A STATISTICAL APPROACH BASED ON THE TOTAL ERROR CONCEPT FOR VALIDATION THE BIOANALYTICAL METHOD: APPLICATION TO THE SPECTROPHOTOMETRIC DETERMINATION OF TRACES AMOUNT OF ACETAMINOPHEN IN HUMAN PLASMA

Authors

  • Ihssane Bouchaib Application Organic Chemistry Laboratory, Faculty of Sciences and Techniques, BP 2202 route d’immouzer-Fes-University Sidi Mohamed Ben Abedallah-Morocco
  • Sbai Mohamed Application Organic Chemistry Laboratory, Faculty of Sciences and Techniques, BP 2202 route d’immouzer-Fes-University Sidi Mohamed Ben Abedallah-Morocco
  • El Hadrami El Mestafa Application Organic Chemistry Laboratory, Faculty of Sciences and Techniques, BP 2202 route d’immouzer-Fes-University Sidi Mohamed Ben Abedallah-Morocco
  • Achour Sanae Central Laboratory of analysis Medical, University Hospital Hassan II-Fes, Morocco
  • Saffaj Taoufiq Application Organic Chemistry Laboratory, Faculty of Sciences and Techniques, BP 2202 route d’immouzer-Fes- University Sidi Mohamed Ben Abedallah-Morocco. Regional University Center of Interface, University Sidi Mohamed Ben Abedallah, BP 2626 route d’immouzer-Fes, Morocco

Keywords:

Error total, Analytical validation, Uncertainty profile, Accuracy profile, Measurement uncertainty, Bioanalytical method

Abstract

Objective: The use of the classical approach of analytical validation, in practice or in the literature, is common. However, statistical verification, that looks separately the two errors (such as bias and precision) to make a decision, presents a risk to declare that an analytical method is valid while it is not, or conversely. To minimize this risk, a new approach based on the concept of total error was proposed.

Methods: This approach proposes a calculation the two sided tolerance interval by combining the two errors; bias and precision, in order to examine the validity of an analytical and bioanalytical method at each concentration level. In this paper, we aim to demonstrate the applicability and simplicity of the both methods based on the total error approach: accuracy profile and uncertainty profile. This study will be illustrated by validation case of a spectrophotometric method for the determination of trace amounts of acetaminophen in human plasma.

Results: After the introduction of the correction coefficient which is worth 1.16, the results obtained with accuracy profile approach show clearly that the bioanalytical method is valid over a concentrations range of [100.34- 500] µg mL-1 since the upper and lower 90%-expectation tolerance limits have fallen within the two acceptance limits of ± 20%. The same results found using the uncertainty profile approach because the "two - sided 66.7%-content, 90% -confidence tolerance intervals "are found within two acceptance limits of ± 20% over the range of [170; 500] µm mL-1.

Conclusion: The excellence of the total error approach was showen since it enables successfully to validate the analytical procedure as well the calculation of the measurement uncertainty at each concentration level.

Downloads

Download data is not yet available.

References

Eurachem guide, the fitness for purpose of analytical methods, a laboratory guide to method validation and related topics, 1st Edition UK; 1998.

Guidance for Industry: Bioanalytical Method Validation, US Department of Health and Human Services, US Food and Drug Administration (FDA), Center for Drug Evaluation and Research (CDER), Center for Biologics Evaluation and Research (CBER), Rockville; 2001.

Text on validation of analytical procedures: definitions and terminology (Q2A), Tripartite International Conference on Harmonization (ICH) text, ICH Tech coordination; 1994.

Text on validation of analytical procedures: methodology (Q2B), Tripartite International Conference on Harmonization (ICH) text, ICH Tech coordination; 1995.

International Standards Organization, General Requirements for the Competence of Calibration and Testing Laboratories, ISO/IEC Guide 17025, ISO, Geneva; 1999.

Bouabidi A, Rozet E, Fillet M, Ziemons E, Chapuzet E, Mertens B, et al. Critical analysis of several analytical method validation strategies in the framework of the fit for purpose concept. J Chromatography A 2010;1217(19):3180–92.

Feinberg M, Laurentie M. A global approach to method validation and measurement uncertainty. Accred Qual Assur 2006;11(1-2):3-9.

Feinberg M, Boulanger B, Dewé W, Hubert P. New advances in method validation and measurement uncertainty aimed at improving the quality of chemical data. Anal Bioanal Chem 2004;380(3):502–14.

Hubert Ph, Nguyen-Huu JJ, Boulanger B, Chapuzet E, Chiap P, Cohen N, et al. Harmonization of strategies for the validation of quantitative analytical Procedures-A SFSTP proposal-part I. J Pharm Biomed Anal 2004;36(3):579–86.

Hubert Ph, Nguyen-Huu JJ, Boulanger B, Chapuzet E, Chiap P, Cohen N, et al. Harmonization of strategies for the validation of quantitative analytical procedures-A SFSTP proposal-Part II. J Pharm Biomed Anal 2007;45(1):70-81.

Hubert Ph, Nguyen-Huu JJ, Boulanger B, Chapuzet E, Cohen N, Compagnon PA, Dewé W, Feinberg M, Lallier M, Laurentie M, Mercier M, Muzard G, Nivet C, Valat L, Rozet E. Harmonization of strategies for the validation of quantitative analytical procedures-A SFSTP proposal-Part III. J Pharm Biomed Anal 2007;45(1):82-96.

Mee RW. β-expectation and β-content tolerance limits for balanced one-way ANOVA random model. Technometrics 1984;26(3):251-4.

Saffaj T, Ihssane B. Uncertainty profiles for the validation of analytical methods. Talanta 2011;85(3):1535-42.

Saffaj T, Ihssane B. Response to comments on Uncertainty profiles for the validation of analytical methodsâ€. Talanta 2012;94(1):361-2.

Saffaj T, Ihssane B. Remarks on ‘‘Reply to the responses to the comments on ‘‘uncertainty profiles for the validation of analytical methods’’ by Saffaj and Ihssane’’. Talanta 2012;106(1):155-7.

Saffaj T, Ihssane B, Jhilal F, Bouchafra H, Laslami S, Sosse SA. An overall uncertainty approach for the validation of analytical separation methods. Analyst 2013;138(16):4677-91.

Saffaj T, Ihssane B. Comments on innovative method for carbon dioxide determination in human postmortem cardiac gas samples using headspace-gas chromatography–mass spectrometry and stable labeled isotope as internal standard†by Varlet et al. Anal Chim Acta 2014;810(1):39-42.

Feinberg M. Labo-Stat: guide de validation des méthodes d’analyse; ed. by Lavoisier; 2009. p. 239-42.

Feinberg M. Validation of analytical methods based on accuracy profiles. J Chromatography A 2007;1158(1-2):174–83.

Bouabidi A, Talbi M, Bouklouze A, El Karbane M, Bourichi B, El Guezzar M, et al. Do placebo based validation standards mimic real batch products behavior? Case studies. J Pharm Biomed Anal 2011;55(3):583–90.

Gibelin N, Dupont D, Imbert S, Rozet E. Use of Total Error concept in the validation of viral activity in cell cultures. J Chromatography B 2009;877(23):2407–11.

Rozet E, Hubert C, Ceccato A, Dewé W, Ziemons A, Moonen F, et al. Using tolerance intervals in pre-study validation of analytical methods to predict in-study results. The fit-for-future-purpose concept. J Chromatography A 2007;1158(1-2):126–37.

Dao T, Nguyen T, Guillarme D, Rudaz S, Veuthey J. Validation of an ultra-fast UPLC-UV method for the separation of antituberculosis tablets. J Sep Sci 2008;31(6-7):1050-6.

Rozet E, Ceccato A, Hubert C, Ziemons E, Oprean R, Rudaz S, et al. Analysis of recent pharmaceutical regulatory documents on analytical method validation. J Chromatography A 2007;1158(1-2):111–25.

Hoffman D, Kringle R. A total error approach for the validation of quantitative analytical methods. Pharm Res 2007;24(6):1157-64.

Gonzalez AG, Herrador MA. Accuracy profiles from uncertainty measurements. Talanta 2006;70(4):896-901.

Gonzalez AG, Herrador MA, Asuero AG. Practical digest for evaluating the uncertainty of analytical assays from validation data according to the LGC/VAM protocol. Talanta 2005;65(4):1022-30.

Quintela M, Baguena J, Gotor G, Blanco MJ, Broto F. Estimation of the uncertainty associated with the results based on the validation of chromatographic analysis procedures: application to the determination of chlorides by high performance liquid chromatography and of fatty acids by high resolution gas chromatography. J Chromatography A 2012;1223(3):107-17.

Galban J, Ubide C. Uncertainty due to the quantification step in analytical methods. Talanta 2007;71(3):1339-44.

Hund E, Massart DL, Smeyers-Verbeke J, Hund E, Massart DL, Smeyers-Verbeke J. Comparison of different approaches to estimate the uncertainty of a liquid chromatographic assay. Anal Chim Acta 2003;480(1):39-52.

De Beer JO, Baten P, Nsengyumva C, Smeyers-Verbeke J. Measurement uncertainty from validation and duplicate analysis results in HPLC analysis of multivitamin preparations and nutrients with different galenic forms. J Pharm Biomed Anal 2003;32(4-5):767-811.

Dehouck P, Vander Heyden Y, Smeyers-Verbeke J, Massart DL, Crommen J, Hubert P, et al. Determination of uncertainty in analytical measurements from collaborative study results on the analysis of a phenoxymethylpenicillin sample. Anal Chim Acta 2003;481(2):261-72.

Dıaz A, Vazquez L, Ventura F, Galceran MT. Estimation of measurement uncertainty for the determination of nonylphenol in water using solid-phase extraction and solid-phase micro extraction procedures. Anal Chim Acta 2004;506(1):71-80.

Maroto A, Riu J, Boque R, Rius XT. Estimating uncertainties of analytical results using information from the validation process. Anal Chim Acta 1999;391(2):173-85.

Gonzalez AG, Herrador MA. A practical guide to analytical method validation, including measurement uncertainty and accuracy profiles. TrAC, Trends Anal Chem 2007;26(3):227-38.

Hoffman D, Kringle R. Two-sided tolerance intervals for balanced and unbalanced random effects models. J Biopharm Stat 2005;15(2):283-93.

Hoffman D. Statistical considerations for assessment of bioanalytical incurred sample. J AAPS 2009;11(3):570-80.

Published

01-11-2014

How to Cite

Bouchaib, I., S. Mohamed, E. H. El Mestafa, A. Sanae, and S. Taoufiq. “A STATISTICAL APPROACH BASED ON THE TOTAL ERROR CONCEPT FOR VALIDATION THE BIOANALYTICAL METHOD: APPLICATION TO THE SPECTROPHOTOMETRIC DETERMINATION OF TRACES AMOUNT OF ACETAMINOPHEN IN HUMAN PLASMA”. International Journal of Pharmacy and Pharmaceutical Sciences, vol. 6, no. 11, Nov. 2014, pp. 413-21, https://journals.innovareacademics.in/index.php/ijpps/article/view/3093.

Issue

Section

Original Article(s)