Int J Pharm Pharm Sci, Vol 7, Issue 11, 13-16Original Article


RATIO SPECTRA DERIVATIVE UV SPECTROPHOTOMETRIC METHOD FOR SIMULTANEOUS ESTIMATION OF KETOROLAC TROMETHAMINE AND PHENYLEPHRINE HYDROCHLORIDE IN IMMEDIATE RELEASE TABLET

RIDDHI PARMARa*, FALGUNI TANDELa, NIKITA PATELa

aDepartment of Quality Assurance, Parul Institute of Pharmacy, Limda, Waghodia, Vadodara
Email: riddhi.parmar91@yahoo.com  

 Received: 08 Feb 2015 Revised and Accepted: 10 Sep 2015


ABSTRACT

Objective: To develop and validate robust, accurate and precise UV spectrophotometric method for determination of Ketorolac Tromethamine and Phenylephrine Hydrochloride.

Methods: Ratio spectra derivative method was developed using water as solvent. The developed method was validated as per International Conference on Harmonization (ICH) guidelines.

Results: Linearity of the developed method was 0.9958 and 0.9987 in the range of 4-20 ppm and 12-60 ppm for Ketorolac Tromethamine and Phenylephrine Hydrochloride respectively. % Relative standard deviation (RSD) was found to be less than 2 for Intraday precision and Intermediate precision. % recovery was found to be 98.5–100.27 % and 98.38–101.99 % for Ketorolac Tromethamine and Phenylephrine Hydrochloride respectively.

Conclusion: A robust, accurate and precise UV spectrophotometric method was developed and validated as per ICH guidelines.

Keywords: Ketorolac Tromethamine, Phenylephrine Hydrochloride, UV spectroscopy, Ratio spectra derivative method.


INTRODUCTION

Ketorolac Tromethamine, chemically known as 5-benzoyl-2,3-dihydro-1H-pyrrolizine-1-carboxylic acid with 2-amino-2-(hydroxymethyl) propane-1,3-diol (fig. 1), is a Non-steroidal anti-inflammatory drug. Ketorolac Tromethamine is used as an analgesic and antipyretic agent. It is freely soluble in water and methanol.

Ketorolac Tromethamine is official in IP, BP and USP and estimated by Potentiometric Titration as per IP [1] and BP [2] and HPLC as per USP [3]. Literature survey showed several reported spectro-photometric [4, 5] and HPLC [6, 7] methods for determination of Ketorolac Tromethamine in pharmaceutical dosage form either as the single component or in combination with other drugs.

Phenylephrine Hydrochloride, chemically known as R)-1-(3-Hydroxy phenyl)-2-methyl amino ethanol hydrochloride (fig. 2), is a sympathomimetic drug. Phenylephrine Hydrochloride is used as an anti-allergic agent. It is freely soluble in water and methanol [8].Phenylephrine Hydrochloride is official in IP, BP and USP and estimated by potentiometric titration as per IP [9] and BP [10] and Titration as per USP [11].

Fig. 1: Structure of ketorolac tromethamine

Fig. 2: Structure of phenylephrine hydrochloride

Literature survey showed several reported spectrophotometric [12, 13] and HPLC [14, 15] methods for determination of Phenylephrine Hydrochloride in pharmaceutical dosage form either as the single component or in combination with other drugs [16, 17]. There are no UV methods available for simultaneous estimation of Ketorolac Tromethamine and Phenylephrine Hydrochloride in combination.

So the rational this works to develop robust, accurate and precise UV method for simultaneous estimation of Ketorolac Tromethamine and Phenylephrine Hydrochloride in immediate release tablet.

MATERIALS AND METHODS

Materials

Instrument

UV visible spectrophotometer (Shimadzu, model 1800) with UV probe software was used for recording the spectra and measuring the absorbance.

Chemicals and reagents

Ketorolac Tromethamine was supplied by the Cadila healthcare, Ahmedabad as the gift sample. Phenylephrine Hydrochloride was supplied by the Tuton Pharmaceuticals, Mumbai as the gift sample. For the preparation of solution double distill water was used.

Methods

Selection of solvent

Double distil water was used as a solvent for the preparation of the sample solution.

Selection of method

Standard solution of Ketorolac Tromethamine (4.2 µg/ml) and Phenylephrine Hydrochloride (12.6 µg/ml) was scanned separately in the range of 200-400 nm. From the overlay spectra of Ketorolac Tromethamine and Phenylephrine Hydrochloride, Absorbance correction method, Absorbance ratio and derivative spectroscopic method are possible. But in all these methods % recovery was found out of the limit. So that Ratio spectra derivative spectroscopic method was developed.

Fig. 3: Overlay spectra of ketorolac tromethamine and phenylephrine hydrochloride

Ratio spectra derivative spectroscopic method

Selection of wavelength

4µg/ml of Ketorolac Tromethamine and 12µg/ml Phenylephrine Hydrochloride were scanned separately in the range of 200-400 nm. For this method development, Ketorolac Tromethamine (12 µg/ml) and Phenylephrine Hydrochloride (36 µg/ml) were used as devisor and first derivative spectra were traced using 1 scaling factor. From the first derivative ratio spectra, 290 nm selected for the Ketorolac Tromethamine (fig. 4) and 227 nm selected for the Phenylephrine Hydrochloride (fig. 5).

Preparation of stock solution for ketorolac tromethamine
(100 µg/ml)

A stock solution of 100 µg/ml of Ketorolac Tromethamine was prepared by accurately weighing 10 mg of Ketorolac Tromethamine and transferred in 100 ml volumetric flasks. 50 ml of water was added in the volumetric flask and the drug was dissolved properly and then the volume was made up to 100 ml with water.

Preparation of stock solution for phenylephrine hydrochloride (100 µg/ml)

A stock solution of 100 µg/ml of Phenylephrine Hydrochloride was prepared by accurately weighing 10 mg of Phenylephrine Hydrochloride and transferred in a 100 ml volumetric flask. 50 ml of water was added in the volumetric flask and the drug was dissolved properly and then the volume was made up to 100 ml with water.

Preparation of sample solution

Twenty tablets were weighed; average weight was determined and finely powdered. An accurately weighed quantity of tablet powder equivalent to 4.2 mg Ketorolac Tromethamine and 12.6 mg Phenylephrine Hydrochloride was transferred to 100 ml volumetric flask and 50 ml of water was added and dissolved by sonication for 30 min and then volume up was made up to the mark with water. Then solution is filtered through whatman filter paper. 1 ml of filtrate was taken in a 10 ml of volumetric flask and volume was made up to mark with water. The above mixture was analysed in UV spectrophotometer. % assay was calculated. Results are reported in table 6.

RESULTS AND DISCUSSION

Selection of wavelength

4µg/ml of Ketorolac Tromethamine and 12µg/ml Phenylephrine Hydrochloride were scanned separately in the range of 200-400 nm. For this method development Ketorolac Tromethamine (12 µg/ml) and Phenylephrine Hydrochloride (36 µg/ml) were used as devisor and first derivative spectra were traced using 1 scaling factor. From the first derivative ratio spectra, wavelength selected were 290 nm for Ketorolac Tromethamine and 227 nm for Phenylephrine Hydrochloride.

Fig. 4: First derivative ratio spectra of Phenylephrine Hydrochloride (12µg/ml), divisor Ketorolac Tromethamine (12µg/ml)

Fig. 5: First derivative ratio spectra of ketorolac tromethamine (4 µg/ml), divisor phenylephrine hydrochloride (36 µg/ml)

Validation of ratio spectra derivative spectroscopic method

Linearity and range

Linearity study was carried out for both the drugs at different concentration levels. Linearity of Ketorolac Tromethamine and Phenylephrine Hydrochloride was in the range of 4-20 µg/ml and 12-60 µg/ml and results are reported in following table.

Fig. 6: Calibration curve for ketorolac tromethamine at 290 nm

Fig. 7: Calibration curve for phenylephrine hydrochloride at 227 nm

Precision

Intraday precision

Three replicates of three concentration of standard solution of Ketorolac Tromethamine (8, 12 and 16 µg/ml) and Phenylephrine Hydrochloride (24, 36 and 48 µg/ml), total nine determination were analysed at three consecutive times on same day and absorbance was measured. % RSD was found to be within acceptable limit that is<2 %. The data for intraday precision of are reported in table 2.

Intermediate precision

Inter day precision

Three replicates of three concentration of standard solution of Ketorolac Tromethamine (8, 12 and 16 µg/ml) and Phenylephrine Hydrochloride (24, 36 and 48 µg/ml), total nine determination were analysed at three consecutive day and absorbance was measured. % RSD was found to be within acceptable limit that is<2 %. The data for inter day precision of are reported in table 2.

Different instruments

Three concentration of standard solution of Ketorolac Tromethamine (8, 12 and 16 µg/ml) and Phenylephrine Hydrochloride (24, 36 and 48 µg/ml) were analysed by two different instruments (UV-1800 and UV-1700) on a single day and absorbance was measured. % RSD was found to be within acceptable limit that is<2 %. The data are reported in table 2.

Different analysts

Three concentrations of the standard solution of Ketorolac Tromethamine (8, 12 and 16 µg/ml) and Phenylephrine Hydrochloride (24, 36 and 48 µg/ml) were analysed by three different analysts and absorbance was measured. % RSD was found to be within acceptable limit that is<2 %. The data for different analyst are reported in table 2.

Table1: Data for linearity and range

Ketorolac tromethamine

Phenylephrine hydrochloride

Concentration (µg/ml)

Response at 290 nm (n=5)

Concentration (µg/ml)

Response at 227 nm (n=5)

4

0.330

12

-0.131

8

0.632

24

-0.253

12

0.812

36

-0.386

16

1.075

48

-0.491

20

1.337

60

-0.629

Regression equation

y = 0.0614x+0.1005

Regression equation

y =-0.0103x-0.0078

R2 value

0.9958

R2value

0.9987


Table 2: Precision data for ketorolac tromethamine and phenylephrine hydrochloride

Precision

Concentration (µg/ml)

% RSD

Ketorolac Tromethamine

Phenylephrine Hydrochloride

Ketorolac Tromethamine

Phenylephrine Hydrochloride

Intraday precision (n=3)

8

24

0.34

0.37

12

36

0.09

0.52

16

48

0.11

1.00

Inter day precision (n=3)

8

24

0.78

0.81

12

36

1.20

0.80

16

48

0.16

1.28

Different instrument precision (n=2)

8

24

0.55

0.78

12

36

0.74

0.64

16

48

0.42

1.00

Different analyst precision (n=3)

8

24

0.94

0.67

12

36

1.16

1.06

16

48

0.33

0.58


Table 3: Accuracy data for ketorolac tromethamine and phenylephrine hydrochloride

Level

% Recovery

Ketorolac tromethamine

Phenylephrine hydrochloride

80 % (n=3)

100.17

99.51

100 % (n=3)

98.87

100.37

120 % (n=3)

100.20

100.50


Accuracy

Accuracy study was performed using the placebo. The accuracy of the method was tested by 3 replicate analyses of 3 different levels (80%, 100% and 120%). Accuracy was determined by % recovery study. % recovery was found to be 98.19–100.71 % and 98.50–101.72% for Ketorolac Tromethamine and Phenylephrine Hydrochloride respectively. The data of accuracy are reported in following table.

Limit of detection (LOD) and limit of quantitation (LOQ)

LOD and LOQ were determined by repeating the calibration curve five times. The results of LOD and LOQ are reported in following table.

Table 4: Data of LOD and LOQ

Parameter

Ketorolac tromethamine

Phenylephrine hydrochloride

S. D. of five intercept

0.01703

0.00133

Mean of five slope

0.215

0.0102

LOD(µg/ml)

0.261

0.433

LOQ(µg/ml)

0.792

1.30


Table 5: Robustness data for ketorolac tromethamine and phenylephrine hydrochloride

Drug

Concentration (µg/ml)

% RSD

Ketorolac tromethamine

8

0.91

12

1.57

16

1.36

Phenylephrine hydrochloride

24

1.16

36

1.27

48

1.24


Table 6: Data for analysis of IR tablet by UV method

Tablet

Dose

Amount found (mg) (mean±SD) (n=3)

% amount found (mean±SD) (n=3)

Ketorolac tromethamine

4.2 mg

4.21±0.02

100.36±0.59

Phenylephrine hydrochloride

12.6 mg

12.73±0.19

    1. ±1.51

Robustness

Three concentrations of the standard solution of Ketorolac Tromethamine (8, 12 and 16 µg/ml) and Phenylephrine Hydrochloride (24, 36 and 48 µg/ml) were analysed at 3 different wave length and % RSD was calculated. % RSD was found to be within acceptable limit that is <2 %. The data for Robustness (different wavelength) are reported in the following table.

Analysis of drugs in immediate release tablet

Analysis of drugs in immediate release (IR) tablet was done by UV method and the % Assay was calculated. % amount of both the drug was found within the limit that is 98-102%.

CONCLUSION

In present work, ratio spectra derivative spectroscopic method was developed for simultaneous estimation of Ketorolac Tromethamine and Phenylephrine Hydrochloride in using water as a solvent. The developed method is robust, accurate and precise. The method is successfully validated as per ICH guidelines.

ACKNOWLEDGEMENT

I would like to thank the Cadila healthcare, Ahmedabad and Tuton pharmaceuticals, Mumbai for supplying the required drugs as a gift sample. I would like to thank Parul Institute of Pharmacy for providing working place for conducting the research.

CONFLICT OF INTERESTS

Declared None

REFERENCES

  1. Indian pharmacopoeia. Ministry of health and family welfare government of India. Published by the Indian pharmacopeia commission, Ghaziabad. Vol. II. 2014. p. 2037.
  2. The British Pharmacopoeia. The department of health Published by the British pharmacopeia commission, London. Vol. II. 2009. p. 1175.
  3. The United States Pharmacopoeia 30/National Formulary 25. Published by the US Pharmacopoeial Convention, Rockville. Vol. II. 2008. p. 2490-91.
  4. Shah J, Maheshwari D. Development and validation of first derivative UV spectrometric method for simultaneous estimation of Fluorometholone acetate and ketorolac tromethamine in ophthalmic dosage form. Indian J Drugs2014;2:56-64.
  5. Bhatt Y, Sharma S, Multani P. A Validated UV Spectro-photometric method for estimation of olopatadine and ketorolac tromethamine in ophthalmic dosage form. Int J Pharm Sci Rev Res 2013;20:118-20.
  6. Sunil G, Jambulingam M, Thangadurai SA, Kamalakannan D. Development and validation of ketorolac tromethamine in eye drop formulation by RP-HPLC method. Arabian J Chem 2012;8:1-8.
  7. Ramakrishna V, Satyanarayana P, Haribabu B. Development and validation of liquid chromatographic method for the simultaneous estimation of ofloxacin and ketorolac tromethamine in combined dosage form. Am J PharmTech Res 2014;4:282-91.
  8. Indian pharmacopoeia. Ministry of health and family welfare Government of India. Published by the Indian pharmacopeia commission, Ghaziabad. Vol. II; 2014. p. 2478.
  9. Indian pharmacopoeia. Ministry of health and family welfare Government of India. Published by the Indian pharmacopeia commission, Ghaziabad Vol. II; 2014. p. 2478-80.
  10. The British Pharmacopoeia. The department of health Published by the British pharmacopeia commission, London. Vol. II; 2009. p. 1609.
  11. The United States Pharmacopoeia 30/National Formulary 25. Published by the US Pharmacopoeial Convention, Rockville. Vol. II; 2008. p. 2985-6.
  12. Soni L, Narsinghani T, Saxena C. Spectrophotometric assay protocol for estimation of ebastine and phenylephrine hydrochloride. Int J Chem Tech Res 2011;3:1918-25.
  13. Ghodasara R, Prajapati A. Spectrophotometric method for simultaneous estimation of Ambroxolol hydrochloride, Levocetrizine dihydrochloride and Phenylephrine hydrochloride in tablet dosage form. Int Res J Pharm 2013;4:197-200.
  14. Parmar K, Bhavsar A, Patel B. Method development and Validation for simultaneous estimation of Ebastine and Phenylephrine in bulk and solid oral dosage form by HPLC. World J Pharm Pharm Sci 2014;3:1312-20.
  15. Bhusan B, Baghel U, Singh R, Kumar Y. RP-HPLC method development for estimation of Levocetrizine and Phenylephrine hydrochloride in combined form. Int J Pharm Biomed Res 2013;1:85-90.
  16. Tandel F, Shah S, Hiren P, Nikita P, Rajesh K. Development and validation of ratio derivative spectrophotometric method for determination of aliskiren hemifumarate and valsartan. Pharmagene 2013;1:49-53.
  17. Patel N, Patel P, Tandel F, Kothari C, Shah S. Ratio derivative spectrophotometric method for simultaneous estimation of olmesartanmedoxomil and atorvastatin calcium in their combined tablet dosage form. Int J Pharm Pharm Sci 2012;4:222-6.