SIMULTANEOUS ESTIMATION AND METHOD DEVELOPMENT FOR L-CARNITINE AND METFORMIN IN HUMAN PLASMA USING LC-MS/MS

Authors

  • JAMES D TERISH BINO Annamalai University
  • Kannappan Faculty of Pharmacy, Annamalai University
  • Sasi Jith Sequent Research Limited, Mangalore
  • Suresh Kumar Sequent Research Limted, Mangalore

Abstract

 

Objectives: The objective was to develop a robust simultaneous method for quantifying L-Carnitine and metformin using liquid chromatography
mass spectrometer (LCMSMS).
Methods: LCMSMS method was developed considering its selectivity, sensitivity, and accuracy. Blank correction method was followed as L-Carnitine
is found endogenously. Range was selected based on the Normal value for L-Carnitine and maximum concentration for metformin.
Results: Rugged simultaneous method using ultra flow LC mass spectrometer for the estimation of L-Carnitine and metformin, in human plasma
with low limit of quantification and upper limit of quantification of 2.289 μg/mL and 33.675 μg/mL for L-Carnitine, 43.483 ng/ml and 639.450 ng/ml
for metformin, respectively, was developed which can be used for therapeutic drug monitoring and as a prognostic tool for Type 2 diabetes mellitus
(T2DM). Extraction was optimized using Protein precipitation (PP) in which sample clean-up was simple, and recovery was consistent. Linearity was
checked and found to be within acceptance criteria with 1/×2 as the weighing factor. As L-Carnitine is present endogenously blank correction was
carried by injecting blank six times and calculated for normalization. As the extraction is by simple PP internal standard was not used, and method is
meeting the criteria for validation using LCMSMS.
Conclusion: Hence, considering L-Carnitine as the biomarker for T2DM and metformin as the drug for treatment, a simultaneous method was
developed which can be validated as per regulatory requirements and can be verified for the applicability of the method as a therapeutic drug
monitoring and prognostic tool.
Keywords: L-Carnitine, Metformin, Type 2 diabetes, Liquid chromatography mass spectrometer.

Downloads

Download data is not yet available.

References

Mingrone G, Greco AV, Capristo E, Benedetti G, Giancaterini A, De Gaetano A, et al. L-carnitine improves glucose disposal in type 2 diabetic patients. J Am Coll Nutr 1999;18(1):77-82.

Sowell J, Fuqua M, Wood T. Quantification of total and free carnitine in human plasma by hydrophilic interaction liquid chromatography tandem mass spectrometry. J Chromatogr Sci 2011;49(6):463-8.

Zhang W, Han F, Zhao H, Lin ZJ, Huang QM, Weng N. Determination of metformin in rat plasma by HILIC-MS/MS combined with Tecan automation and direct injection. Biomed Chromatogr 2012;26(10):1163‑9.

Heinig K, Bucheli F. Fast liquid chromatographic-tandem mass spectrometric (LC-MS-MS) determination of metformin in plasma samples. J Pharm Biomed Anal 2004;34(5):1005-11.

Bioanalytical Method Validation - Food and Drug Administration. Available from: http://www.fda.gov/downloads/drugs/guidancecomplianceregulatoryinformation/guidances/ucm368107.pdf

Q2(R1) – ICH. Available from: http://www.ich.org/fileadmin/Public_Web_Site/ICH_Products/Guidelines/Quality/Q2_R1/Step4/Q2_R1__Guideline.pdf.

Published

01-01-2015

How to Cite

BINO, J. D. T. ., Kannappan, S. Jith, and S. Kumar. “SIMULTANEOUS ESTIMATION AND METHOD DEVELOPMENT FOR L-CARNITINE AND METFORMIN IN HUMAN PLASMA USING LC-MS/MS”. Asian Journal of Pharmaceutical and Clinical Research, vol. 8, no. 1, Jan. 2015, pp. 185-91, https://journals.innovareacademics.in/index.php/ajpcr/article/view/3897.

Issue

Section

Original Article(s)