Objective: The development and use of bio-sampling techniques for the analysis of COVID-19 drugs oseltamivir and azithromycin using the Dried Blood Spot technique simultaneously using LC-MS/MS aims to obtain optimal conditions and validated analytical methods using LC-MS/MS according to Food and Drug Administration 2018 recommendations.

Methods: Azithromycin and oseltamivir analyses were performed using LC-MS/MS with C18 Acquity® Bridged Ethylene Hybrid (BEH) column 1.7 m, 100 x 2.1 mm. The matrix sample used is Dried Blood Spot (DBS) with azithromycin and Oseltamivir as the raw material and acyclovir as the internal standard. Optimum analytical conditions were obtained on a gradient mobile phase using 0.1% formic acid-methanol solution with a flow rate of 0.2 ml/minute. The quantification of the analysis was carried out using triple quadrupole mass spectrometry with positive electrospray ionization (ESI) mode.

Results: The calibration curve ranged from 0.5 to 160 g/ml, and the Lower Limit of Quantification (LLOQ) achieved was 25.31 and 25.37 ng/ml. Sensitivity, selectivity, linearity, precision, carry-over, accuracy, stability, and recovery were found to be within the suitable limits and fully validated by the guidelines from the Food and Drug Administration 2018.

Conclusion: The method developed successfully passed all of the FDA's 2018 full validation guidelines, with the LLOQ achieved for azithromycin and Oseltamivir was 25.31 and 25.37 ng/ml.

Downloads

Download data is not yet available.

References

WHO. Timeline of WHO’s response to COVID-19; 2020. Available from: https://www.who.int/news-room/detail/29-06-2020-covidtimeline. [Last accessed on 18 Jul 2020]

McClellan K, Perry CM. Oseltamivir: a review of its use in influenza. Drugs. 2001;61(2):263-83. doi: 10.2165/00003495-200161020-00011, PMID 11270942.

Tan Q, Duan L, Ma Y, Wu F, Huang Q, Mao K. Is Oseltamivir suitable for fighting against COVID-19: in silico assessment, in vitro and retrospective study. Bioorg Chem. 2020;104:104257. doi: 10.1016/j.bioorg.2020.104257, PMID 32927129.

Parnham MJ, Erakovic Haber V, Giamarellos Bourboulis EJ, Perletti G, Verleden GM, Vos R. Azithromycin: mechanisms of action and their relevance for clinical applications. Pharmacol Ther. 2014 Aug;143(2):225-45. doi: 10.1016/j.pharmthera.2014.03.003, PMID 24631273.

Bleyzac N, Goutelle S, Bourguignon L, Tod M. Azithromycin for COVID-19: more than just an antimicrobial? Clin Drug Investig. 2020;40(8):683-6. doi: 10.1007/s40261-020-00933-3, PMID 32533455.

Rosenberg ES, Dufort EM, Udo T, Wilberschied LA, Kumar J, Tesoriero J. Association of treatment with hydroxychloroquine or azithromycin with in-hospital mortality in patients with COVID-19 in New York State. JAMA. 2020;323(24):2493-502. doi: 10.1001/jama.2020.8630, PMID 32392282.

Hooff GP, Meesters RJ, van Kampen JJ, van Huizen NA, Koch B, Al Hadithy AF. Dried blood spot UHPLC-MS/MS analysis of Oseltamivir and oseltamivircarboxylate-a validated assay for the clinic. Anal Bioanal Chem. 2011;400(10):3473-9. doi: 10.1007/s00216-011-5050-z, PMID 21537911.

Ghari T, Kobarfard F, Mortazavi SA. Development of a Simple RP-HPLC-UV method for determination of azithromycin in bulk and pharmaceutical dosage forms as an alternative to the USP method. Iran J Pharm Res. 2013;12 Suppl:57-63. PMID 24250672.

Bacharier LB, Guilbert TW, Mauger DT, Boehmer S, Beigelman A, Fitzpatrick AM. Early administration of azithromycin and prevention of severe lower respiratory tract illnesses in preschool children with a history of such illnesses: a randomized clinical trial. JAMA. 2015;314(19):2034-44. doi: 10.1001/jama.2015.13896, PMID 26575060.

Imperi F, Leoni L, Visca P. Antivirulence activity of azithromycin in pseudomonas aeruginosa. Front Microbiol. 2014;5:178. doi: 10.3389/fmicb.2014.00178, PMID 24795709.

Bioanalytical method validation guidance for industry. United States Food and Drug Administration; 2021. Available from: https://www.fda.gov/regulatory-information/search-fda-guidance-documents/bioanalytical-method-validation-guidance-industry. [Last accessed on 16 Aug 2021]

Cabo Calvet E, Ortiz Bolsico C, Baeza Baeza J, Garcia Alvarez Coque M. Description of the retention and peak profile for chromolith columns in isocratic and gradient elution using mobile phase composition and flow rate as factors. Chromatography. 2014;1(4):194-210. doi: 10.3390/chromatography1040194.

Azzahra Rahmadhani C, Harahap Y, Aisyah Rahmania T. HPLC-DAD quantification of favipiravir in whole blood after extraction from volumetric absorptive microsampling devices. J Chromatogr B Analyt Technol Biomed Life Sci. 2023;1215:123547. doi: 10.1016/j.jchromb.2022.123547, PMID 36586341.

Zhang Y, Bala V, Chhonker YS, Aldhafiri WN, John LN, Bjerum CM. A simple, high-throughput and validated LC–MS/MS method for the determination of azithromycin in human plasma and its application to a clinical pharmacokinetic study. Biomed Chromatogr. 2022;36(10):e5443. doi: 10.1002/bmc.5443, PMID 35789011.

Lakeev AP, Abdrashitova NY, Bryushinina OS, Frelikh GA, Tsuran DV, Zyuz’kova YG. Determination of Oseltamivir in human plasma by HPLC-MS/MS. Pharm Chem J. 2023;57(1):116-22. doi: 10.1007/s11094-023-02858-5, PMID 37152765.

Kumar A, Mhatre S, Godbole S, Jha P, Dikshit R. Optimization of extraction of genomic DNA from archived dried blood spot (DBS): potential application in epidemiological research and bio banking. Gates Open Res. 2018;2:57. doi: 10.12688/gatesopenres.12855.3, PMID 31815249, PMCID PMC6883222.

Zhang Y, Bala V, Chhonker YS, Aldhafiri WN, John LN, Bjerum CM. A simple, high-throughput and validated LC-MS/MS method for the determination of azithromycin in human plasma and its application to a clinical pharmacokinetic study. Biomed Chromatogr. 2022;36(10):e5443. doi: 10.1002/bmc.5443, PMID 35789011.

Nagase K, Nishiyama T, Inoue M, Kanazawa H. Temperature responsive chromatography for therapeutic drug monitoring with an aqueous mobile phase. Sci Rep. 2021;11(1):23508. doi: 10.1038/s41598-021-02998-2, PMID 34873248.

Published

07-03-2024

How to Cite

RAHMANIA, T. A., HARAHAP, Y., & SANDY, K. (2024). AZITHROMYCIN AND OSELTAMIVIR QUANTIFICATION METHOD DEVELOPED AND VALIDATED USING LIQUID CHROMATOGRAPHY-TANDEM MASS SPECTROMETRY IN DRIED BLOOD SPOT. International Journal of Applied Pharmaceutics, 16(2), 182–187. https://doi.org/10.22159/ijap.2024v16i2.49051

Issue

Section

Original Article(s)

Most read articles by the same author(s)

<< < 1 2 3