A COMPARATIVE STUDY OF OPTICAL BIOMETRY AND IMMERSION A-SCAN ULTRASOUND IN PATIENTS UNDERGOING PHACOEMULSIFICATION WITH FOLDABLE INTRAOCULAR LENS IMPLANTATION SURGERY

Authors

  • SUMIT DILIP DONGARE Department of Ophthalmology, Bharati Vidyapeeth (Deemed to be) University Medical College and Hospital, Sangli, Maharashtra, India.
  • AJIT KAMALAKAR JOSHI Department of Ophthalmology, Bharati Vidyapeeth (Deemed to be) University Medical College and Hospital, Sangli, Maharashtra, India.
  • NISARG PACHAURI Department of Ophthalmology, Bharati Vidyapeeth (Deemed to be) University Medical College and Hospital, Sangli, Maharashtra, India.

DOI:

https://doi.org/10.22159/ajpcr.2024.v17i2.50443

Keywords:

Cataract, Phacoemulsification, Optical biometry, Immersion A-scan

Abstract

Objective: The objective of this study was to compare optical biometry with immersion A-scan ultrasound biometry in terms of axial length (AL) and post-operative refractive error by assessing 1-month post-operative refraction in patients undergoing phacoemulsification with foldable intraocular lens (IOL) implantation surgery.

Methods: The study was carried out in the Ophthalmology Department of Bharati Vidyapeeth (Demeed to be University) Medical College and Hospital Sangli, from November 2019 to April 2021. A total of 60 eyes of 60 patients were included in the study. All patients underwent both techniques of biometry, namely, optical and immersion A-scan biometry. Mean AL was calculated and compared between the two methods. Then patients were divided into two groups: Group A and Group B; randomization was done on the basis of odd and even numbers. All patients underwent phacoemulsification with foldable IOL implantation surgery and followed up on 1 week and then on 1 month. All patients were operated by single surgeon and a single technique was used. Actual post-operative refractive error, that is, mean of spherical equivalent was compared between two groups on 1-month follow-up.

Results: At 1-month follow-up, actual post-operative refractive error was obtained after calculating spherical equivalent for all the patients and we found that, the mean of actual post-operative refractive error for Group A was higher (−0.371±0.24 D) compared to Group B (−0.264±0.16 D) and the comparison was statistically significant (p=0.049).

Conclusion: Optical biometry is slightly more accurate than ultrasound biometry, in terms of accuracy and reproducibility of the IOL power calculation, but ultrasound biometry is adequate in case optical biometry is unavailable.

Downloads

Download data is not yet available.

References

Vashist P, Talwar B, Gogoi M, Maraini G, Camparini M, Ravindran RD, et al. Prevalence of cataract in an older population in India: The India study of age-related eye disease. Ophthalmology 2011;118:272-8.e82. doi:10.1016/j.ophtha.2010.05.020

Eballé AO, Ellong A, Ella GP, Dohvoma VA, Bella AL, Mvogo CE. Secondary cataract: An epidemiologic and clinical survey at the Yaounde Gynaeco-obstetric and paediatric hospital. Clin Ophthalmol 2011;5:847-51. doi:10.2147/OPTH.S19929

Espiritu CR, Bernardo JP Jr. Incision sizes at different stages of phacoemulsification with foldable intraocular lens implantation. J Cataract Refract Surg 2009;35:2115-20. doi:10.1016/j. jcrs.2009.06.037

Khoramnia R, Auffarth G, Łabuz G, Pettit G, Suryakumar R. Refractive outcomes after cataract surgery. Diagnostics (Basel) 2022;12:243. doi:10.3390/diagnostics12020243

Chung J, Bu JJ, Afshari NA. Advancements in intraocular lens power calculation formulas. Curr Opin Ophthalmol 2022;33:35-40. doi:10.1097/ICU.0000000000000822

Kassa MS, Gessesse GW. Ocular biometry and intra ocular lens power among cataract patients in rural eastern Ethiopia. Ethiop J Health Sci 2021;31:823-30. doi:10.4314/ejhs.v31i4.17

VanderVeen DK, Nizam A, Lynn MJ, Bothun ED, McClatchey SK, Weakley DR, et al. Predictability of intraocular lens calculation and early refractive status: The infant Aphakia treatment study. Arch Ophthalmol 2012;130:293-9. doi:10.1001/archophthalmol.2011.358

Petrella L, Perdigão F, Caixinha M, Santos M, Lopes M, Gomes M, et al. A-scan ultrasound in ophthalmology: A simulation tool. Med Eng Phys 2021;97:18-24. doi:10.1016/j.medengphy.2021.09.005

Whang WJ, Yoo YS, Kang MJ, Joo CK. Predictive accuracy of partial coherence interferometry and swept-source optical coherence tomography for intraocular lens power calculation. Sci Rep 2018;8:13732. doi:10.1038/s41598-018-32246-z

Ison M, Scott J, Apel J, Apel A. Patient expectation, satisfaction and clinical outcomes with a new multifocal intraocular lens. Clin Ophthalmol 2021;15:4131-40. doi:10.2147/OPTH.S327424

Abdelaziz A, Mousa A. Ocular axial length measurement using regular ultrasound and IOL master for different refractive errors in Egyptician population. Med J Cairo Univ 2014;82:159-65.

Joshi AK, Tripthi AK, Parashar AD, Patil SR. Comparative evaluation of accuracy of immersion ultrasound biometry and optical biometry in cases undergoing small incision cataract surgery. J Clin Diagn Res 2019;13:NC04-8.

Fontes BM, Fontes BM, Castro E. Intraocular lens power calculation by measuring axial length with partial optical coherence and ultrasonic biometry. Arq Bras Oftalmol 2011;74:166-70.

Attar HH. Measurement of axial length and intraocular lens power using IOL master and A scan biometry. Pharm Sci Res 2017;9:1975-8.

Németh J, Fekete O, Pesztenlehrer N. Optical and ultrasound measurement of axial length and anterior chamber depth for intraocular lens power calculation. J Cataract Refract Surg 2003;29:85-8.

Eleftheriadis H. IOL master biometry refractive results of 100 consecutive cases. Br J Ophthalmol 2003;87:960-3.

Published

07-02-2024

How to Cite

DONGARE, S. D., A. K. JOSHI, and N. PACHAURI. “A COMPARATIVE STUDY OF OPTICAL BIOMETRY AND IMMERSION A-SCAN ULTRASOUND IN PATIENTS UNDERGOING PHACOEMULSIFICATION WITH FOLDABLE INTRAOCULAR LENS IMPLANTATION SURGERY”. Asian Journal of Pharmaceutical and Clinical Research, vol. 17, no. 2, Feb. 2024, pp. 54-57, doi:10.22159/ajpcr.2024.v17i2.50443.

Issue

Section

Original Article(s)