OCULAR DRUG DELIVERY SYSTEM USING OPEN-SOURCE SYRINGE PUMP

Authors

  • Surya V Department of Biomedical Engineering, Sri Ramaswamy Memorial Institute of Science and Technology, Chennai, Tamil Nadu – 603 203, India.
  • Srutartha Bose Department of Biomedical Engineering, Sri Ramaswamy Memorial Institute of Science and Technology, Chennai, Tamil Nadu – 603 203, India.
  • Shreya Kulkarni Department of Biomedical Engineering, Sri Ramaswamy Memorial Institute of Science and Technology, Chennai, Tamil Nadu – 603 203, India.
  • Kathirvelu D Department of Biomedical Engineering, Sri Ramaswamy Memorial Institute of Science and Technology, Chennai, Tamil Nadu – 603 203, India.

DOI:

https://doi.org/10.22159/ajpcr.2018.v11i6.24151

Keywords:

Arduino UNO, Drug delivery, Glaucoma, Leadscrew mechanism, Syringe pump

Abstract

Objective: The objective of this study is to design an automated ocular drug delivery system controlled by an open-source syringe pump, which in turn is to be controlled using Arduino UNO.

Methods: The Arduino UNO is the microcontroller used to control the entire setup. The drug delivery is executed using the open-source syringe pump. A 12V NEMA 17 stepper motor is used to execute the pump movements. To facilitate the ease of use, an Android application developed using Android Studio has been developed, where inputs are given in the form of the required drug dose. The application plays the role of converting the requested drug dose into the number of steps required to run the stepper motor. The stepper motor works based on simple lead-screw mechanism.

Results: The designed syringe pump has been tested using water. It was determined that for 42 steps a drug volume of 30 μl is dispensed by the designed device. Typically, a drop volume ranges from 33.8 μl to 63.4 μl and the device that has been discussed in this article can deliver a minimum volume of 30 μl.

Conclusion: Using this device, excessive use of drugs can be minimalized, thus reducing the capital investment put behind ocular drugs.

Downloads

Download data is not yet available.

References

Michael R. Device for measuring and improving adherence of eye drops; 2016.

Ball SF, Schneider E. Cost of β-adrenergic receptor blocking agents for ocular hypertension. Arch Ophthalmol 1992;110:654-57.

Hermann MM, Diestelhorst M. Microprocessor controlled compliance monitor for eye drop medication. Br J Ophthalmol 2006;90:830-2.

Rylander NR, Vold SD. Cost analysis of glaucoma medications. Am J Ophthalmol 2008;145:106-13.

Wijnen B, Hunt EJ, Anzalone GC, Pearce JM. Open source syringe pump library. PLoS One 2014;9:e107216.

Akash K, Kumar MP, Venkatesan N, Venkatesan M. A Single Acting Syringe Pump Based on Raspberry Pi-SOC; 2015. p. 1-3.

Appaji MS, Reddy SG, Arunkumar S, Venkatesan M. An 8051 microcontroller based syringe pump control system for surface micromachining. Proce Mater Sci 2014;5:1791-800.

Harisudhan R, Kumar MG, Prakash AU Sathya P. Stepper motor control using Arduino-ATMEGA328 microcontroller. Int J Sci Res Dev 2015;2:778-80.

Kumar S, Karki R, Meena M, Prakash T, Rajeswari T, Goli D, et al. Reduction in drop size of ophthalmic topical drop preparations and the impact of treatment. J Adv Pharm Technol Res 2011;2:192-4.

German EJ, Hurst MA, Wood D. Reliability of drop size from multidose eye drop bottles: Is it cause for concern? Eye (Lond) 1999;13:93-100.

Sklubalová Z, Zatloukal Z. Study of eye drops dispensing and dose variability by using plastic dropper tips. Drug Dev Ind Pharm 2006;32:197-205.

Cunningham MG, O’Connor RP, Wong SE. Construction and implantation of a microinfusion system for sustained delivery of neuroactive agents. J Vis Exp 2008;17:7-16.

Chen H. Recent developments in ocular drug delivery. J Drug Target 2015;23:597-604.

Powar PV. Development status in the meadow of nanostructure magnetic drug delivery system and its promising applications. Int J Pharm Pharm Sci 2017;9:10-7.

Rajendran R, Balan R, Ganesan N, Thiruvengadam D. Recent modalities in drug delivery via inhalation therapy-An advanced treatment strategy for pulmonary carcinoma. Int J Pharm Pharm Sci 2015;7:8-21.

Vinupritha P, Hariharan M, Kathirvelu D, Chinnadurai S. Estimation of HbA1c using the complete blood count measures in thediagnosis of diabetes. Asian J Pharm Clin Res 2017;10:214-8.

Alex SM, Sreelekshmi BS, Smitha S, Jiji KN, Menon AS, Umadevi P. Drug utilization pattern of anti-diabetic drugs among diabetic outpatients in a tertiary care hospital. Asian J Pharm Clin Res 2015;8:144-6.

Adepu R, Madhu S. Influence of post discharge counselling on health outcomes in diabetic and hypertensive patients. Asian J Pharm Clin Res 2011;4:28-33.

File RR, Patton TF. Topically applied pilocarpine. Human pupillary response as a function of drop size. Arch Ophthalmol 1980;98:112-5.

Published

07-06-2018

How to Cite

V, S., S. Bose, S. Kulkarni, and K. D. “OCULAR DRUG DELIVERY SYSTEM USING OPEN-SOURCE SYRINGE PUMP”. Asian Journal of Pharmaceutical and Clinical Research, vol. 11, no. 6, June 2018, pp. 152-7, doi:10.22159/ajpcr.2018.v11i6.24151.

Issue

Section

Original Article(s)