BIOANALYTICAL METHOD DEVELOPMENT AND VALIDATION OF CANAGLIFLOZIN IN HUMAN PLASMA BY LIQUID CHROMATOGRAPHY–TANDEM MASS SPECTROMETRY
CANAGLIFLOZIN IN HUMAN PLASMA BY LC-MS/MS
DOI:
https://doi.org/10.22159/ajpcr.2019.v12i18.33228Keywords:
Canagliflozin,, Nil, Human plasma, Nil, Validation, Stability studiesAbstract
Objective: A validated liquid chromatography–tandem mass spectrometry (LC–MS/MS) method was developed for canagliflozin in human plasma along with stability studies.
Methods: The chromatographic separation of canagliflozin was performed on Zorbax XDB phenyl (75 × 4.6 mm, 3.5 mm) using methanol:acetate buffer (80:20 v/v) at a flow rate of 1.0 ml/min. The LC–MS/MS system consists of API 4000 triple quadrupole mass spectrometer equipped with turbospray ionization and an AS8020 automatic sample injector.
Results: The retention time of canagliflozin was 1.15 min and total runtime was 2 min. The multiple reaction monitoring was 462.5/267.1 (m/z) for canagliflozin and 466.4/267.2 (m/z) for internal standard (canagliflozin D4), respectively. The method was linear over the range of 10–7505 ng/ml. The calculated slope ranged from 0.0451 to 0.0502 and intercepts from 0.0102 to 0.0456 with coefficients of the determination of 0.9970. The overall mean recovery of internal standard and canagliflozin was 76.66 and 79.77, respectively.
Conclusion: The method was successfully validated and it was found to be within the limits for accuracy, precision, and linearity and it is stable under analytical conditions used.
Downloads
References
Available from: https://www.drugbank.ca/drugs/DB08907.
Kaushal S, Singh H, Thangaraju P, Singh J. Canagliflozin: A Novel SGLT2 inhibitor for Type 2 diabetes mellitus. N Am J Med Sci 2014;6:107-13.
Kaur I, Wakode S, Singh HP. Development and validation of stability indicating UV spectroscopic method for determination of canagliflozin in bulk and pharmaceutical dosage form. Pharm Methods 2016;7:63-9.
Iqbal M, Khalil NY, Alanazi AM, Al-Rashooda KA. A simple and sensitive high-performance liquid chromatography assay with a fluorescence detector for determination of canagliflozin in human plasma. Anal Methods 2015;7:3028-35.
Sen G, Babu KR, Annapurna N, Vekariya NA. Validation of stability-indicating reverse phase HPLC method for the determination of related substances in canagliflozin drug substance. IOSR J Pharm Biol Sci 2017;12:86-94.
Suneetha A, Sharmila D. A validated stability indicating RP-HPLC method for estimation of canagliflozin in dosage form. Res J Pharm Biol Chem Sci 2015;6:1186-94.
Ashok P, Narenderan ST, Meyyanathan SN, Babu B, Vadivelan R. Development and validation of a RP-HPLC method for estimation of telmisartan in human plasma. Int J Appl Pharm Res 2019;11:237-40.
Kaur I, Wakode S, Singh HP. Development and validation of a stability-indicating high-performance thin layer chromatography (HPTLC) method for estimation of canagliflozin in bulk and pharmaceutical dosage form. J Appl Pharm Sci 2016;6:51-7.
Kobuchi S, Yano K, Ito Y, Sakaeda T. A validated LC-MS/MS method for the determination of canagliflozin, a sodium-glucose co-transporter 2 (SGLT-2) inhibitor, in a lower volume of rat plasma: Application to pharmacokinetic studies in rats. Biomed Chromatogr 2016;30:1549-55.
Prasad PB, Satyanarayana K, Mohan GK. Simultaneous determination of metformin, linagliptin in jentadueto and metformin, saxagliptin in kombiglyze by LC-MS method. Int J Pharm Pharm Sci 2018;10:110-6.
Pinjari MJ, Somani RS, Gilhotra RM. A rapid, sensitive and validated ultra-performance liquid chromatography and tandem mass spectrometry method for determination of paromomycin in Mice plasma: Application to pharmacokinetic study. Int J Pharm Pharm Sci 2017;9:86-90.
Guidance for Industry Bioanalytical Method Validation; 2013.
Published
How to Cite
Issue
Section
The publication is licensed under CC By and is open access. Copyright is with author and allowed to retain publishing rights without restrictions.