou ZL, Yang M, Yu XY, Peng HY, Shan ZX, Chen SZ, et al. A rapid and simple high-performance liquid chromatography method for the determination of human plasma levofloxacin concentration and its application to bioequivalence studies. Biomed Chromatogr 2007;21:1045-51.
  15. Ji HY, Jeong DW, Kim YH, Kim HH, Sohn DR, Lee HS. Hydrophilic interaction liquid chromatography-tandem mass spectrometry for the determination of levofloxacin in human plasma. J Pharm Biomed Anal 2006;41:622-7.
  16. Kumara TM, Srikanthb G, Venkateshwar Raoa J, Sambasiva Rao K. Development and validation of HPLC-UV method for the estimation of levofloxacin in human plasma. Int J Pharm Pharm Sci 2011;3:247-50.
  17. Watabe S, Yokoyama Y, Nakazawa K, Shinozaki K, Hiraoka R, Takeshita K, et al Simultaneous measurement of pazufloxacin, ciprofloxacin, and levofloxacin in human serum by high-performance liquid chromatography with fluorescence detection. J Chromatogr B 2010;878:1555-61.
  18. Ballesteros O, Vilchez JL, Navalon A. Determination of the antibacterial ofloxacin in human urine and serum samples by solid-phase spectro fluorimetry. J Pharm Biomed Anal 2002;30:1103-10.
  19. Okazaki O, Aoki H, Hakusui H. High-performance liquid chromatographic determination of (S)-(−)-ofloxacin and its metabolites in serum and urine using a solid-phase clean-up. J Chromatogr B: Biomed Sci Appl 1991;563:313-22.
  20. Khandagle KS, Gandhi SV, Deshpande PB, Gaikwad NV. A simple and sensitive RPHPLC method for simultaneous estimation of cefixime and ofloxacin in combined tablet dosage form. Int J Pharm Pharm Sci 2011;3:46-8.
  21. Patel DM, Soneji JA, Patel PB, Patel CN. Development and validation of a method for simultaneous estimation of ofloxacin and ornidazole in different dissolution media. Pharm Methods 2012;3:102-5.
  22. Liu CG, Xu YZ, Wei YJ, Zhao J, Qi J, Wang XH, et al. Spectral properties, protonation and fluorescence quantum yield of ciprofloxacin. Guang Pu Xue Yu Guang Pu Fenxi 2005;25:1446-50.
  23. Kontou P, Chatzika K, Pitsiou G, Stanopoulos I, Argyropoulou-Pataka P, Kioumis I. Pharmacokinetics of ciprofloxacin and its penetration into bronchial secretions of mechanically ventilated patients with the chronic obstructive pulmonary disease. Antimicrob Agents Chemother 2011;55:4149-53.
  24. Grasshoff C, Thiermann H, Gillessen T, Zilker T, Szinicz L. Internal standard high-performance liquid chromatography method for the determination of obidoxime in urine of organophosphate-poisoned patients. J Chromatogr B: Biomed Sci Appl 2001;753:203-8.
  25. Denney RC. Dictionary of Spectroscopy. Wiley: New York; 1982. p. 2.
  26. Vinay KB, Revanasiddappa HD, Divya MR, Rajendraprasad N. Spectrophotometric determination of ofloxacin in pharmaceuticals and human urine. Eclética Química 2009;34:65-78.
  27. Lister AS. 7 Validation of HPLC methods in pharmaceutical analysis. In: Separation Science and Technology. Satinder A, Michael WD. Eds. Academic Press: New York; 2005;6:191-217.
  28. Arayne MS, Sultana N, Sajid SS, Ali SS. Cleaning validation of ofloxacin on pharmaceutical manufacturing equipment and validation of desired HPLC method. PDA J Pharm Sci Technol 2008;62:353-61.
  29. Maraschiello C, Cusido E, Abellan M, Vilageliu J. Validation of an analytical procedure for the determination of the fluoroquinolone ofloxacin in chicken tissues. J Chromatogr B: Biomed Sci Appl 2001;754:311-8.
  30. Kazusaki M, Ueda S, Takeuchi N, Ohgami Y. Validation of analytical procedures by high−performance liquid chromatography for pharmaceutical analysis. Chromatography 2012;33:65-73.
  31. Hennion MC. Solid-phase extraction: method development, sorbents, and coupling with liquid chromatography. J Chromatogr A 1999;856:53-4.
  32. Caufield WV, Stewart JT. Determination of zidovudine and levofloxacin in human plasma by reversed-phase HPLC and solid phase extraction. J Liq Chromatogr Relat Technol 2002; 25:1791-05.
  33. Guideline IHT. In: Validation of Analytical Procedures: Text and Methodology, Q2 (R1), Complementary Guideline on Methodology, International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use, London; 2005.
  34. Épshtein NA. Structure of chemical compounds, methods of analysis and process control. Pharm Chem J 2004;38:212-27.
  35. Chatzimichalakis PF, Samanidou VF, Verpoorte R, Papadoyannis IN. Development of a validated HPLC method for the determination of B-complex vitamins in pharmaceuticals and biological fluids after solid phase extraction. J Sep Sci 2004;27:1181-8.